APPLICATION NOTE NO. 9
Calculating Calibration Coefficients for Sea Tech Fluorometer & WET Labs Flash Lamp Fluorometer (FLF)
Revised September 2008

Print version

Some Sea-Bird CTDs can be equipped with a Sea Tech or WET Labs Flash Lamp Fluorometer. For these fluorometers, the CTD configuration file requires the scale factor and offset to enable the software to calculate the concentration:

Concentration = (voltage * scale factor / 5) + offset

These fluorometers have internal, switch-selectable ranges. The scale factor is dependent on the fluorometer range:

Fluorometer Range (milligrams/cubic meter) Scale Factor
Sea Tech 0 - 3 3
0 - 10 (factory-set default) 10
0 - 30 30
0 - 100 100
0 - 300 300
0 - 1000 1000
WET Labs FLF 0 - 100 100
0 - 300 (factory-set default) 300
0 - 1000 1000
   

The offset can be calculated by measuring the fluorometer voltage output when its light sensor is completely blocked from the strobe light by an opaque substance such as heavy black rubber:

offset = (-scale factor * output voltage ) / 5

Note: Scale factor and offset can be adjusted to fit a linear regression of fluorometer responses to known chlorophyll a concentrations.

 

Using Seasave V7 to Determine Offset

  1. In Seasave V7, select Configure Inputs; the Configure Inputs dialog box appears. On the Instrument Configuration tab, click Open (open an existing .con file), Create (create a new .confile), or Modify (modify an existing .con file). Input / verify the scale factor for the fluorometer from the table above.
  2. In the Display menu, select Add New Fixed Display Window. In the Fixed Display dialog box:
  1. Double click on the voltage channel used by the fluorometer from the list of available voltage numbers (see CTD Configuration Sheet at beginning of CTD manual for voltage channel assignments as shipped from the factory).
  2. Modify other fixed display setup parameters (such as seconds between updates, font size, etc.) as desired.
  3. Click OK.
  1. Connect the CTD to the desired COMM port on the computer (verify that it matches the COMM port selected on the Serial Ports tab in the Configure Inputs dialog box).
  2. Block the fluorometer light path with a completely opaque substance.
  3. Start logging data in the CTD:
  1. In Seasave V7’s Real-Time Data menu, select Start. The Start Real-Time Data Acquisition dialog box appears.
  1. Click Start to start real-time data acquisition.
  2. Record the fluorometer voltage and use this value in the equation to calculate the offset.
  3. In the Real-Time Data menu, select Stop.
  1. Enter the calculated offset in the .con file (select Configure Inputs to access the .con file).

Application Note Revision History
Date Description
  Initial release.
October 2000 Covers both Sea Tech fluorometer and WET Labs Flash Lamp fluorometer.
September 2008 - Incorporate Seasave V7.
- Previously said that software does not list WET Labs FLF, but it was updated in 2000 to include it.