PSA file: C:\Users\sci0\AppData\Local\Sea-Bird\Seasave\Seasave_7.psa Date: 02/11/2018 Instrument configuration file: C:\data\ctd\09P-0749_at39-05.xmlcon Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Frequency channels suppressed : 0 Voltage words suppressed : 0 Computer interface : RS-232C Deck unit : SBE11plus Firmware Version >= 5.0 Scans to average : 1 NMEA position data added : Yes NMEA depth data added : No NMEA time added : No NMEA device connected to : PC Surface PAR voltage added : Yes Scan time added : No 1) Frequency 0, Temperature Serial number : 2265 Calibrated on : 11-Mar-17 G : 4.33180814e-003 H : 6.43843418e-004 I : 2.35638501e-005 J : 2.22412609e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 2670 Calibrated on : 14-Mar-17 G : -9.76668185e+000 H : 1.30706976e+000 I : -3.47412812e-004 J : 1.01242880e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 0749 Calibrated on : 08-Apr-14 C1 : -4.970438e+004 C2 : -4.005750e-001 C3 : 1.365700e-002 D1 : 3.669600e-002 D2 : 0.000000e+000 T1 : 3.027760e+001 T2 : -4.522040e-004 T3 : 4.002070e-006 T4 : 1.971140e-009 T5 : 0.000000e+000 Slope : 0.99992397 Offset : -1.86008 AD590M : 1.283700e-002 AD590B : -8.706643e+000 4) Frequency 3, Temperature, 2 Serial number : 4148 Calibrated on : 11-Mar-17 G : 4.37331760e-003 H : 6.61226926e-004 I : 2.43591904e-005 J : 2.05481777e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 2770 Calibrated on : 14-Mar-17 G : -1.00880142e+001 H : 1.48322515e+000 I : -3.93661281e-003 J : 3.52192948e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Free 7) A/D voltage 1, Oxygen, SBE 43 Serial number : 0444 Calibrated on : 30-Mar-17 Equation : Sea-Bird Soc : 5.32570e-001 Offset : -4.99000e-001 A : -4.68830e-003 B : 2.25720e-004 C : -3.43710e-006 E : 3.60000e-002 Tau20 : 1.09000e+000 D1 : 1.92634e-004 D2 : -4.64803e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 8) A/D voltage 2, Transmissometer, WET Labs C-Star Serial number : CST-854DR Calibrated on : 8Dec16 M : 21.7440 B : -1.2610 Path length : 0.250 9) A/D voltage 3, PAR/Irradiance, Biospherical/Licor Serial number : QSP200L4S-4553 Calibrated on : 21Nov2016 M : 1.00000000 B : 0.00000000 Calibration constant : 3448000000.00000000 Multiplier : 1.00000000 Offset : -0.38230000 10) A/D voltage 4, Fluorometer, WET Labs ECO-AFL/FL Serial number : FLTURT-964 Calibrated on : 2016-Nov-22 Dark output : 0.0610 Scale factor : 1.10000000e+001 11) A/D voltage 5, Turbidity Meter, WET Labs, ECO-NTU Serial number : FLTURT-964 Calibrated on : 2016-Nov-22 ScaleFactor : 5.000000 Dark output : 0.058000 12) A/D voltage 6, Altimeter Serial number : PSA916-40852 Calibrated on : Scale factor : 15.000 Offset : 0.000 13) A/D voltage 7, Free 14) SPAR voltage, Unavailable 15) SPAR voltage, SPAR/Surface Irradiance Serial number : QSR2240-16500 Calibrated on : 1 Aug 2017 Conversion factor : 1447.86680000 Ratio multiplier : 1.00000000 Scan length : 40 --------------------------------------------- Pump Control This setting is only applicable to a custom build of the SBE 9plus. Enable pump on / pump off commands: NO --------------------------------------------- Data Acquisition: Archive data: YES Delay archiving: NO Data archive: C:\data\ctd\at3905999.hex Timeout (seconds) at startup: 60 Timeout (seconds) between scans: 20 --------------------------------------------- Instrument port configuration: Port = COM1 Baud rate = 19200 Parity = N Data bits = 8 Stop bits = 1 --------------------------------------------- Water Sampler Data: Water Sampler Type: SBE Carousel Number of bottles: 24 Port: COM2 Enable remote firing: NO Firing sequence: Sequential Tone for bottle fire confirmation uses PC sound card. --------------------------------------------- Header information: Header Choice = Prompt for Header Information --------------------------------------------- TCP/IP - port numbers: Data acquisition: Data port: 49163 Status port: 49165 Command port: 49164 Remote bottle firing: Command port: 49167 Status port: 49168 Remote data publishing: Converted data port: 49161 Raw data port: 49160 --------------------------------------------- Miscellaneous data for calculations Depth, Average Sound Velocity, and TEOS-10 Latitude when NMEA is not available: 0.00000000 Longitude when NMEA is not available: 0.00000000 Average Sound Velocity Minimum pressure [db]: 20.00000000 Minimum salinity [psu]: 20.00000000 Pressure window size [db]: 20.00000000 Time window size [s]: 60.00000000 Descent and Acceleration Window size [s]: 2.00000000 Plume Anomaly Theta-B: 0.00000000 Salinity-B 0.00000000 Theta-Z / Salinity-Z 0.00000000 Reference pressure [db] 0.00000000 Oxygen Window size [s]: 2.00000000 Apply hysteresis correction: 1 Apply Tau correction: 1 Potential Temperature Anomaly A0: 0.00000000 A1: 0.00000000 A1 Multiplier: Salinity --------------------------------------------- Serial Data Output: Output data to serial port: NO --------------------------------------------- Mark Variables: No variables are selected. --------------------------------------------- Shared File Output: Output data to shared file: NO --------------------------------------------- TCP/IP Output: Raw data: Output raw data to socket: NO XML wrapper and settings: NO Seconds between raw data updates: 0.00000000 Converted data: Output converted data to socket: NO XML format: NO --------------------------------------------- SBE 11plus Deck Unit Alarms Enable minimum pressure alarm: NO Enable maximum pressure alarm: NO Enable altimeter alarm: NO --------------------------------------------- SBE 14 Remote Display Enable SBE 14 Remote Display: NO --------------------------------------------- PC Alarms Enable minimum pressure alarm: NO Enable maximum pressure alarm: NO Enable altimeter alarm: NO Enable bottom contact alarm: NO Alarm uses PC sound card. --------------------------------------------- Options: Prompt to save program setup changes: YES Automatically save program setup changes on exit: NO Confirm instrument configuration change: YES Confirm display setup changes: YES Confirm output file overwrite: YES Check scan length: YES Compare serial numbers: YES Maximized plot may cover Seasave: NO