Sea-Bird Electronics              Products            Support            Software            Sales            Service            Search/Sitemap

APPLICATION NOTE NO. 48
Entering Calibration Coefficients for the Seapoint Turbidity Meter
Revised August 2010

Print version

The Seapoint Turbidity Meter measures turbidity by detecting scattered light from suspended particles in the water using dual 880 nm light sources and dual silicon photodiode detectors with visible light blocking filters. The design of the optical sensors confines the sensing volume to within 5 centimeters of the sensor windows.

Sensitivity of the Seapoint Turbidity Meter is determined by two control lines that allow the user to change the range and resolution as required for a particular application. Jumper cables may be purchased from Sea-Bird to allow the sensor range to be changed by inserting the jumper cable in line with the original purchased cable.

Gain Sensitivity (mV/FTU) Range (FTU)
100X 200  25
20X 40  125
5X 10  500
1X 750
     

In our SEASOFT V2 suite of programs, edit the CTD configuration (.con or .xmlcon) file using the Configure Inputs menu in Seasave V7 (real-time data acquisition software) or the Configure menu in SBE Data Processing (data processing software). Select OBS/Nephelometers/Turbidity -- Turbidity Meter, Seapoint as a voltage sensor when editing the configuration file; the software prompts for the gain and scale factor. SEASOFT calculates turbidity as:

OUTPUT (FTU) = (500 * Scale Factor * Voltage) / Gain

where

Response to turbidity levels greater than 750 FTU is non-linear. The linear values calculated by our software will be incorrect for water samples with turbidity levels greater than 750 FTU. An approximate response can be calculated using the second order polynomial equation:

Voltage = 2.2 x 10 -3 (FTU) - 3.0 x 10 -7 (FTU2)

 


Application Note Revision History
Date Description
March 1996 Initial release.
September 2001 Incorporate Sea-Bird Windows software.
June 2006 Provide more details on using software.
May 2007 Incorporate Seasave V7.
February 2010 - Change range from < 750 to 750.
- Update Seasoft-Win32 to Seasoft V2.
- Add information on .xmlcon configuration file.
- Update Sea-bird address.
August 2010 SBE Data Processing and Seasave 7.20g software revision: OBS/Nephelometer sensor listing in software changed to OBS/Nephelometer/Turbidity.

[HRule Image]

Last modified: 02-Aug-2010

Sea-Bird Home     Phone: (+1) 425-643-9866     Fax: (+1) 425-643-9954     E-mail: seabird@seabird.com