PSA file: D:\process_sbe9_rosette\9_carousel_settings\9_carousel_psa\9c_20111122_250m.psa Date: 01/31/2012 Instrument configuration file: D:\process_sbe9_rosette\sbe_9c_20110816.con Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Frequency channels suppressed : 0 Voltage words suppressed : 0 Computer interface : RS-232C Scans to average : 1 NMEA position data added : No NMEA depth data added : No NMEA time added : No Surface PAR voltage added : No Scan time added : No 1) Frequency 0, Temperature Serial number : 5144 Calibrated on : 2010/08/10 G : 4.42333517e-003 H : 6.49321897e-004 I : 2.39916854e-005 J : 2.26379830e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 3679 Calibrated on : 2010/08/04 G : -1.01087198e+001 H : 1.52563181e+000 I : 1.43926364e-005 J : 9.68135383e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 114729 Calibrated on : 06/02/09 C1 : -3.966108e+004 C2 : -1.980940e-001 C3 : 1.218650e-002 D1 : 3.592600e-002 D2 : 0.000000e+000 T1 : 3.001725e+001 T2 : -3.301490e-004 T3 : 4.005720e-006 T4 : 2.907080e-009 T5 : 0.000000e+000 Slope : 1.00000000 Offset : 0.00000 AD590M : 1.281500e-002 AD590B : -9.225127e+000 4) Frequency 3, Temperature, 2 Serial number : 2492 Calibrated on : 2010/08/05 G : 4.40005424e-003 H : 6.61191657e-004 I : 2.57849933e-005 J : 2.46118936e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 3677 Calibrated on : 2010/08/04 G : -9.91248176e+000 H : 1.39805027e+000 I : -2.45121786e-003 J : 2.54713906e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Fluorometer, Wetlab ECO-AFL/FL Serial number : FLRTD_1536 Calibrated on : 2010/08/16 Vblank : 0.0513 Scale factor : 6.00000000e+000 7) A/D voltage 1, User Polynomial Serial number : BBRTD-501 Calibrated on : 2010/17/08 Sensor name : ECO-backscatter A0 : -0.03320000 A1 : 0.06633300 A2 : 0.00000000 A3 : 0.00000000 8) A/D voltage 2, Transmissometer, Chelsea/Seatech/Wetlab CStar Serial number : CST-1269DR Calibrated on : 2010/08/18 M : 21.4611 B : -1.2319 Path length : 0.250 9) A/D voltage 3, Free 10) A/D voltage 4, Oxygen, SBE 43 Serial number : 1639 Calibrated on : 2010/08/18 Equation : Sea-Bird Soc : 4.38900e-001 Offset : -5.20200e-001 A : -2.78230e-003 B : 1.58510e-004 C : -2.59070e-006 E : 3.60000e-002 Tau20 : 1.84000e+000 D1 : 1.92630e-004 D2 : -4.64800e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 11) A/D voltage 5, Fluorometer, Wetlab ECO-AFL/FL, 2 Serial number : FLCDRTD-1539 Calibrated on : 2010/08/17 Vblank : 0.0650 Scale factor : 2.50000000e+001 12) A/D voltage 6, User Polynomial, 2 Serial number : UVP001 Calibrated on : 2010/08/27 Sensor name : Underwater Vision Profiler 5 A0 : 0.00000000 A1 : 125.00000000 A2 : 0.00000000 A3 : 0.00000000 13) A/D voltage 7, User Polynomial, 3 Serial number : ISUS - 0155 Calibrated on : 08 Sept 2008 Sensor name : nitrate sensor A0 : -8.37102500 A1 : 27.07651000 A2 : 0.00000000 A3 : 0.00000000 --------------------------------------------- Pump Control This setting is only applicable to a custom build of the SBE 9plus. Enable pump on / pump off commands: NO --------------------------------------------- Data Acquisition: Archive data: YES Delay archiving: NO Data archive: Timeout (seconds) at startup: 20 Timeout (seconds) between scans: 20 --------------------------------------------- Instrument port configuration: Port = COM1 Baud rate = 9600 Parity = N Data bits = 8 Stop bits = 1 --------------------------------------------- Water Sampler Data: Water Sampler Type: SBE Carousel Number of bottles: 10 Port: COM2 Enable remote firing: NO Firing sequence: Sequential Tone for bottle fire confirmation uses PC sound card. --------------------------------------------- Header information: Header Choice = Prompt for Header Information --------------------------------------------- TCP/IP - port numbers: Data acquisition: Data port: 49163 Status port: 49165 Command port: 49164 Remote bottle firing: Command port: 49167 Status port: 49168 Remote data publishing: Converted data port: 49161 Raw data port: 49160 --------------------------------------------- Miscellaneous data for calculations Depth and Average Sound Velocity Latitude when NMEA is not available: 0.00000000 Average Sound Velocity Minimum pressure [db]: 20.00000000 Minimum salinity [psu]: 20.00000000 Pressure window size [db]: 20.00000000 Time window size [s]: 60.00000000 Descent and Acceleration Window size [s]: 2.00000000 Plume Anomaly Theta-B: 0.00000000 Salinity-B 0.00000000 Theta-Z / Salinity-Z 0.00000000 Reference pressure [db] 0.00000000 Oxygen Window size [s]: 2.00000000 Apply hysteresis correction: 1 Apply Tau correction: 1 Potential Temperature Anomaly A0: 0.00000000 A1: 0.00000000 A1 Multiplier: Salinity --------------------------------------------- Serial Data Output: Output data to serial port: NO --------------------------------------------- Mark Variables: Variables: Digits Variable Name [units] ------ --------------------- 0 Scan Count 4 Depth [salt water, m] 7 Conductivity [S/m] 5 Salinity [PSU] --------------------------------------------- Shared File Output: Output data to shared file: NO --------------------------------------------- TCP/IP Output: Raw data: Output raw data to socket: NO XML wrapper and settings: NO Seconds between raw data updates: 0.00000000 Converted data: Output converted data to socket: NO XML format: NO --------------------------------------------- SBE 11plus Deck Unit Alarms Enable minimum pressure alarm: NO Enable maximum pressure alarm: NO Enable altimeter alarm: NO --------------------------------------------- SBE 14 Remote Display Enable SBE 14 Remote Display: NO --------------------------------------------- PC Alarms Enable minimum pressure alarm: NO Enable maximum pressure alarm: NO Enable altimeter alarm: NO Enable bottom contact alarm: NO Alarm uses PC sound card. --------------------------------------------- Options: Prompt to save program setup changes: YES Automatically save program setup changes on exit: NO Confirm instrument configuration change: YES Confirm display setup changes: YES Confirm output file overwrite: YES Check scan length: YES Compare serial numbers: YES Maximized plot may cover Seasave: NO