Backscatter instrument calibration 15 July 2005 Boss, Slade, Loftin ECOVSF 004r red connected via minilogger, logged in procomm LSS546, LSS585 connected via AquaScat system, data read from AquaTalk software, Data Acquisition panel ac-s used to determine bead concentration in horse bucket ac-s scrapped since the water was too cold and was degassing... 1) DIW, ecored-1.txt 2) 6 drops, 2um Duke poly, ecored-2.txt 3) 25 drops, ecored-3.txt 4) 50 drops, ecored-4.txt 5) 100 drops, ecored-5.txt 6) ~175 drops (ran out of beads), ecored-6.txt Look at instrument resonse after saturation 7) added Maalox (mint flavored!), ecored-7.txt 8) doubled Maalox, ecored-8.txt 9) doubled Maalox, ecored-9.txt 10) doubled Maalox, ecored-10.txt 11) doubled Maalox, ecored-11.txt 12) doubled Maalox, ecored-12.txt 13) Black tape in DIW water, ecored-blacktape.txt Voltages read from AquaTalk, each LSS placed in water independently 1) 546 = 0.08 V, 585=0.03 V, Supply 11.98V 2) 546 = 0.14 V, 585 = 0.05 V 3) 546 = 0.47 V, 585 = 0.14 V, numbers seem reasonable down to Supply ~9.25 V 4) 546 = 1.02 V, 585 = 0.30 V 5) 546 = 2.07 V, 585 = 0.57 V 6) 546 = 3.53 V, 585 = 0.97 V, Supply 11.98 7) 546 = 5.00 V, 585 = 3.56 V 8) 546 = 5.00 V, 585 = 5.00 V 9) 546 = 5.00 V, 585 = 5.00 V 10) 546 = 5.00 V, 585 = 5.00 V 11) 546 = 5.00 V, 585 = 5.00 V 12) 546 = 5.00 V, 585 = 5.00 V 13) Tested 546 = 3.53 V for changes in supply voltage (as read by AquaTalk) - things seem good down to about LSS 546 is much more sensitive to concentration (both using "low gain" output) - don't put it close to the bottom for OASIS 2005