Selection Guide -- Products at a Glance![]()
A CTD measures Conductivity, Temperature, and Depth. Despite the name, all CTDs actually measure pressure, which is not quite the same thing as depth. The relationship between pressure and depth is a complex one involving water density and compressibility as well as the strength of the local gravity field. The CTD data can be used to calculate salinity, density, sound velocity, and other parameters of interest. The term CTD is often used today to describe a package that includes the actual CTD as well as auxiliary sensors to measure other parameters (such as dissolved oxygen, pH, turbidity, fluorometer, altimeter, etc.) and a water sampler to collect water samples for later analysis in the lab.
Sea-Bird manufactures CTDs as well as a number of other products. Our products are categorized into six groups, each with unique applications:
Shown below are comparisons of features within each product group, to assist you in selecting the appropriate instrument for your application.
| SBE | Sampling Rate | Channels for Auxiliary Sensors | Memory | Power | Real-Time Data | Comments | |
| Internal | External | ||||||
| 911plus (9plus CTD & 11plus Deck Unit) | 24 Hz |
8 A/D |
16 Mb with optional SBE 17plus V2 |
√ (with optional SBE 17plus V2) |
√ | √ | World's most accurate, high resolution CTD, premium sensors, multi-parameter support, water sampler control |
| 25plus | 16 Hz |
8 A/D; 2 RS-232 |
2 Gb | √ | √ |
√
(may require SBE 36 Deck Unit & PDIM) |
High resolution logging CTD with multi-parameter support, water sampler control with SBE 33 Deck Unit |
| 25 | 8 Hz | 7 A/D | 8 Mb | √ | √ |
√
(may require SBE 36 Deck Unit & PDIM) |
Replaced by SBE 25plus in 2012 |
| 19plus V2 | 4 Hz |
6 A/D; 1 RS-232 |
64 Mb | √ | √ |
√
(may require SBE 36 Deck Unit & PDIM) |
Personal CTD, small, self-contained, adequate resolution, water sampler control with SBE 33 Deck Unit |
| 19plus | 4 Hz | 4 A/D; optional PAR | 8 Mb | √ | √ |
√
(may require SBE 36 Deck Unit & PDIM) |
Replaced by SBE 19plus V2 in 2008 |
| 19 | 2 Hz | 4 A/D | 1 - 8 Mb | √ | √ |
√
(may require SBE 36 Deck Unit & PDIM) |
Replaced by SBE 19plus in 2001 |
| 49 | 16 Hz | √ |
√
(may require SBE 36 Deck Unit & PDIM) |
Intended for towed vehicle, ROV, AUV, or other autonomous profiling applications. Water sampler control with SBE 33 Deck Unit. | |||
| 52-MP | 1 Hz | 1 frequency channel for dissolved oxygen sensor | 28,000 samples | √ | √ | Intended for moored profiling applications on device that is winched up and down from a buoy or bottom-mounted platform | |
| 41 / 41CP | OEM CTD for sub-surface oceanographic float that surfaces at regular intervals, transmits new drift position and in situ measurements to ARGOS satellite system. CTD obtains latest temperature and salinity profile for transmission on each ascent. Also available is a Navis profiling float, Navis profiling float with Biogeochemical sensors, and Navis profiling float with Integrated Biogeochemical sensors. | ||||||
| GPCTD & Slocum CTD | OEM CTD for autonomous gliders. Generic Glider Payload CTD (GPCTD) is modular, low-power profiling instrument that measures C, T, P, and (optional) Dissolved Oxygen. Slocum Glider Payload CTD provides retrofit/replacement for CTDs on Slocum gliders. Designs share many features, but there are differences in packaging, sampling abilities, power consumption, and installation (see individual data sheets). | ||||||
|
Notes: 1. See Application Note 82: Guide to Specifying a CTD. 2. CTDs in gray are no longer in production. See our Model List page for links to manuals for those CTDs. |
|||||||
| Water Sampler | Number of Bottles | Bottle Size (liters) | CTD | Control | Deployment Depth | |
|
Real-Time
Configuration |
Self-Contained
Configuration (auto bottle firing) |
|||||
|
SBE
32 (full size) |
12 | 1.7 - 30 | SBE 9plus |
SBE 11plus V2 Deck Unit |
SBE 17plus V2 | 6800 m (aluminum), 7000 m (titanium) |
| 24 | 1.7 - 12 | |||||
| 36 | Consult factory | |||||
| 12 | 1.7 - 30 | SBE 19, 19plus, 19plus V2, 25, or 25plus |
SBE 33 Deck Unit (can also be used with SBE 49 CTD) |
Auto Fire Module (can also be used with SBE 50 Pressure Sensor) |
||
| 24 | 1.7 - 12 | |||||
| 36 | Consult factory | |||||
|
SBE
32C (compact) |
12 | 1.7 - 8 | SBE 9plus |
SBE 11plus V2 Deck Unit |
SBE 17plus V2 | 6800 m (aluminum), 7000 m (titanium) |
| SBE 19, 19plus, 19plus V2, 25, or 25plus |
SBE 33 Deck Unit (can also be used with SBE 49 CTD) |
Auto Fire Module (can also be used with SBE 50 Pressure Sensor) |
||||
|
SBE
32SC (sub-compact) |
12 | 1.7 or 2.5 | SBE 19, 19plus, 19plus V2, 25, or 25plus |
SBE 33 Deck Unit (can also be used with SBE 49 CTD) |
Auto Fire Module (can also be used with SBE 50 Pressure Sensor) |
6800 m (aluminum), 7000 m (titanium) |
| SBE 55 | 3 or 6 | 4 | SBE 19, 19plus, 19plus V2, 25, or 25plus |
SBE 33 Deck Unit (can also be used with SBE 49 CTD) |
Built-in auto bottle firing, no additional equipment required | 600 m |
|
Notes: 1. Operation of all SBE 32 Carousel Water Sampler models with the SBE 33 Deck Unit requires an optional sea cable / serial data interface in the Carousel. 2. See Application Note 45: Guide to SBE 32 Carousel Water Sampler Configuration Options for information on selecting the appropriate Carousel options for your application. |
||||||
| SBE |
Measures (C, T, P) |
Auxiliary Sensors | Memory | Power | Communication |
Real-Time Data |
Comments | |
| Internal | External | |||||||
| 16plus V2 | C, T, P* | 6 A/D; 1 RS-232 | 64 Mb | √ | √ | RS-232 | √ | Optional pump |
| 16plus | C, T, P* | 4 A/D; optional RS-232 or PAR | 8 Mb | √ | √ | RS-232 or -485 | √ | Replaced by 16plus V2 in 2008 |
| 16 | C, T, P* | 4 A/D | 1 Mb | √ | √ | RS-232 | √ | Replaced by 16plus in 2001 |
| 16plus-IM V2 | C, T, P* | 6 A/D; 1 RS-232 | 64 Mb | √ | Inductive Modem | √ | Optional pump | |
| 16plus-IM | C, T, P* | 4 A/D; optional RS-232 or PAR | 8 Mb | √ | Inductive Modem | √ | Replaced by 16plus-IM V2 in 2008 | |
| 19plus V2 | C, T, P |
6 A/D; 1 RS-232 |
64 Mb | √ | √ | RS-232 | √ | Programmable mode -- profiling or moored |
| 19plus | C, T, P | 4 A/D; optional PAR | 8 Mb | √ | √ | RS-232 | √ | Replaced by 16plus V2 in 2008 |
| 19 | C, T, P | 4 A/D | 1 - 8 Mb | √ | √ | RS-232 | √ | Replaced by 16plus in 2001 |
| 37-SM | C, T, P* | 8 Mb | √ | √ | RS-232 or -485 | √ | ||
| 37-SMP | C, T, P* | 8 Mb | √ | √ | RS-232, RS-485, or SDI-12 | √ | Integral pump | |
| 37-SMP-IDO | C, T, P* | Integrated DO | 8 Mb | √ | √ | RS-232 or -485 | √ | Integral pump |
| 37-SMP-ODO | C, T, P* | Integrated Optical DO | 8 Mb | √ | √ | RS-232, RS-485, or SDI-12 | √ | Integral pump |
| 37-IM | C, T, P* | 8 Mb | √ | Inductive modem | √ | |||
| 37-IMP | C, T, P* | 8 Mb | √ | Inductive modem | √ | Integral pump | ||
| 37-IMP-IDO | C, T, P* | Integrated DO | 8 Mb | √ | Inductive modem | √ | Integral pump | |
| 37-IMP-ODO | C, T, P* | Integrated Optical DO | 8 Mb | √ | Inductive modem | √ | Integral pump | |
| 37-SI | C, T, P* | 8 Mb | √ | RS-232 or -485 | √ | |||
| 37-SIP | C, T, P* | 8 Mb | √ | RS-232 or -485 | √ | Integral pump | ||
| 37-SIP-IDO | C, T, P* | Integrated DO | 8 Mb | √ | RS-232 or -485 | √ | Integral pump | |
| 37-SIP-ODO | C, T, P* | Integrated Optical DO | 8 Mb | √ | RS-232 or -485 | √ | Integral pump | |
| WQM | C, T, P | Integrated DO, chlorophyll fluorescence, turbidity (WQMx adds CDOM fluorescence, replaces turbidity with backscattering) | √ | √ | RS-232 | √ | Jointly developed by Sea-Bird and WET Labs; manufactured by WET Labs | |
| 39plus | T, P* | 64 Mb | √ | Optional | USB & RS-232 | Optional |
Replaces SBE 39 (Feb 2014) |
|
| 39 | T, P* | 32 Mb | √ | Optional | RS-232 | Optional | ||
| 39-IM | T, P* | 32 Mb | √ | Inductive modem | √ | |||
| 56 | T | 64 Mb | √ | USB | ||||
| 26plus | T, P | C optional | 32 Mb | √ | √ | RS-232 |
√
(tides, waves, & wave statistics) |
Wave & tide recorder |
| 26 | T, P | C optional | 8 Mb | √ | √ | RS-232 | Replaced by SBE 26plus in 2004 | |
| 53 | T, P | C optional | 32 Mb | √ | √ | RS-232 | √ | Bottom pressure recorder |
| 54 | T, P | 128 Mb | Optional | √ | RS-232 | √ | Tsunami pressure sensor | |
|
C = conductivity, T = temperature, P = pressure,
DO = dissolved oxygen * = optional Note: Instruments in gray are no longer in production. See our Model List page for links to manuals for those instruments. |
||||||||
| SBE |
Measures (C, T, P) |
Auxiliary Sensors | Memory | Power | Comments | |
| Internal | External | |||||
| 16plus-IM V2 | C, T, P* | 6 A/D; 1 RS-232 | 64 Mb | √ | Optional pump | |
| 16plus-IM | C, T, P* | 4 A/D; optional RS-232 or PAR | 8 Mb | √ | Replaced by 16plus-IM V2 in 2008 | |
| 37-IM | C, T, P* | 8 Mb | √ | |||
| 37-IMP | C, T, P* | 8 Mb | √ | Integral pump | ||
| 37-IMP-IDO | C, T, P* | Integrated DO | 8 Mb | √ | Integral pump | |
| 37-IMP-ODO | C, T, P* | Integrated Optical DO | 8 Mb | √ | Integral pump | |
| 39-IM | T, P* | 32 Mb | √ | |||
| 44 | 1 RS-232 | 30 Kb buffer | √ | √ | Integrates sensor with standard serial interface (current meter, Doppler profiler, etc.) to inductive modem telemetry system | |
| UIMM | 1 RS-232 | 16 Kb Sample Data Memory | √ | Integrates sensor with standard serial interface (current meter, Doppler profiler, etc.) to inductive modem telemetry system; more flexible and economical than SBE 44 | ||
|
C = conductivity, T = temperature, P = pressure,
DO = dissolved oxygen * = optional Notes: 1. See Inductive Modem Module or Surface Inductive Modem (SIM), and Inductive Cable Coupler (ICC) to complete your inductive modem system. 2. See Typical IM Mooring Components for an overview of an inductive modem system, and Application Note 92: Real-Time Oceanography with Inductive Moorings and the Inductive Modem Module (IMM) for background information. 3. Instruments in gray are no longer in production. See our Model List page for links to manuals for those instruments. |
||||||
| SBE | Measures | Auxiliary A/D Sensors | Remote Temperature Sensor Input | NMEA Data Input | Memory | Power | Communication | Real-Time Data | |
| Internal | External | ||||||||
| 21 | T, C | 4 | √ | √ | 64 Mb | √ | RS-232 | √ | |
| 45 | T, C | √ * | √ * | √ | RS-232 | √ | |||
|
C = conductivity, T = temperature * With optional PN 90402 SBE 45 Power, Navigation, and Remote Temperature Interface Box |
|||||||||
| SBE | Sensor Type | Output | Depth (m) |
User- Programmable? |
Can be Integrated with |
Use |
| 43 | membrane | 0 - 5 volts | 600, 7000 |
|
Moored CTDs (with 1.0-mil membrane), or profiling CTDs (with 0.5-mil membrane) |
|
| 43F | membrane | frequency | 600, 7000 |
|
Moored CTDs (with 1.0-mil membrane), or profiling CTDs (with 0.5-mil membrane) |
|
| 63 | optical | RS-232 | 600, 2000, 7000 | √ |
|
Moored CTDs or slow profiling CTDs |
Sea-Bird manufactures many other products, including:
Print version of our Product Quick Guide, with photos and specifications
Specifications are subject to change without notice.
![]()
Sea-Bird Home Phone: (+1) 425-643-9866 E-mail: seabird@seabird.com