APPLICATION NOTE NO. 18-1
SBE 18, 27, and 30, & AMT pH Sensor Calibration (PHFIT Version 2.1)
Revised April 2013
This application note applies to the SBE 18 pH sensor, SBE 27 pH/ORP (Redox) sensor, SBE 30 DO/pH/ORP sensor, and AMT Analysenmesstechnik GmBh pH sensor.
Sea-Bird software calculates pH as:
pH = 7 + (Vout - offset) / (1.98416 x 10 -4 x T * slope) (see Appendix for derivation of equation)
where
T = temperature (ºK)
Vout = output voltage from pH sensor (0 - 5 volts)
Offset and slope = calibration coefficients, determined by a least-squares fit of voltage and pH in a series of buffer solutions, using the measured temperature of the buffer solutions; coefficients are calculated using PHFIT software
Sea-Bird includes a calibration sheet with the shipment. The calibration sheet provides values for offset and slope, which have been input by Sea-Bird in the configuration (.con or .xmlcon) file.
Note: If you purchase an AMT sensor as a separate item, not integrated
with a Sea-Bird CTD, Sea-Bird provides only the calibration sheet from AMT. This
calibration sheet calculates pH as:
pH = a + bV
where a and b are calculated by AMT using a least-squares fit of voltage and pH
in a series of buffer solutions. If you will be integrating the AMT sensor with
a Sea-Bird CTD, use the voltage and pH values from AMT’s calibration sheet
(and 25 ºC for the temperature) to generate the offset and slope as described
below in User Recalibration.
Sea-Bird provides PHFIT software to use when calibrating pH sensors. PHFIT is installed when you install SBE Data Processing (part of the Seasoft software package). PHFIT can be used on computers running Windows XP service pack 2 or later, Windows Vista, Windows 7, or Windows 8.
When needed, recalibrate the pH sensor as follows:
- At the prompt, enter the sensor serial number and the temperature (in ºC) of the buffer solutions.
- At the prompt, enter the pH and output voltage (Vout) for up to 25 buffer solutions. When you have finished, the program outputs the offset and slope, along with the residuals.
- Double click on SBEDataProc.exe.
- In the Configure menu, select the applicable CTD.
- In the dialog box, click Open and select the applicable .con or .xmlcon file for the CTD.
- In the sensor list, double click on the pH sensor.
- Enter the new offset and slope in the dialog box and click OK.
- Click Save or Save As to save the changed configuration (.con or .xmlcon) file.
Vout = offset + [ slope * (R * T / F) * ln (10) * (pH - 7) ]
Where
R = gas constant = 8.31434
F = Faraday constant = 9.64867 x 10 +4
T = temperature (ºK)
Vout = output voltage from pH sensor (0 - 5 volts)
Substituting for R, F, and ln (10):
Vout = offset + [ slope * 1.98416 x 10 -4 * T * (pH - 7) ]
Therefore,
pH = 7 + (Vout - offset) / (1.98416 x 10 -4 x T * slope)
Date | Description |
June 1992 | - Initial release. |
July 2001 | - Add information on AMT pH sensor. |
March 2005 | - Correct typo -- Faraday constant was 9.64867e-04, should be 9.64867e+04. |
February 2010 |
- Change Seasoft-Win32 to Seasoft V2. - Add information on .xmlcon files. - Update address. |
April 2013 | - Update for newer version of PHFIT, which is now installed with Windows software, in same directory as SBE Data Processing. |