Sea-Bird Electronics              Products            Support            Software            Sales            Service            Search/Sitemap

APPLICATION NOTE NO. 59
A Load-Bearing Underwater Cable for Hand-Hauled, Real-Time Profiling
Revised February 2010

div.gif (3429 bytes)

Print version

Profiling in real time, from small boats not equipped with a winch, calls for hand hauling a CTD using a load-bearing data cable. Sea-Bird can supply cables up to 300 meters (985 feet) in length that provide a practical way to hand-haul SBE 19 / 19plus /19plus V2 SEACAT Profilers or SBE 25 SEALOGGER CTDs, and acquire CTD profile data in real time. In these cases the CTD is powered by its battery, and the cable allows two-way* RS-232 communications (3 conductors - transmit, receive, and common) and bears the weight of the CTD.

The cable is not intended for static working loads above 45 kg (100 lbs); working loads above 18 kg (40 lbs) may be difficult to handle without a winch. The minimum recommended cable bend radius is 10 cm (4 in.) (e.g., 20 cm sheave block nominal diameter). Use with loads exceeding 45 kg (100 lbs) or cable bend radius less than 10 cm (4 inches) will reduce the cable's useful life and void the warranty.

Drawings show the cable assembly and wiring details, and include tables of Sea-Bird part numbers for cables of specific lengths:

If the length you need is not listed on the drawing, contact Sea-Bird with the drawing number and desired length (300 meter [985 feet] maximum).

* Notes:

 

Using Load-Bearing Underwater Cable for Real-Time Data Acquisition with Seasave
  1. Communicate with the CTD via the terminal program* to set up sampling parameters, initialize memory, etc. (see terminal program Help Files).
  2. Configure Seasave for real-time acquisition and display (see Seasave Help files).
  3. Start the CTD sampling via magnetic switch (or with computer command in the terminal program* per CTD manual instructions).
  4. Deploy the CTD to begin profile.
  5. Recover the CTD.
  6. Stop the CTD sampling via magnetic switch (or with computer command in the terminal program*).
  7. Communicate with the CTD via the terminal program* to upload data.

*Note:
Sea-Bird has several terminal programs for communicating with instruments --

Cable Specifications

div.gif (3429 bytes)

Last modified: 23 Feb 2011

Sea-Bird Home     Phone: (+1) 425-643-9866     Fax: (+1) 425-643-9954     E-mail: seabird@seabird.com