APPLICATION NOTE NO. 47
Using Biospherical Instruments Surface PAR Light Sensor
with SBE 33 or SBE 36 Deck Unit
Revised
March 2014
If the SBE 33 or 36 Deck Unit includes an A/D converter, it can acquire the output of a Biospherical Instruments Surface PAR sensor and integrate this into the CTD data stream (A/D converter in Deck Unit was optional prior to 2012; now standard). This application note applies to the following Surface PAR sensors (the corresponding underwater PAR sensor, mounted on the CTD or cage, is also shown):
Surface PAR Sensor | Corresponding Underwater PAR Sensor | |
If underwater PAR sensor connected directly to CTD A/D voltage channel |
If underwater PAR sensor connected directly to CTD PAR connector, or interfacing to PN 90310 Log Amp Module (which is connected to CTD A/D voltage channel) |
|
QSR-240 or QSR-2200 | QSP-200L, QSP-2300L, or QSP-2350L | QSP-200(PD) or QSP-2200(PD) |
QCR-240 or QCR-2200 | QCP-200L, QCP-2300L, or QCP-2300L-HP | QCP-2200(PD) |
Notes:
Our Seasoft V2 software fully supports the acquisition and display of the data from these surface reference PAR sensors and the corresponding underwater PAR sensor.
- Deck Unit firmware < 3.0: Locate the NMEA Interface PCB with optional A/D converter under the bottom cover panel of the Deck Unit . Verify that dip switch S1 position 8 is in the OFF position.
- Deck Unit firmware > 3.0: In Seaterm, verify that the SBE 33/36 status message on power-up shows Surface PAR enabled. If it does not, type @ to access the setup menu, and then type 6 and press the Enter key to enable Surface PAR acquisition.
Deck Unit | Function | Biospherical Surface PAR Sensor with Switchcraft Connector |
pin A | Signal (gnd) | pin 3 |
pin B | Power (+12 volts) | pin 4 |
pin C | Power (ground) -- Deck Units with NMEA PCB Assembly 40785b/40786b or greater use Pin C. All previous versions do not use Pin C. | pin 1 |
pin D | Signal | pin 2 |
-- | -- | pin 5 |
Note: Biospherical sold the Surface PAR sensor with other connector types in the past. See the appropriate drawing for pinout details if your sensor does not have a Switchcraft connector. |
Set up the configuration (.con or .xmlcon) file for the appropriate CTD (SBE 16 / 16plus / 16plus V2 / 19 / 19plus / 19plus V2 / 25 / 25plus / 49). Enable the recording of the Surface PAR sensor by selecting Surface PAR voltage added. This alters the display by adding two additional external voltages. The highest numbered voltage is labeled SPAR/surface irradiance. Enter the surface light conversion factor and the ratio multiplier in the configuration file.
Note: In our Seasoft V2 suite of programs, edit the CTD configuration (.con or .xmlcon) file using the Configure Inputs menu in Seasave V7 (real-time data acquisition software) or the Configure menu in SBE Data Processing (data processing software).
Seasoft calculates surface light as:
surface light (μEinstein/m2.sec) = volts * conversion factor
To compute the conversion factor, calculate the output of the Surface PAR sensor in μEinstein/m2.sec/volt using the data from the Surface PAR calibration sheet that was provided by Biospherical (located in the CTD manual).
Conversion factor = Output in Air / Probe Net Response
For example, if Probe Net Response = 85.6 mV (0.0856 volts) and Output in Air = 0.01384 μEinsteins/cm2.sec:
Conversion factor = (0.01384 μEinsteins/cm2.sec) * (10000 cm2/m2) / 0.0856 volts = 1851 μEinsteins/m2.sec/volt
Seasoft calculates corrected PAR as:
corrected PAR = 100 * ratio multiplier * underwater light / surface light
(underwater light is the calculated light output from the underwater sensor)
To compare the shape of data sets taken at disparate light levels, the ratio multiplier can be used to scale the data. For example, a ratio multiplier of 10 would make a 100 μEinsteins/m2.sec light level plot as 1000 μEinsteins/m2.sec. The ratio multiplier should be set to 1 for normal operations.
Notes:
Date | Description |
1995 | Initial release. |
October 2004 | Previously referred to SEACON in discussions of .con files. Add references to modifying .con file using Configure menu in Seasave or SBE Data Processing in Windows software. |
June 2005 | - Update with new (2003)
Biospherical PAR sensor part numbers. - Update Surface PAR sensor connector information. - Expand/rewrite/reorganize text. |
May 2007 | Provide output in microEinsteins/m2 sec, and refer to application note 11General for conversion to other units. |
May 2008 | Add applicability to V2 SeaCATs (16plus V2, 19plus V2). |
March 2010 | - Add information on QSP-2350L (same
functioning as QSP-2300L, just a different connector) and
QCP-2300L-HP (same functioning as QCP-2300L, just higher depth
range), which are on current price list. - Update description of configuring SBE 33 or 36 -- no dip switches in new version (3.0). - Change Seasoft-Win32 to Seasoft V2. - Add information on .xmlcon. - Update address. |
October 2012 | Update for SBE 25plus. |
March 2014 | - Add applicability to SBE 49
(compatible with Seasave 7.23.2 and later). - Update to indicate that Surface PAR capability is standard in SBE 33 and 36 Deck Units sold in 2012 and later. |