APPLICATION NOTE NO. 41
Calculating Calibration Coefficients
for
WET Labs WETStar Fluorometer
(voltage or RS-232 output sensor)
Revised
May 2011
Note: Procedures in this application note are valid for Seasave V7 and SBE Data Processing version 7.21d and later.
This Application Note applies to the use of the WET Labs WETStar fluorometer with Sea-Bird CTDs. The WETStar fluorometer is factory configured to measure Chlorophyll-a, CDOM, Uranine, Rhodamine, or Phycoerythrin concentration, and provides a linear output. See the WET Labs documentation provided with your WETStar for your instrument range.
Note:
The WETStar comes with a calibration sheet that lists values for Clean Water Offset and Scale Factor. Each of these values is supplied in terms of both voltage and counts.
Concentration (μg/l or ppb, as applicable) = (Output – Blank Output) * Scale Factor
where:
- Output (volts for analog output sensor; counts for digital output sensor) = in situ output of the fluorometer
- Blank Output (volts for analog output sensor; counts for digital output sensor) = measured output for a seawater blank (pure, de-ionized water); called Clean Water Offset on WETStar calibration sheet
- Scale Factor (μg/l – volt, μg/l – count, ppb/volt, or ppb/count, as applicable) = multiplier
Note: The configuration file can only be saved as an .xmlcon file (not a .con file) if the RS-232 output WETStar is one of the sensors.
Example Concentration Calculation in Sea-Bird Software:
Blank Output = 0.05 volts and Scale Factor = 14.10 µg/l-volts (from calibration sheet)
Measured voltage from WETStar= Vsample = 4.65 volts
Concentration = (Output – Blank Output) * Scale Factor = (4.65 - 0.05) * 14.10 = 64.9 µg/l
While the factory-supplied Scale Factor can be used to obtain approximate values, field calibration is highly recommended.
Example Calculation of Scale Factor from field calibration:
Seawater sample analysis indicates chlorophyll-a concentration is 50 µg/l when fluorometer reads 3.2 volts; measured signal for a sea water blank is 0.05 volts.
concentration = (Output – Blank Output) * Scale Factor
50 = (3.2 - 0.05) * Scale FactorSolving for Scale Factor:
Scale Factor = (50) / (3.2 - 0.05) = 15.87 µg/lEnter the new Scale Factor in the configuration (.xmlcon) file.
Date | Description |
Initial release. | |
March 2003 | Reflect new calibration procedure used by WET Labs, which now provides Vblank and Scale Factor on calibration sheet. |