APPLICATION NOTE NO. 21
Instructions for Use of Nickel-Cadmium Battery Packs
Revised February 2010
Note: As of January 2011, Sea-Bird does not manufacture the 90226 Ni-Cad battery charger, because parts are no longer available. We are still able to provide the battery pack, for customers who already have a charger.
Nickel-Cadmium (Ni-Cad) battery packs are shipped uncharged. Follow this procedure to charge and install a battery pack, and to set up the instrument to run with Ni-Cad batteries:
- The posts inside the housing are not placed symmetrically, so the battery pack fits into the housing only when aligned correctly. Looking at the battery bottom cover, note that one circular cutout is closer to the edge than the others, corresponding to the post that is closest to the housing.
- Some older Sea-Bird instruments have PVC plastic chocks (spacers) used to position and reinforce the posts when the instrument is used with loose alkaline batteries. Cut away the silicon rubber securing these chocks and remove them before installing the Ni-Cad pack.
- SBE 17plus V2 SEARAM -- Run SeatermAF. In the Configure menu, select SBE 17plus V2. In the dialog box, on the 17plus Communications tab, select Ni-Cad batteries and click OK or Save As. On SeatermAF's Toolbar, click Program to program the updated configuration into the instrument.
- SBE 19plus SEACAT CTD -- Run Seaterm and connect to the CTD.
- (19plus firmware version < 1.5) Send the BatteryType=1 command to set up for Ni-Cad batteries.
- (19plus firmware version 1.5 and later) Send the BatteryType=nicad command to set up for Ni-Cad batteries.- SBE 19plus V2 SEACAT CTD -- Run SeatermV2, and connect to the CTD. Send the BatteryType=nicad command to set up for Ni-Cad batteries.
- SBE 25 SEALOGGER CTD -- Run Seaterm and connect to the CTD. Send the CC command. The SBE 25 responds with a number of setup prompts; for battery type, select Ni-Cad.
Note: when using Ni-Cad batteries with an SBE 19 (not plus) SEACAT Profiler CTD, no change in the instrument setup is required.
These Ni-Cad cells have a rated capacity of 4.4 amp-hours at temperatures between 0 and 30 degrees C. Ni-Cad batteries lose charge with time and should be re-charged if not used within 30 days.
Remove the battery pack from the instrument housing before recharging. If you recharge while the battery pack is in the housing, you may damage the instrument electronics.
Additional Information:
Application Note 43 -- Operating Instructions for the SBE Universal (Ni-Cad) Battery Charger.