FIELD SERVICE BULLETIN NO. 3
Retrofit for SBE 37 MicroCATs with Optional Pressure
Sensor --
Pressure Port Plug Removal, Oil Refill, & Nylon
Capillary Fitting Installation
April 2001
Print version
Link to all Field Service Bulletins
Equipment Affected
SBE 37 MicroCATs (37-IM, 37-SM,
and 37-SI) that meet all of
the following criteria:
- Paine pressure sensor,
- Full scale pressure < 1500 psi, and
- Serial number < 1789
Note: This Field Service Bulletin does not apply
to the SBE 37-IMP, 37-SMP, or 37-SIP.
Description of Problem
The MicroCAT’s low pressure (< 1500 psia) Paine pressure sensor
is made of stainless steel. When installed in the MicroCAT, the sensor is in
electrical contact with the titanium end cap. The pressure sensor and end cap
pressure port were originally filled with a silicon oil, and a pressure port
plug was used to retain the oil. In some cases, there has been intrusion of salt
water into the pressure sensor cavity, leading to corrosion.
This retrofit kit is being provided to prevent corrosion from occurring in
the future. The kit consists of a pressure port capillary fitting to replace the
pressure port plug and a higher viscosity oil.
Solution
Retrofit
is simple and requires:
- removing the pressure port plug,
- refilling the internal pressure capillary tube with a higher viscosity
oil, and
- installing a nylon pressure capillary fitting
The retrofit procedure follows.
|
|
P/N 50284 MicroCAT Retrofit Oil Fill Kit:
P/N |
Description |
Quantity |
50283 |
Nylon pressure capillary fitting |
1 |
30419 |
Syringe for injecting DC-200 |
1 |
30420 |
Needle (21 gauge) for syringe |
1 |
30421 |
76 mm (3 inch) long FEP tubing (1/32 inch I.D; 1/16 inch O.D.), to place
on end of syringe needle |
1 |
31152 |
Dow Corning DC-200 silicon oil (10,000 centistokes viscosity) |
1 ounce |
To
replace the needle (if necessary):
- Use a 21-gauge surgical needle, minimum 25 mm (1 inch) long.
- Grind off the tip so the needle is no longer sharp.
- Remove any burrs.
- Verify that the needle still has a hole through it.
- Screw-lock the needle into place on the syringe.
|
|
Procedure:
- Seal
the end of the glass conductivity cell (using Tygon tubing or similar
material) to avoid contaminating the cell with oil.
- Unscrew the nylon pressure port plug (P/N 231593) from
the end cap pressure port. The fitting contains silicon oil, so there will
be some spillage.
CAUTION: Be careful not to hit the glass conductivity cell with the
screwdriver, as doing so will break the cell.
|
|
- Rinse
the pressure port with warm, de-ionized water to remove any corrosion.
CAUTION: Do not put a brush or any object in the pressure port. Doing so
may damage or break the pressure sensor.
- Drain all water from the pressure port cavity. Let the cavity dry
overnight or use low-pressure (< 100 psi) air to dry it before
continuing.
- Calibrate the pressure sensor (if desired):
- The pressure port is tapped with a 7/16-20 straight thread to
accept your pressure fitting.
- Calibrate the pressure sensor.
- Remove the pressure fitting.
|
|
- Turn the instrument so that the end cap faces up.
- Fill the pressure sensor internal capillary with oil:
- Fill the syringe with the DC-200 oil provided with the retrofit kit.
NOTE: This oil is of a thicker viscosity than the oil normally used
with Sea-Bird products. The oil was specifically purchased for this
application and should only be used as part of this kit.
- Insert the syringe’s FEP tubing as far as it will go into the pressure
port’s internal capillary.
- Fill the internal capillary with oil, slowly pulling the syringe tubing
out of the capillary as the capillary fills. Make sure to keep the tip of
the syringe tubing below the oil surface as you pull out the syringe. When
filled, a slight meniscus forms at the surface of the end cap.
|
|
|
- With
the end cap still facing up, insert the new nylon pressure capillary fitting
on the end cap. Do not overtighten the fitting. The oil from the internal
capillary should fill the fitting and overflow the top of the external
capillary tube as the fitting is screwed in. If it does not, remove the
capillary fitting and repeat Steps 6 through 8 to fill with additional oil.
- Clean up any spilled oil on the instrument.
- Remove the material used to seal the end of the glass conductivity cell.
|
|
Note: The oil in the nylon pressure capillary fitting transmits
hydrostatic pressure to the pressure sensor inside the instrument, and prevents
corrosion that might occur if the sensor diaphragm was exposed to water. Because
of the viscosity of the silicone oil and capillary action, the silicone oil does
not run out of the external capillary tube. However, due to temperature and
pressure cycling over long periods, it is normal for some oil to slowly leak out
of the external capillary tube. When the oil is not visible or is receding
inside the translucent tube, or if the fitting has been damaged, refill the oil
using the procedure described above.
Sea-Bird Home Phone: (+1) 425-643-9866
E-mail:
seabird@seabird.com