Sea-Bird Electronics            Products            Support            Software            Sales            Service            Search/Sitemap

Dissolved Oxygen Sensor 
SBE 13
   (Obsolete -- reference only; see Trade-In Program)

See the SBE 43 Dissolved Oxygen Sensor or SBE 63 Optical Dissolved Oxygen Sensor  in place of the SBE 13.

 

The SBE 13 Dissolved Oxygen (DO) Sensor is available in two models:

The sensor is easy to install, service, and calibrate, since the sensor element and associated interface electronics are a modular, self-contained package. The SBE 13 design is electrically isolated and primarily intended as a bolt-on auxiliary sensor for Sea-Bird's SBE 9plus CTD Underwater Unit, but is also well suited to many custom instrumentation applications. A power and signal interface cable and mounting hardware are available separately.

The optional flow-through plenum improves data quality when the SBE 13 is used with the SBE 9 Underwater Unit. Pumping water over the sensor membrane reduces errors caused by oxygen depletion during periods of slow or intermittent flushing (i.e., stopping for bottle samples). Since the SBE 13 is plumbed in-line between the pump and the CTD's conductivity cell, the oxygen measurement is coordinated with the CTD measurement. The plenum is also useful when calibrating the sensor.

The SBE 13 interface electronics outputs voltages proportional to membrane current (oxygen current) and to the sensor element's membrane temperature (oxygen temperature) used for internal temperature compensation. Computation of dissolved oxygen in engineering units is typically done with Sea-Bird's Seasoft© software, which uses the equation described by Owens and Millard (1985, Journal of Physical Oceanography, v15 [5]) for computing dissolved oxygen.

Sea-Bird calibrates the sensor with a solution of air-saturated water and sodium sulfite. These calibration results are tabulated on a certificate furnished with each sensor.

SPECIFICATIONS

Measurement Range: 0 - 15 ml/l

Accuracy¹: 0.1 ml/l

Resolution: 0.01 ml/l

Time Response²:

Power required: 10-24 VDC, 25 mA

Outputs:

Depth and Housing Materials:

¹ Stated accuracy is achievable with frequent field calibrations.
² Time to reach 63% of final value following a step change in oxygen concentration.

ADDITIONAL INFORMATION / LINKS:

Sales Information -- options, accessories, cables, mount kits, spares, etc.