APPLICATION NOTE NO. 41
Calculating Calibration Coefficients for
WET Labs WETStar Fluorometer
(voltage or RS-232 output sensor)
Revised May 2011

Print version

Note: Procedures in this application note are valid for Seasave V7 and SBE Data Processing version 7.21d and later.

This Application Note applies to the use of the WET Labs WETStar fluorometer with Sea-Bird CTDs. The WETStar fluorometer is factory configured to measure Chlorophyll-a, CDOM, Uranine, Rhodamine, or Phycoerythrin concentration, and provides a linear output. See the WET Labs documentation provided with your WETStar for your instrument range.

Note:

The WETStar comes with a calibration sheet that lists values for Clean Water Offset and Scale Factor. Each of these values is supplied in terms of both voltage and counts.

 

Setting Up Configuration (.xmlcon) File

  1. Use the Configure Inputs menu in Seasave V7 (real-time data acquisition software), or the Configure menu in SBE Data Processing (post-processing software), to create / modify the .xmlcon file (see software Help files).
  1. The software prompts for Blank Output and Scale Factor, and calculates:

Concentration (μg/l or ppb, as applicable) = (Output – Blank Output) * Scale Factor

where:

Note: The configuration file can only be saved as an .xmlcon file (not a .con file) if the RS-232 output WETStar is one of the sensors.

Example Concentration Calculation in Sea-Bird Software:

Blank Output = 0.05 volts and Scale Factor = 14.10 µg/l-volts (from calibration sheet)

Measured voltage from WETStar= Vsample = 4.65 volts

Concentration = (Output – Blank Output) * Scale Factor = (4.65 - 0.05) * 14.10 = 64.9 µg/l

Field Calibrations

While the factory-supplied Scale Factor can be used to obtain approximate values, field calibration is highly recommended.

Example Calculation of Scale Factor from field calibration:

Seawater sample analysis indicates chlorophyll-a concentration is 50 µg/l when fluorometer reads 3.2 volts; measured signal for a sea water blank is 0.05 volts.

concentration = (Output – Blank Output) * Scale Factor
50 = (3.2 - 0.05) * Scale Factor

Solving for Scale Factor:
Scale Factor = (50) / (3.2 - 0.05) = 15.87 µg/l

Enter the new Scale Factor in the configuration (.xmlcon) file.

 


Application Note Revision History
Date Description
  Initial release.
March 2003 Reflect new calibration procedure used by WET Labs, which now provides Vblank and Scale Factor on calibration sheet.
May 2007 Incorporate Seasave V7.
February 2010 - Change Seasoft-Win32 to Seasoft V2.
- Add information on .xmlcon file.
- Update address.
May 2011 SBE Data Processing and Seasave 7.21d software revision: WET Labs RS-232 output sensors, including Triplet, are compatible with RS-232 sensor channel on SBE 16plus V2, 16plus-IM V2, and 19plus V2.