APPLICATION NOTE NO. 27
Minimizing Strain Gauge Pressure Sensor Errors
Revised February 2010

Print version

Note: See Application Note 27Druck for current production (Druck) strain gauge pressure sensors.

The following Sea-Bird instruments use strain gauge pressure sensors manufactured by Paine Corporation:

Note: A small number of instruments that fall outside of the above categories also have Paine sensors; consult the instrument configuration information supplied with your instrument.

 

DEFINITION OF PRESSURE TERMS

The term psia means pounds per square inch, absolute (absolute means that the indicated pressure is referenced to a vacuum). The Paine sensors as supplied to Sea-Bird are designed to respond to pressure in nominal ranges 0 to 100, 0 - 300, 0 - 500 psia etc.

For oceanographic purposes, pressure is most often expressed in decibars (1 dbar = 1.4503774 psi). A dbar is 0.1 bar; a bar is approximately equal to a standard atmosphere (1 atmosphere = 1.01325 bar). For historical reasons, pressure at the water surface (rather than absolute or total pressure) is treated as the reference pressure (0 dbar); this is the value required by the UNESCO formulas for computation of salinity, density, and other derived variables.

Some oceanographers express pressure in Newtons/meter2 or Pascals (this is the accepted SI unit). A Pascal is a very small unit (1 psi = 6894.757 Pascals), so the mega-Pascal (1 MPa = 106 Pascals) is frequently substituted (1 MPa = 100 dbar).

Since the pressure sensors used in Sea-Bird CTDs are absolute types, they inherently indicate atmospheric pressure (about 14.7 psi) when in air at sea level. Sea-Bird outputs pressure in one of the following ways:

 

RELATIONSHIP BETWEEN PRESSURE AND DEPTH

Despite the common nomenclature (CTD = Conductivity - Temperature - Depth), all CTD instruments measure pressure, which is not quite the same thing as depth. The relationship between pressure and depth is a complex one involving water density and compressibility as well as the strength of the local gravity field, but it is convenient to think of a dbar as essentially equivalent to a meter, an approximation that is correct within 3% for almost all combinations of conductivity, temperature, depth, and gravitational constant.

SEASOFT offers two methods for estimating depth from pressure.

 

CHOOSING THE RIGHT SENSOR

Initial accuracy and resolution are expressed as a percentage of the full scale range for the pressure sensor. Paine sensors offer accuracies of 0.25% (100 - 1500 psia units) and 0.15% (3000 - 15,000 psia units). For best accuracy and resolution, select a pressure sensor full scale range to correspond to no more the greatest depths to be encountered. The effect of this choice on CTD accuracy and resolution is shown below:

Range Maximum Error (meters) Resolution (meters)
 Pressure (psia) Depth (meters)
100 60 0.2 0.01
300 200 0.5 0.03
500 340 0.9 0.05
1,000 680 1.7 0.10
1,500 1,020 2.5 0.15
3,000 2,000 3.0 0.30
5,000 3,400 5.1 0.50
10,000 6,800 10.2 1.00
15,000 10,200 15.3 1.50
       

The meaning of accuracy, as it applies to these sensors, is that the indicated pressure will conform to true pressure to within ± maximum error (expressed as equivalent depth) throughout the sensor's operating range. Note that a 10,000-psia sensor reading + 10 meters at the water surface is operating within its specifications; the same sensor would be expected to indicate 6800 meters ±10 meters when at depth.

Resolution is the magnitude of indicated increments of depth. For example, a 10,000-psia sensor subjected to slowly increasing pressure will produce readings approximately following the sequence 0, 1.00, 2.00, 3.00 (meters). Resolution is limited by the design configuration of the CTD's A/D converter. For example, for the SBE 25 this restricts the possible number of discrete pressure values for a given sample to somewhat less than 8192 (13 bits); an approximation of the ratio 1:7000 is the source of the 0.015% resolution specification indicated for the SBE 25.

Note: SEASOFT presents temperature, conductivity, and other variables as a function of depth or pressure, so the CTD's pressure resolution limits the number of plotted data points in the profile. For example, a CTD with a 10,000-psia sensor might acquire several values of temperature and conductivity during the time required to descend from 1- to 2-meters depth. However, all the temperature and conductivity values will be graphed in clusters appearing at either 1- or 2-meters on the depth axis.

 

ACCURACY CONSIDERATIONS USING HIGH-RANGE SENSORS IN SHALLOW WATER

High-range sensors used in shallow water will generally provide better accuracy than their absolute specifications would indicate. With careful use, they may exhibit accuracy approaching their resolution limits. For example, a 3000-psia sensor has a nominal accuracy (irrespective of actual operating depth) of ±3 meters. Most of the error, however, derives from variation over time and temperature of the sensor's offset, while little error occurs as a result of changing sensitivity. Furthermore, only the offset is of real significance in shallow-water applications (by definition sensitivity error cannot be greater than 0.15%, and thus the contribution to total error from this source cannot be greater than 0.15% of reading).

The primary offset error due to drift over time can be eliminated by letting the CTD take some readings in air before beginning the profile (if your CTD has a high-range -- 3000 psia or more -- pressure sensor, wait 2 minutes before taking the in air reading; see the discussion under TURN-ON TRANSIENTS below). The pressure value indicated is the offset and can be eliminated by making an entry in the configuration (.con or .xmlcon) file of equal magnitude and opposite sign; the same data subsequently reevaluated with SEASOFT will then show an in air reading of 0 pressure.

The second source of offset error is temperature-induced drifts. These can be estimated for the conditions of the profile by observing the pre- and post-cast in air readings and using their mean value in the configuration (.con or .xmlcon) file. Because Paine sensors are carefully temperature compensated, errors from this source will be small. Hysteresis is the term used to describe the failure of pressure sensors to repeat previous readings after exposure to other (typically higher) pressures. Hysteresis errors rise sharply as the sensor's upper pressure limit is approached, and are very small when only the low end of the pressure range is in use.

Note: In our SEASOFT V2 suite of programs, edit the CTD configuration (.con or .xmlcon) file using the Configure Inputs menu in Seasave V7 (real-time data acquisition software) or the Configure menu in SBE Data Processing (data processing software).

 

TURN-ON TRANSIENTS

Sensors with 3000-, 5000-, 10,000-, and 15,000-psia ranges exhibit a turn-on transient caused by self-heating of their internal bridge resistors. The transient lasts for approximately 2 minutes, as indicated by the plots below. The plots below show the character of the turn-on transient for 5000- and 10,000-psia sensors. Note also the CTD's pressure resolution limits, which can be seen in the step-wise change in pressure readings. The transient has a magnitude of about 10 dbar (10,000-psia sensor) and 5 dbar (5,000-psia sensor), approximately 0.15% of full scale range in both cases. This proportionality is maintained in all sensors observed to date. Because of their different internal construction, the turn-on transient is negligible in sensors with ranges of 1500-psia and below.

During calibration, the sensors are allowed to warm-up before calibration points are recorded. Similarly, for best depth accuracy the user should allow the CTD to warm-up for about 2 minutes before beginning a profile; this can be part of the soak time in the surface water. Soaking also allows the CTD housing to approach thermal equilibrium (minimizing the housing's effect on measured temperature and conductivity) and permits a Beckman- or YSI-type dissolved oxygen sensor (if present) to polarize.