APPLICATION NOTE NO. 21
Instructions for Use of Nickel-Cadmium Battery Packs
Revised February 2010

Print version

Note: As of January 2011, Sea-Bird does not manufacture the 90226 Ni-Cad battery charger, because parts are no longer available. We are still able to provide the battery pack, for customers who already have a charger.

Nickel-Cadmium (Ni-Cad) battery packs are shipped uncharged. Follow this procedure to charge and install a battery pack, and to set up the instrument to run with Ni-Cad batteries:

  1. To charge the pack, follow the instructions in Application Note 43, which came with the Sea-Bird Universal Battery Charger. The batteries should charge for 15 hours (overcharging will not damage the batteries).
  2. Disconnect the battery pack from the charger and the charger from the power source. Check the voltage at BAT+ and BAT- on the battery cover. It should be approximately 10.8 volts for a 9-cell pack or 14.4 volts for a 12-cell pack.
  3. To install the battery pack, remove the individual alkaline D-cells (if applicable) from your instrument. Lower the Ni-Cad pack (pin jacks up) into the battery compartment.
  1. While pressing the pack firmly into the battery compartment to compress the springs, secure the pack in place with 3 Phillips-head screws and washers. The screws must be fully tightened, with the washers beneath the screw heads, or battery power to the circuitry will be intermittent.
  2. Update the battery type setting for the instrument:

Note: when using Ni-Cad batteries with an SBE 19 (not plus) SeaCAT Profiler CTD, no change in the instrument setup is required.

These Ni-Cad cells have a rated capacity of 4.4 amp-hours at temperatures between 0 and 30 degrees C. Ni-Cad batteries lose charge with time and should be re-charged if not used within 30 days.

Remove the battery pack from the instrument housing before recharging. If you recharge while the battery pack is in the housing, you may damage the instrument electronics.

Additional Information:

Application Note 43 -- Operating Instructions for the SBE Universal (Ni-Cad) Battery Charger.

 


Application Note Revision History
Date Description
  Initial release.
July 2004 - Update with Windows software.
- Remove deployment length information for SeaCAT, which is inconsistent with SBE 19plus (was probably based on old SBE 19 without pump).
Marc 2008 - Update BatteryType= command for SBE 19plus. Was changed from 0 or 1 to alkaline, nicad, or nimh in firmware revision 1.5.
- Add SBE 19plus V2.
February 2010 - Delete SCPlusV2 terminal program, change it to SeatermV2..
- Update address.