APPLICATION NOTE NO. 72
Setting Up WET Labs ECO Sensor with Bio-Wiper™ for integration with SeaCATplus
CTDs (SBE 16plus, 16plus-IM, 19plus)
Revised May 2013
Note: See Application Note 72-V2 for integration with SeaCATplus V2 (16plus V2, 19plus V2).
This application note applies to the following models of the WET Labs ECO sensor with Bio-Wiper:
These sensors can be configured to operate in profiling mode or in moored mode:
When Sea-Bird integrates the ECO with one of our CTDs, we set up the ECO to operate in the mode appropriate for the CTD. However, you may want to change the mode in the future, if:
To change the ECO configuration, connect the ECO directly to the computer and a 12 V power source, using the cable supplied by WET Labs. Configure the ECO using a terminal program or WET Labs' ECOView program.
The required comm port settings are: 19200 baud, 8 data bits, no parity.
Configure ECO for Profiling Mode | Configure ECO for Moored Mode |
Send the following commands: !!!!! $set 0 $pkt 0 $sto |
Send the following commands: !!!!! $set 1 $pkt 1 $sto |
Notes:
|
After you have configured the ECO for your CTD, reconnect the ECO to the CTD.
General
Delay before sampling sets the amount of time to wait after switching on external voltages before sampling. The ECO requires approximately 4 seconds to provide time for the sensor to open the Bio-Wiper before sampling (sensor dependent; see WET Labs documentation). Therefore, set:
Newer SeaCATplus firmware
Sea-Bird added a Bio-Wiper command to the command set, to ensure that the Bio-Wiper automatically closes after sending the status command:
Command | Description |
Biowiper=x (16plus RS-232, and 19plus) or #iiBiowiper=x |
x=Y:
Configuration includes WET Labs sensor with Bio-Wiper.
CTD is powered longer (a total of 8 sec) for status command (DS or #iiDS),
providing 4 sec for Bio-Wiper to open and then shut again if in Moored mode. Note: 4 sec for the Bio-Wiper to open and shut is sufficient for the typical application, with the ECO set up to take a single measurement for each sample. However, the ECO can be user-programmed to take and average a number of measurements for each sample; if averaging multiple measurements, 4 seconds may not provide sufficient time for the Bio-Wiper to close. Use the TV, TVR, or TS command (#iiTV, #iiTVR, or #iiTS for RS-485 or IM) after the status command, as described below for older CTD firmware, to ensure the Bio-Wiper closes for these applications. x=N (default): Configuration does not include WET Labs sensor with Bio-Wiper. |
Older SeaCATplus firmware
In older SeaCATplus firmware, power was not provided for enough time for the Bio-Wiper to close after the status command (DS for 16plus RS-232 and 19plus; #iiDS for 16plus RS-485 and 16plus-IM) was sent if the Bio-Wiper was in Moored mode. If the user did not notice that the Bio-Wiper was still open, and then deployed the CTD on a mooring with a delayed start time, the WET Labs sensor could become fouled, because the Bio-Wiper would not close until after the first sample was taken.
After sending the status command to these
older firmware versions:
Send the TV, TVR, or TS command (#iiTV, #iiTVR, or
#iiTS for RS-485 or IM) and then wait at least 10 sec to cycle the Bio-Wiper open/closed. Verify that the Bio-Wiper
has closed before deploying.
Date | Description |
February 2004 | - Initial release. |
December 2004 | - Add description of how to handle ECO when integrated with a moored instrument (16plus RS-232, 16plus RS-485, 16plus-IM, 19plus in moored mode). |
May 2005 | - Add Bio-Wiper= command for updated firmware for 16plus RS-485, 16plus-IM, and 19plus. |
March 2008 | - Add applicability to V2 SeaCATplus (16plus V2 RS-232, 16plus-IM V2, 19plus V2). |
May 2013 | - Remove V2 SeaCATplus (16plus V2 RS-232, 16plus-IM V2, 19plus V2) from application note, information on those instruments is in separate application note (Application Note 72-V2). |