APPLICATION NOTE NO. 77
Entering Calibration Coefficients for the Seapoint
Ultraviolet Fluorometer
Revised August 2010
Print version
(Note: This application note is for use with Seasave and SBE Data Processing
version 7.20g and later.
An earlier
version of the application note is compatible with earlier
software versions.
The Seapoint ultraviolet fluorometer is used for in-situ measurements of chromophoric dissolved organic matter (CDOM). This fluorometer uses modulated ultraviolet LED lamps and excitation filter to excite CDOM present in water. The fluorescent light emitted by the CDOM passes through a blue emission filter and is detected by a silicon photodiode. The low level signal is then processed, generating an output voltage proportional to CDOM concentration.
Sensitivity of this fluorometer is determined by two control lines that allow the user to change the range and sensitivity as required for a particular application. Jumper cables may be purchased from Sea-Bird to allow the sensor range to be changed by inserting the jumper cable in line with the original cable purchased.
Gain | Sensitivity [V/( µg/l)] | Range [ µg/l] |
30X | 0.1 | 50 |
10X | 0.033 | 150 |
3X | 0.01 | 500 |
1X | 0.0033 | 1500 |
NOTES:
Date | Description |
May 2005 | Initial release. |
May 2007 | Incorporate Seasave V7. |
February 2010 | - Change Seasoft-Win32 to Seasoft
V2. - Add information on .xmlcon configuration file. - Update Sea-bird address. |
August 2010 | SBE Data Processing and Seasave 7.20g software revision: Seapoint Ultraviolet Fluorometer added to list of fluorometers (previously needed to select User Polynomial for this sensor). |
Sea-Bird Home Phone: (+1) 425-643-9866 E-mail: seabird@seabird.com