FIELD SERVICE BULLETIN NO. 9
SeaCAT FlashInit
Error
November 2006
Print version Link to all Field Service Bulletins
The following instruments, shipped prior to November 2006 (the firmware version is shown in the first line of the status command response):
A firmware bug causes SeaCATs to store data incorrectly after the FlashInit or #iiFlashInit command (as applicable) is sent, if the customer starts logging before sending the Initialize Logging command. It is anticipated that very few customers will have a problem with this bug; FlashInit and #iiFlashInit map bad blocks and erase FLASH memory, and are typically used only if there are FLASH Read errors in the Status command response.
Send the following command after the instrument has completed processing FlashInit or #iiFlashInit:
This corrects the problem, and the SeaCAT will store data correctly.
Sea-Bird Home Phone: (+1) 425-643-9866 E-mail: seabird@seabird.com