Sea-Bird Electronics              Products            Support            Software            Sales            Service            Search/Sitemap

SBE Data Processing - Seasave Revision History
SBE Data Processing and Seasave V7 Version 7.23.2

div.gif (3429 bytes)

Brief description (SBE Data Processing and Seasave V7) | Software manuals (SBE Data Processing and Seasave V7) | Reference Sheet (Seasave V7)
Download instructionsDownload from FTP siteSoftware FAQs

The Revision History describes each version of software -- including bug fixes, modifications, and new features. You may not need the latest version; our revisions often include improvements and new features which may have little or no impact on your operation. Review this site regularly to view the latest changes.

Version # listing:

  7.22.1 7.21d 7.20c 7.17 7.10 5.37a 5.32 5.28f
(SBE Data Processing only)
5.27 5.25
  7.22.0 7.21c
(SBE Data Processing only)
7.20b 7.16a 7.0 & 7.0h
(Seasave only)
5.37 5.31b
(SBE Data Processing only)
5.28e 5.26f (Seasave only) 5.24
7.23.2 7.21k 7.21b 7.20a 7.16 5.37m 5.36 5.31a 5.28d 5.26d (Seasave only) 5.23
7.23.1 7.21j 7.21a 7.19 7.15 5.37e
(SBE Data Processing only)
5.35 5.31 5.28c 5.26c (Seasave only) 5.22
7.22.5a
(SBE Data Processing only)
7.21i 7.21 7.18d 7.14e
(SBE Data Processing only)
5.37d 5.34b
(SBE Data Processing only)
5.30b
(Seasave only)
5.28b 5.26b
(Seasave only)
5.21
7.22.5 7.21h 7.20g 7.18c 7.14d 5.37c
(SBE Data Processing only)
5.34a
(SBE Data Processing only)
5.30a 5.28a 5.26a
(Seasave only)
5.2
7.22.4 7.21g 7.20f 7.18b 7.14c 5.37c
(Seasave only)
5.34 5.30 5.27c 5.26 5.1
7.22.3 7.21f 7.20e 7.18 7.12 5.37b
(SBE Data Processing only)
5.33 5.29b 5.27b 5.25b 5.0
7.22.2 7.21e 7.20d 7.17a 7.11 5.37b
(Seasave only)
5.32a
(SBE Data Processing only)
5.29a 5.27a 5.25a
(Seasave only)
Older versions

 


VERSION 7.23.2
18 March 2014

SBE Data Processing --

1. MODIFICATIONS
a. Update to SBE 49 configuration dialog box, to allow Surface PAR data to be appended when SBE 49 used with SBE 33 or 36 deck unit.
b. Update to SBE 9 configuration dialog box, to add explanation on voltage words.

2. BUG FIXES
a. Derive module – Calculation of SBE43 oxygen and oxygen saturation gave incorrect negative values at temperatures below -2 deg C; calculation has been fixed for temperatures down to -5 deg C (ITS-90).

Seasave --

1. MODIFICATIONS
a. Update to SBE 49 configuration dialog box, to allow Surface PAR data to be appended when SBE 49 used with SBE 33 or 36 deck unit.
b. Update to SBE 9 configuration dialog box, to add explanation on voltage words.


VERSION 7.23.1
17 September 2013

SBE Data Processing --

1. ADDITIONS
a. Add Derive Teos-10 module.
b. Replace Sea Calc II with Sea Calc III, adding and changing TEOS-10 calculations.

2. MODIFICATIONS
a. Update default location of .psa, .ini, and .txt diagnostics files for compliance with more operating systems.
b. Increase # of sensors that can be on CTD from 5 to 6 for the following WET Labs sensors: C-Star, ECO-AFL (or ECO-FL), ECO CDOM, WETStar.
c. Derive module now has a note on the Data Setup tab stating that it is for EOS-80.

3. BUG FIXES
a. Derive -- Calculation of secondary SBE 43 oxygen was incorrect if the input .cnv file had secondary conductivity in mS/cm.

Seasave --

1. MODIFICATIONS
a. Update default location of .psa, .ini, and .txt diagnostics files for compliance with more operating systems.
b. Increase # of sensors that can be on CTD from 5 to 6 for the following WET Labs sensors: C-Star, ECO-AFL (or ECO-FL), ECO CDOM, WETStar.
 


VERSION 7.22.5a
29 April 2013

SBE Data Processing --

1. ADDITIONS
a. Add installation of Windows program phfit.exe, which can be used to calculate new pH coefficients for an SBE 18 or 27 pH sensor. phfit.exe is installed in same directory as SBE Data Processing, and is compatible with a PC running Windows XP service pack 2 or later, Windows Vista, Windows  7, or Windows 8. This version of phfit replaces the version that was installed as part of the old Seasoft-DOS software; phfit in the DOS install is incompatible with Windows 7 and 8.
Double click on phfit.exe, and respond to the prompts to run the software.
Note: SBE Data Processing install file name is SBEDataProcessing_Win32_V7_22_5a.exe (for software version 7.22.5a). However, because the data processing software did not actually change (the change is simply the installation of the phfit software in the same directory as SBE Data Processing), selecting Help / About in SBE Data Processing still shows the version as 2.22.5.


VERSION 7.22.5
21 March 2013

SBE Data Processing --

1. ADDITIONS
a. Add installation of new program StripNULLchars.exe (installed in same directory as SBE Data Processing). This new program removes null characters from uploaded SBE 25plus data file and should be used only if:
    (1) SBE 25plus includes auxiliary serial (RS-232) sensor data that was saved in the .xml file instead of a separate .txt file (SetInline#=y), and
    (2) There are 2 auxiliary serial sensors of the same type on the 25plus, and
    (3) Running the Data Conversion module in SBE Data Processing produced a file with bad flags for the second auxiliary serial sensor data.
Run StripNULLchars.exe from a DOS window, following instructions provided in the software. The data file can then be processed in SBE Data Processing's Data Conversion module.
Note: The null characters in the file also prevent uploading of the data from the 25plus via RS-232. You must open the 25plus and upload via the internal USB connector.

Seasave V7 --

1. MODIFICATIONS
a. Update for compatibility with internal Sea-Bird testing; no impact to customers.


VERSION 7.22.4
04 February 2013

SBE Data Processing and Seasave V7 --

1. MODIFICATIONS
a. In voltage sensor selection list, change name of METS sensor to Franatech (acquired Capsum in 2006).


VERSION 7.22.3
14 January 2013

Seasave V7 --

1. ADDITIONS
a. Add Discover Ports button on Serial  Ports tab in Configure Inputs and Configure Outputs.


VERSION 7.22.2
31 December 2012

SBE Data Processing --

1. BUG FIXES
a. Modified location of temporary files, to prevent failure if user does not have write permission on directory.

Seasave V7 --

No changes; revision number updated to keep synchronized with SBE Data Processing


VERSION 7.22.1
05 December 2012

SBE Data Processing --

1. MODIFICATIONS
a. Derive, SBE 9plus CTD: Previously, Derive calculated SBE 43 dissolved oxygen (DO) using temperature and conductivity values from primary T & C sensors, for both primary and secondary dissolved oxygen sensors, regardless of whether data from secondary T & C sensors was in the input file. Modification forces matching of T, C, & DO sensor outputs for DO calculation; i.e., primary T & C are used for primary DO, secondary T & C are used for secondary DO. If data from a matching sensor pair is not in the input .cnv file, Derive uses data from the other pair (for example, if the input .cnv file contains primary DO and only secondary T & C [perhaps because the primary T & C data indicates sensor malfunctioning or bad cables], it uses the secondary T & C in the calculation).

2. BUG FIXES
a. Derive: Correct calculation of dissolved oxygen in micromol/kg (previously output 0 if those units were selected).
b. SBE 16plus, 16plus-IM, 19plus, 16plus V2, 16plus-IM V2, 19plus V2: Revised so Data Conversion does not use remote temperature from an SBE 38 if calculating secondary SBE 43 dissolved oxygen.
c. SBEBatch and SBE 25plus: Fix bug in conversion of SBE 25plus XML data file.

Seasave V7 --

1. MODIFICATIONS
a. Allow for selection of higher number COM ports (up to 255).

2. BUG FIX
a. SBE 16plus, 16plus-IM, 19plus, 16plus V2, 16plus-IM V2, 19plus V2: Revised so Seasave does not use remote temperature from an SBE 38 if calculating secondary SBE 43 dissolved oxygen.


VERSION 7.22.0
10 August 2012

SBE Data Processing --

1. ADDITIONS
a. Add selection of units for calculating and outputting Specific Conductance.

2. MODIFICATIONS
a. Data Conversion: Increase speed of data conversion.

Seasave V7 --

1. ADDITIONS
a. Add ability to calculate and output Specific Conductance.


VERSION 7.21k
23 May 2012

SBE Data Processing and Seasave V7 --

1. MODIFICATIONS
a. SBE 25plus: Add information on deck unit and baud rate to configuration dialog box, to allow software to correctly calculate elapsed time.


VERSION 7.21j
22 May 2012

SBE Data Processing and Seasave V7 --

1. ADDITIONS
a. Add compatibility with SBE 37-IMP-ODO MicroCAT.
b. Add compatibility with SBE 19plus V2 integrated with SBE 63 Optical Dissolved Oxygen Sensor.
c. Bottle Summary in SBE Data Processing: Add Gordon and Garcia oxygen saturation to list of derived variables.

2. MODIFICATIONS
a. SBE 9plus: Calculate secondary SBE 43 oxygen using temperature and conductivity from secondary T and C sensors.

3. BUG FIXES
a. Fix bug in scan matching in old SBE 19s (not plus) and 25s (not plus), when used with AFM.


VERSION 7.21i
07 May 2012

SBE Data Processing --

1. ADDITIONS
a. Add compatibility with SBE 37-SMP-ODO MicroCAT.


VERSION 7.21h
16 April 2012

SBE Data Processing --

1. ADDITIONS
a. Add compatibility with SBE 25plus.

Seasave V7 --

1. ADDITIONS
a. Add compatibility with SBE 25plus.


VERSION 7.21g
12 March 2012

SBE Data Processing --

1. ADDITIONS
a. Add compatibility with SBE 63 Optical Dissolved Oxygen Sensor integrated with SBE 16plus V2.

Seasave V7 --

1. ADDITIONS
a. Add compatibility with SBE 63 Optical Dissolved Oxygen Sensor integrated with SBE 16plus V2.
b. Add visual alarms.


VERSION 7.21f
23 September 2011

SBE Data Processing --

1. BUG FIXES
a. Derive: If processing SBE 43 dissolved oxygen data and Tau correction was selected, there were sometimes clearly out of range values.

Seasave V7 --

1. BUG FIXES
a. Seasave and Seasave Remote: If some axes were reversed, there were sometimes plot errors when zooming in.


VERSION 7.21e
29 August 2011

SBE Data Processing --

1. MODIFICATIONS
a. Configuration (.xmlcon) file now supports up to 5 of each of the following WET Labs sensors on a CTD: ECO-AFL/FL fluorometer, ECO CDOM fluorometer, WETstar fluorometer, ECO-NTU turbidity meter, ECO-BB turbidity meter, C-star transmissometer.

Seasave V7 --

1. MODIFICATIONS
a. Configuration (.xmlcon) file now supports up to 5 of each of the following WET Labs sensors on a CTD: ECO-AFL/FL fluorometer, ECO CDOM fluorometer, WETstar fluorometer, ECO-NTU turbidity meter, ECO-BB turbidity meter, C-star transmissometer.

2. BUG FIXES
a. SBE 19plus / 19plus V2: If integrated with a bottom contact switch, Seasave alarm turned on when bottom was not in contact, turned off when bottom was in contact.


VERSION 7.21d
19 April 2011

SBE Data Processing --

1. ADDITIONS

a. SBE 16plus V2, 16plus-IM V2, and 19plus V2: Add support for WET Labs RS-232 sensor(s) to configuration files.

2. MODIFICATIONS

a. Sea Plot:
- Plot legend can now be copied and pasted into a document (Word file, etc.).
- Modify Overlay Setup Line Colors, Line Symbols, and Line Types dialog boxes, replacing Set with File in column headings to clarify.
- Warning is now provided if user specifies too many ticks in an axis (previously it provided no warning but did not draw axis properly).
b. Data Conversion:
- Now ignores a 'w' in conductivity sensor serial number in configuration file (indicating wide range conductivity calibration) when checking against serial number in data file. Previously, it would give an error message saying that serial numbers did not match.
- Now does not make a .ros file if no bottles have been fired, even if user specifies to output .ros file.

3. BUG FIXES
a. Filter: Fix bug that caused crashes with large data sets.

Seasave V7 --

1. ADDITIONS
a. SBE 16plus V2, 16plus-IM V2, and 19plus V2: Add support for WET Labs RS-232 sensor(s) to configuration files.

2. MODIFICATIONS
a. Modify Table Driven and Auto Fire bottle firing column headings, to clarify that bottle position numbers are entered in tables.


VERSION 7.21c
03 March 2011

SBE Data Processing --

1. MODIFICATIONS
a. SBE 37-SI, SIP, SIP-IDO MicroCATs: .xmlcon configuration file accepts a (non-integer) 1.5 sec value for Sample Interval, to accommodate 1.5 sampling speed for MicroCAT with pressure sensor that is sampling continuously (SampleMode=3).

2. BUG FIXES
a. Fix several small bugs.


VERSION 7.21b
27 January 2011

SBE Data Processing --

1. BUG FIXES
a.
A few minor bug fixes.

Seasave V7 --

1. MODIFICATIONS
a. Seasave and SeasaveAcq now communicate via anonymous pipes instead of sockets, to resolve problems that some customers were having when running the software in a Windows 7 environment. This resulted in a modification of the TCP/IP Ports tab in Configure Inputs and Configure Outputs.


VERSION 7.21a
10 November 2010

SBE Data Processing --

1. ADDITIONS
a. SBE 9plus with 11plus Deck Unit: Add selection of Deck Unit (and firmware version) or Searam in configuration file.
b. NMEAtest (software that automatically installs when SBE Data Processing is installed): add new features to test NMEA time.

2. MODIFICATIONS
a. Elapsed time in seconds was formatted as an exponential for time > 1,000,000 seconds, resulting in a loss in precision from the 3 digits to the right of the decimal place for smaller numbers. It is not formatted as an exponential now: format with 2 places to right of decimal for values between 1,000,000 and 10,000,000 seconds, and 1 digit to right of decimal for values between 10,000,000 and 100,000,000 seconds.

3. BUG FIXES
a. SBE 45: If configuration file was set to output salinity but not conductivity, software output 0 for all salinity values.

Seasave V7 --

1. ADDITIONS
a. SBE 9plus with 11plus Deck Unit: Add selection of Deck Unit (and firmware version) or Searam in configuration file. For Deck Unit with firmware version > 5.0, Seasave can now send AddSPAR= command at start of data acquisition, based on whether Surface PAR is enabled or disabled in configuration file (previously, user had to send AddSPAR= to Deck Unit from Seaterm; if setting in Deck Unit was incompatible with setting in configuration file, Seasave indicated that there were scan length errors.).
b. SBE 9plus with 11plus Deck Unit: Status (DS) reply from Deck Unit is now output in header of data file created by Seasave, for Deck Unit with firmware version > 5.0 and operating with RS-232 interface.
c. Playback of archived data: Seasave can now display bottle fire lines when playing back archived data, if it finds a .bl file (of the same name) in the same folder as the data (.hex or .dat) file.

2. MODIFICATIONS
a. NMEA time is now available without RMC when a NMEA device is connected directly to computer. Systems UTC is used as date source.
b. Seasave now interprets cast headers (when playing back archived data) the same way as they are interpreted by SBE Data Processing's Data Conversion module.

3. BUG FIXES
a. SBE 45: If configuration file was set to output salinity but not conductivity, Seasave output 0 for all salinity values.
b. SBE 45: If NMEA time was set up to display in a fixed or scrolled display, and there was no NMEA time in data stream, display NMEA time field showed random numbers (instead of just remaining blank).


VERSION 7.21
06 October 2010

SBE Data Processing --

1. ADDITIONS
a. Add ability to select SBE 37 in Configure menu, to create or modify an .xmlcon file for SBE 37. Add ability to process SBE 37 .hex file data in Data Conversion.
Note: SeatermV2 version 1.1 and later create a .hex data file and an .xmlcon file for SBE 37 when you upload data from SBE 37 memory.

2. MODIFICATIONS
a. Modify ASCII Out's Data Setup tab to allow user selection from a wide range of output date and time formats.
b. Modify Data Conversion's Data Setup tab to allow user selection of the source of the start time to be placed in the output .cnv file header.
c. Modify Sea Plot to allow user selection of date and time format for output plot.

3. BUG FIXES
a. Fix several small bugs.

Seasave V7 --

1. BUG FIXES
a. Fix several small bugs.


VERSION 7.20g
30 July 2010

SBE Data Processing and Seasave V7 --

1. ADDITIONS
a. Add Wet Labs EC-BB and ECO-NTU to OBS/Nephelometer/Turbidity sensor list, and add Turner Cyclops and Seapoint Ultraviolet fluorometer to fluorometer sensor list. Software supports up to 2 of each of these sensors. Previously, the User Polynomial function in the software was used to define these sensors.
b. 16plus V2 and 19plus V2 in moored mode -- add ability to output or display instrument time in Julian days as well as in seconds since January 1, 2000.

Seasave V7 --

1. ADDITIONS
a. 16plus V2 and 19plus V2 in moored mode -- Fixed display and scrolled display now shows instrument time in a day month year hour minute second form, instead of seconds since January 1, 2000.


VERSION 7.20f
28 June 2010

SBE Data Processing and Seasave V7 --

1. MODIFICATIONS
a. Now allows selection of up to two Chelsea UV Aquatracka Fluorometers in the configuration (.xmlcon) file.

2. BUG FIXES
a. Glider Payload CTD: Soc coefficient for integrated frequency oxygen sensor was saved to .xmlcon file with insufficient significant digits.


VERSION 7.20e
11 June 2010

Seasave V7 --

1. BUG FIXES
a. Correct a few minor bugs related to the user interface (no impact on data acquisition).
 

SBE Data Processing --

1. MODIFICATIONS
a. Data Conversion: Remove the line <?xml version="1.0" encoding="UTF-8"?> from the part of the output header that describes the sensors (sensor type and calibration coefficients in XML format). This line was making it difficult for customers to import and process the output .cnv data in Excel. This line has been output by firmware versions 7.19 and higher (software released from July 2009 to now).

2. BUG FIXES
a.
Correct a few minor bugs related to the user interface (no impact on data processing).


VERSION 7.20d
27 May 2010

Seasave V7 --

1. ADDITION
a. Configuration report now includes 'scan length' where applicable, to assist in diagnostics when getting a scan length error (caused by a mismatch between the configuration file and the instrument's setup).

2. BUG FIXES
a. Correct a few minor bugs related to the user interface (no impact on data acquisition).
 

SBE Data Processing --

1. ADDITIONS
a. Add compatibility with Glider Payload CTD (it is now on Configure menu).
b. Configuration report now includes 'scan length' where applicable, to assist in diagnostics when getting a scan length error (caused by a mismatch between the configuration file and the instrument's setup).

2. BUG FIXES
a. Correct a few minor bugs related to the user interface (no impact on data processing).


VERSION 7.20c
02 March 2010

Seasave V7 --

1. BUG FIXES
a. Correct a few minor bugs related to the user interface (no impact on data acquisition).
 

SBE Data Processing --

1. MODIFICATIONS
a. Bottle Summary: Add ability to enable/disable Tau correction for oxygen data, when calculating oxygen as a derived variable in Bottle Summary. Previously, Bottle Summary always applied Tau correction.

2. BUG FIXES
a. Data Conversion: Correct bug that caused software to crash for certain circumstances when processing SBE 19 (not plus) data with data from an AFM.
b. Sea Plot: Correct bug that caused Cursor Position window to show incorrect values for numbers with many digits when viewing plot on your monitor (did not have any effect on saved or printed plot).


VERSION 7.20b
04 February 2010

Seasave V7 --

1. MODIFICATIONS
a. Modify valid limits of temperature and salinity for Gordon and Garcia Oxsol calculation (used as an input for calculating SBE 43 Dissolved Oxygen). Previous limits were -2 < T < 40 and 0 < S < 42; new limits are -5 < T < 50 and 0 < S < 60. Outside of those ranges, software returns a value of -99 for Oxsol.
 

SBE Data Processing --

1. MODIFICATIONS
a. Modify valid limits of temperature and salinity for Gordon and Garcia Oxsol calculation (used as an input for calculating SBE 43 Dissolved Oxygen). Previous limits were -2 < T < 40 and 0 < S < 42; new limits are -5 < T < 50 and 0 < S < 60. Outside of those ranges, software returns a value of -99 for Oxsol.
b. Bottle Summary: If primary temperature is not available, use secondary temperature to calculate SBE 43 Oxygen.
c. Nmea Test: Add options for DPT and DBT strings. Rebuilt using newer Microsoft compiler to eliminate problems with software not running on some PCs.


VERSION 7.20a
14 December 2009

Seasave V7 --

1. ADDITIONS
a. Add ability to save instrument configuration files as xml files (with .xmlcon extension) instead of as .con files. Software will continue to also support .con files.
b. Add ability to plot differences, when 2 sensors are measuring the same parameter; for example, 2 SBE 43 dissolved oxygen sensors.
c. Add ability to easily modify plot ranges by right clicking in the plot and selecting Modify Ranges.
d. Add ability to view position of last bottle fired in fixed or scrolled display.
e. Add Display log file in Tools menu; provides easy access to diagnostic information.

2. MODIFICATIONS
a. Configure Inputs, Configure Outputs, and Plot Display dialog boxes have been reduced in height; all dialogs can now be used with a screen resolution as small as 800 x 600 pixels.
b. Change Plot Display dialog box -- line colors for all axes as well as for bottle fire lines, mark lines, grid lines, etc. are now selected by clicking the Colors button on the Plot Setup tab (previously, each line color was selected independently).
c. Seasave now writes scan interval information to the output data file header for all instruments (previously, wrote this information to the header only for the SBE 9plus and 25).

3. BUG FIXES
a. SBE 25: Correct archived data display of elapsed time, for SBE 25 data uploaded using SEATERM.
b. Correct several small bugs causing software to crash, and provided clearer error messages.

SBE Data Processing --

1. ADDITIONS
a. Add ability to save instrument configuration files as xml files (with .xmlcon extension) instead of as .con files. Software will continue to also support .con files.
b. Add ability to plot differences, when differences are in input .cnv file (when 2 sensors are measuring the same parameter; for example, 2 SBE 43 dissolved oxygen sensors).
c. Sea Calc: Add capabilities to calculate Absolute Salinity as well as Practical Salinity (see Application Note 90: Absolute Salinity and TEOS-10). Note that all other modules only output Practical Salinity (PSU).
d. Data Conversion: Bottle position in water sampler is now available as an output variable, if information is available in input .bl file.
e. Instrument configuration (.xmlcon or .con) files can now be easily viewed by double clicking on the file name. A program that is installed automatically when SBE Data Processing is installed (DisplayConFile.exe) automatically opens configuration files from Windows Explorer; you do not have to open SBE Data Processing or Seasave to view or modify the file.

2. MODIFICATIONS
a. Data Conversion and Derive: Dialog boxes have been reduced in height; all dialogs can now be used with a screen resolution as small as 800 x 600 pixels.
b. Data Conversion: Data Conversion checks the scan interval defined in the input .dat or .hex data file against the scan interval defined in the .xmlcon or .con configuration file. If they are not the same, it gives an error message and does not process the data. Previously, it would process the data, but the results would be incorrect because of the mismatch in scan intervals.

3. BUG FIXES
a. Data Conversion: Previously, if bottles were fired in rapid succession (for example, at one depth), only the last bottle in the sequence had the full number of requested scans in the .ros file output by Data Conversion. Now each bottle has the full number of requested scans.


VERSION 7.19
23 July 2009

Seasave V7 --

1. ADDITIONS
a. Add compatibility with SBE 16plus V2, 16plus-IM V2, and 19plus V2 interfacing with Aanderaa Optode RS-232 sensor.
b. SBE 19 (older instrument): Add ability to append NMEA data from a NMEA device connected to the computer (instead of being connected to the deck unit). Selection of whether NMEA data is coming through the deck unit or the computer is made in the configuration (.con) file. Output data format is not dependent on which method is used for acquiring NMEA data.
c. SBE 49: Add ability to append NMEA data from a NMEA device connected to the computer or to a deck unit. Selection of whether NMEA data is coming through the deck unit or the computer is made in the configuration (.con) file. Output data format is not dependent on which method is used for acquiring NMEA data.
d. Add command line option to automatically start Seasave V7 with a specified .psa file. This command line parameter does not start data acquisition or playback of archived data (both of which have previously existing command line options), it simply launches Seasave V7.

2. MODIFICATIONS
a. Modify the print option for printing in black and white (Print Options tab of Plot Display dialog). Previously this changed the background to white and all plot lines to black. Now it simply changes the background to white, while leaving all plot lines in their specified colors.

3. BUG FIXES
a. If playing back archived data which has S prompts (S>) in the file, Seasave V7 now ignores the S> and does not give scan length errors.
b. Correct several small bugs causing software to crash, and provided clearer error messages.

SBE Data Processing --

1. ADDITIONS
a. Add compatibility with SBE 16plus V2, 16plus-IM V2, and 19plus V2 interfacing with Aanderaa Optode RS-232 sensor.
b. SBE 19 (older instrument, not plus): Add ability to append NMEA data from a NMEA device connected to the computer (instead of being connected to the deck unit). Selection of whether NMEA data is coming through the deck unit or the computer is made in the configuration (.con) file. Output data format is not dependent on which method is used for acquiring NMEA data.
c. SBE 49: Add ability to append NMEA data from a NMEA device connected to the computer or to a deck unit. Selection of whether NMEA data is coming through the deck unit or the computer is made in the configuration (.con) file. Output data format is not dependent on which method is used for acquiring NMEA data.
d. Data Conversion: Data Conversion now adds the calibration coefficients, in XML format, to the header of the output .cnv file.
e. Add Diagnostics in the Options menu of each module.

2. BUG FIXES
a. Data Conversion: If you are using the Merge separate header file capability, all of the header information in the output file will be taken from the separate header. Previously, there was some information from the existing header and some information from the separate header.


VERSION 7.18d
13 May 2009

Seasave V7 --

1. ADDITIONS
a. Command line option /IgnoreNMEAChecksum tells program to ignore checksum when a NMEA device is attached to computer directly (not thru deck unit). Eliminates problems associated with NMEA devices that incorrectly calculate check sum, and allows data to still be acquired by Seasave. Instead of shutting down, a warning appears in SeasaveAcq’s error log.

2. MODIFICATIONS

a. SBE 911plus -- Clean up handling of data with scan length errors.
b. Cleaned up handling of NMEA initialization problems and NMEA time in output .hdr file.
c. SBE 911plus/17plus with one dual sensor (T or C) -- previously allowed you to select output of some secondary derived variables, such as salinity, density, sound velocity, etc. No longer allows you to output secondary derived variables, unless you have a dual T-C pair.
d. Change location of Seasave.ini, to provide better compatibility with computers running Vista. Seasave.ini contains location and file name of most recently used program setup (.psa) files.
e. Change default location of .psa files to provide better compatibility with computers running Vista.
f. Change default location of .txt diagnostic files to provide better compatibility with computers running Vista.
g. Seasave Remote -- Change location of SeasaveRem.ini, to provide better compatibility with computers running Vista. SeasaveRem.ini contains location and file name of most recently used program setup (.psa) files.
h. Seasave Remote -- Change default location of .psa files to provide better compatibility with computers running Vista.
i. Seasave Remote -- Change default location of .txt diagnostic file to provide better compatibility with computers running Vista.

3. BUG FIXES
a. SBE 19plus V2 with SBE 38 RS-232 auxiliary sensor and Surface PAR enabled -- Surface PAR calculation was incorrect.
b. SBE 19plus or 19plus V2 in Moored mode, with Surface PAR enabled -- Surface PAR calculation was incorrect.
c. SBE 911plus/17plus with dual T-C sensors -- Secondary conductivity was calculated using primary temperature, even though there was a secondary temperature sensor.
d. SBE 21 with remote temperature sensor, density calculation – Density is a function of salinity (calculated from SBE 21’s T and C output) and temperature. Seasave incorrectly used temperature from SBE 21 instead of temperature from remote temperature sensor in density calculation.
e. Descent rate and acceleration – If descent rate was output for both metric and English units, both descent rate and acceleration were incorrect (was correct if only 1 was output). If descent rate in English units was output, and acceleration was output, acceleration was incorrect.
f. NMEA -- New position flag for NMEA data always stayed at 0, even if latitude or longitude changed. No impact on NMEA data itself, just the flag (if you chose to display it or output it) was incorrect.
g. Plume anomaly -- Calculation was incorrect if plume anomaly was only derived variable being output for shared file output, tcp/ip output, and/or serial data output.
h. Potential temperature anomaly -- Calculation was incorrect if Sigma-Theta was selected as A1 Multiplier on Miscellaneous tab in Configure Inputs.
i. All instruments except SBE 45 -- Software used IPTS-68 temperature in calculating conductivity; should have been using ITS-90 temperature (T68 = 1.00024 * T90). Bug caused extremely small error in output conductivity, well beyond any rounding error in most cases.

SBE Data Processing --

1. MODIFICATIONS
a. SBE 911plus/17plus with one dual sensor (T or C) -- previously allowed you to select output of some secondary derived variables, such as salinity, density, sound velocity, etc. No longer allows you to output secondary derived variables, unless you have a dual T-C pair.
b. Change location of PostProcSuite.ini, to provide better compatibility with computers running Vista. PostProcSuite.ini contains location and file name of last saved program setup (.psa) file for each module.
c. Change default location of .psa files to provide better compatibility with computers running Vista.

2. BUG FIXES
a. Data Conversion, SBE 911plus/17plus with dual T-C sensors -- Secondary conductivity was calculated using primary temperature, even though there was a secondary temperature sensor.
b. Data Conversion, NMEA -- New position flag for NMEA data always stayed at 0, even if latitude or longitude changed. No impact on NMEA data itself, just the flag (if you chose to output it) was incorrect.
c. Data Conversion, all instruments except SBE 45 -- Software used IPTS-68 temperature in calculating conductivity; should have been using ITS-90 temperature (T68 = 1.00024 * T90). Bug caused extremely small error in output conductivity, well beyond any rounding error in most cases.
d. Data Conversion -- If selected Binary output format, some precision was lost on NMEA time calculation.
e. Data Conversion, SBE 45 -- If selected output temperature in IPTS-68, deg F, software actually output ITS-90, deg C. This applied to data from SBE 45 itself as well as data from a secondary temperature sensor (SBE 38).
f. Data Conversion, SBE 45, specific conductance -- calculation was incorrect.
g. Data Conversion -- Could not process data if output variables included Fluorescence, Wetlab AC3.
h. Data Conversion, Density and Density Difference (density difference if dual T-C sensors) -- software always calculated sigma-theta, regardless of user selection for type of density (density, sigma-theta, sigma-t, sigma-1, sigma-2, sigma-4).
i. Data Conversion, Average Sound Velocity -- Latitude used in calculation was incorrect. Also, if you selected more than one Average Sound Velocity variable for output (available are Chen-Millero, Delgrosso, and Wilson, each in m/s or ft/s), all sound velocity data output corresponded to last one selected.
j. Data Conversion, SBE 911plus/917plus with dual T sensors -- Temperature Difference was incorrect; it always output a constant value.
k. Data Conversion, SBE 21 with NMEA data in input data file header only -- If outputting NMEA data in .cnv file, it output incorrect NMEA data.
l. Data Conversion, SBE 25 -- If selected “frequency2” for output (raw output of pressure sensor), it output incorrect pressure data.
m. Derive, Average Sound Velocity -- If you selected more than one Average Sound Velocity variable for output (available are Chen-Millero, Delgrosso, and Wilson, each in m/s or ft/s), all sound velocity data output corresponded to last one selected.


VERSION 7.18c
31 December 2008

Seasave V7 --

1. ADDITIONS
a. Add capability to create a TS plot (similar to TS plot previously available in SBE Data Processing's Sea Plot module).

2. BUG FIXES
a. Correct several small bugs causing software to crash, and provided clearer error messages.


VERSION 7.18b
04 November 2008

SBE Data Processing --

1. ADDITIONS
a. Add Tau and hysteresis corrections for oxygen data.
b. Add support for a second D&A OBS-3 backscatterance sensor.

2. BUG FIXES
a. Correct several small bugs causing software to crash.

Seasave V7 --

1. ADDITIONS
a. Add Tau and hysteresis corrections for oxygen data.
b. Add GO 1016 Arm offset in Configure Inputs dialog box (Water Sampler tab).
c. Add support for a second D&A OBS-3 backscatterance sensor.

2. BUG FIXES
a. Correct several small bugs causing software to crash.


VERSION 7.18
19 June 2008

SBE Data Processing --

1. ADDITIONS
a. SBE 19plus, 19plus V2, 21, and 25: Add ability to append NMEA data from a NMEA device connected to the computer (instead of being connected to the deck unit). Selection of whether NMEA data is coming through the deck unit or the computer is made in the configuration (.con) file. Output data format is not dependent on which method is used for acquiring NMEA data.

Seasave V7 --

1. ADDITIONS
a. SBE 19plus, 19plus V2, 21, and 25: Add ability to append NMEA data from a NMEA device connected to the computer (instead of being connected to the deck unit). Selection of whether NMEA data is coming through the deck unit or the computer is made in the configuration (.con) file. Output data format is not dependent on which method is used for acquiring NMEA data.
b. Add selection in fixed display dialog to allow short names in display (to take up less room on screen).
c. Add command line option (/nodeckunit), providing smoother operation for SBE 9plus used without SBE 11plus Deck Unit.

2. BUG FIXES
a. Correct several small bugs.


VERSION 7.17a
25 April 2008

SBE Data Processing --

1. ADDITIONS
a. Add support for a second WET Labs ECO AFL fluorometer.

2. MODIFICATIONS
a. Modify Sea-Bird equation (previously called Murphy-Larson equation) for SBE 43 dissolved oxygen.

3. BUG FIXES
a. Correct bug related to dual Turner Scufa sensors and dual Turner OBS sensors.

Seasave V7 --

1. ADDITIONS
a. Add support for a second WET Labs ECO AFL fluorometer.
b. Add user configuration of % for auto paging on plot axis setup tabs.

2. MODIFICATIONS
a. Modify Sea-Bird equation (previously called Murphy-Larson equation) for SBE 43 dissolved oxygen.

3. BUG FIXES
a. Correct bug related to dual Turner Scufa sensors and dual Turner OBS sensors.
b. 19plus: Correct bug in system time calculation if you selected both System time added and NMEA depth data added.
c. Correct several bugs that caused lock-up of the software.


VERSION 7.17
17 March 2008

SBE Data Processing --

1. ADDITIONS
a. SBE 9plus CTD: Add ability to append NMEA data from a NMEA device connected to the computer (instead of being connected to the deck unit). Selection of whether NMEA data is coming through the deck unit or the computer is made in the configuration (.con) file. Output data format is not dependent on which method is used for acquiring NMEA data.
Note: This feature will be added to configuration file setup for other instruments over the next few months.

Seasave V7 --

1. ADDITIONS
a. Add ability to convert an output .xml file (from Shared File Out) to a tab-delimited .txt format that can be opened with Microsoft Excel or other spreadsheet software. See Tools menu -> Convert shared file output .xml file to spreadsheet format.
b. SBE 9plus CTD: Add ability to append NMEA data from a NMEA device connected to the computer (instead of being connected to the deck unit). Selection of whether NMEA data is coming through the deck unit or the computer is made in the configuration (.con) file. Output data format is not dependent on which method is used for acquiring NMEA data.
Note: This feature will be added to configuration file setup for other instruments over the next few months.
c. Add ability to select whether to reset a deck unit (for SBE 16, 16plus, 16plus V2, 19, 19plus, 19plus V2, 21, or 25) to state 1 (echo only, no NMEA data) when data acquisition is complete. Previously, deck unit was automatically reset to state 1.
See Real-Time Data menu -> Start. In the Start Real-Time Data Acquisition dialog box, the new selection is at the bottom (Reset deck unit to state 1 when acquisition stops) if the selected .con file is for one of the applicable instruments and NMEA position data added was selected in the .con file.

2. MODIFICATIONS
a. Update Help file description of how to use -Autostop feature.


VERSION 7.16a
18 January 2008

SBE Data Processing --

1. BUG FIXES
a. ASCII In – If input data file had two temperatures (example: 9plus CTD with dual temperature sensors, or SBE 21 or 45 thermosalinograph with remote temperature sensor), column names for both temperatures were incorrect in output file. Other processing modules (such as Sea Plot) would not run if using this file.

Seasave V7 --

1. BUG FIXES
a. NMEA data display dialog box was too narrow, sometimes resulting in truncation of lat / lon data.
b. Correct error messages received for data acquisition errors.
c. If using –u command line option, output .hex file was created in last directory visited with a File Open dialog. Software now prompts you for where to put output .hex file.
d. IOW Or Hydro-Bios water samplers set up for Auto Fire: after 12 bottles were fired, Auto Fire dialog showed “next bottle’ as #13, but these samplers cannot have more than 12 bottles.

2. ADDITIONS
a. Add audible tone from computer when a bottle fire confirmation is received. Configure and test tone on Water Sampler tab in Configure Inputs.
b. Add computer alarms -- minimum and maximum pressure, altimeter, and bottom contact switch. Configure and test alarms on PC Alarms tab in Configure Outputs.
c. Add mixed mode bottle firing, to allow manual firing of some bottles when set up for auto fire (often used to obtain a water sample at one or more points with the water sampler held in a stationary position, typically at the actual bottom and top of the cast). Configure mixed mode firing in Auto Fire Table Entry dialog box (Configure Inputs / Water Sampler tab; set Firing sequence to Auto Fire and click on Auto-Fire Pressures & Positions button).
d. Add –Autostop=time command line option, to automatically stop real-time data acquisition and shut down Seasave after time. Can be used with a batch file script to create new data files on a defined schedule for continuous acquisition systems, such as a thermosalinograph or a towed vehicle.

3. MODIFICATIONS
a. Temporary (.tmp) files generated by Seasave are now placed in folder designated in your computer for all .tmp files. Previously, .tmp files were placed in same directory as active .psa file.
b. If a .con file was open (for example, in Notepad), you could not start data acquisition.


VERSION 7.16
26 November 2007

SBE Data Processing --

1. BUG FIXES
a. Sea Calc did not open in the previous release.
b. Sea Plot did not allow you to type the path for the input directory; it only was able to open a file if you browsed to the directory using the Select button.

2. ADDITIONS
a. Add compatibility with custom instruments, SBE 16plus V2 and 19plus V2.

3. MODIFICATIONS
a. Data Conversion -- If you select an output variable that requires miscellaneous additional information, a pop-up appears when you select the output variable. User-input value(s) in the pop-up is then automatically placed on the Miscellaneous tab.

Seasave V7 --

1. ADDITIONS
a. Add compatibility with custom instruments, SBE 16plus V2 and 19plus V2.
b. Add Erase all plots feature in Display menu.


VERSION 7.15
24 October 2007

SBE Data Processing --

1. BUG FIXES
a. Data Conversion and Derive -- These modules erroneously calculated oxygen as 0 ml/l if all of the following occurred:
     (1) Using a .con file created and last saved before implementation of the Murphy-Larson equation for SBE 43 Dissolved Oxygen (Seasave V7 and SBE Data Processing version 7.10; Seasave-Win32 version 5.38),
     (2) The .con file includes SBE 43 oxygen and scan time added is selected, and
     (3) Using SBE Data Processing version 7.10 through 7.14e.
b. Sea Plot -- If creating plots for multiple files and selected File All ->JPG, the output .jpg file plots were not scaled correctly.
c. Sea Plot -- If creating TS plots for multiple files and selected File All -> JPG, only the plot that was in the window at that time would include contour lines in the output .jpg file.
d. Sea Plot -- If plotting Julian days or Elapsed time on the X axis, and selected Show as date/time, the dates displayed in the plot were all advanced 1 day from the correct dates.

2. ADDITIONS
a. Sea Plot -- Add Recent Setup Files selection in the File menu, to allow easy selection of the desired .psa file.

3. MODIFICATIONS
a. Batch file operation -- When running through batch file operation, the dialog box in modules that require a .con file now show the actual .con file that is being used. Previously, if Match instrument configuration to input file was selected, SBE Data Processing used the correct .con file in the processing but did not show the actual .con file name as it ran through the batch sequence.

Seasave V7 --

1. ADDITIONS
a. Add Recent Setup Files selection in the File menu, to allow easy selection of the desired .psa file.

2. BUG FIXES
a. In Configure Outputs, on the Header Form tab, Seasave V7 ignored any non-blank lines after blank header lines, and did not save those lines to the .psa file.


VERSION 7.14e
20 September 2007

SBE Data Processing --

1. BUG FIX
a. Align CTD -- For computers set up for some other languages (not English), Align CTD would drop the decimal point or comma in the entered advance number. If this occurred, the incorrect advance value would be shown in the output file header. For example, an entry to advance a variable by 3.25 seconds or 3,25 seconds would be interpreted as 325 seconds; Align CTD would perform the analysis with a 325 second advance, and would show that it used a 325 second advance in the output file header.


VERSION 7.14d
05 September 2007

SBE Data Processing --

1. BUG FIX
a. Data Conversion -- For SBE 17plus, if the input file name (including the path) was longer than 100 characters, the data was not processed.

Seasave V7 --

1. BUG FIX
a. Incorrectly calculated pressure in psi for an instrument with a Druck pressure sensor. Pressure output to display, to serial port, to shared file, and/or via TCP/IP was actually in decibars, even if the label said psi.

2. ADDITIONS
a. Add command line option (/AutoFireOnDowncast) to allow for auto firing capability on downcast (Sea-Bird recommends bottle firing on upcast for almost all applications; this is a custom feature).


VERSION 7.14c
09 July 2007

SBE Data Processing --

1. ADDITION
a. SBE 19plus -- Add support for NMEA depth selection in .con file.

Seasave V7 --

1. ADDITIONS
a. Add XML output formats as an option for Serial Data Output, Shared File Output, and TCP/IP output.
b. Add font size options for Fixed and Scrolled displays.
c. SBE 19plus -- Add support for NMEA depth selection in .con file.
d. Add Remote Bottle Fire (via TCP/IP) capability.
e. Create example Java applications to use for Remote Display (via TCP/IP) and for Remote Bottle Firing (via TCP/IP). These Java applications (.jar files) are installed automatically in the Seasave V7 directory when Seasave V7 is installed. See the Seasave V7 manual and/or Help file for details.

2. MODIFICATIONS
a. Include in the .psa file information on whether the NMEA display, bottle fire display, and/or mark scan control display are open, so that they don't need to be manually opened by the user each time Seasave V7 is launched.
b. Include ability to calculate Oxygen (ml/l) given just oxygen current (and not oxygen temperature) as the input for a Beckman or YSI-type sensor. Seasave sets the WT calibration coefficient to 0 when calculating oxygen. This capability was available in Seasave-Win32, but had been omitted from Seasave V7 until this revision.


VERSION 7.12
09 April 2007

SBE Data Processing --

1. BUG FIX
a. Data Conversion -- For SBE 16plus, did not allow you to select Julian days as a output variable.

2. MODIFICATION
a. Add compatibility with Windows Vista to Help files.

Seasave V7 --

1. MODIFICATION
a. Add compatibility with Windows Vista to Help files.


VERSION 7.11
22 March 2007

Seasave V7 --

1. BUG FIX
a. For SBE 21 Thermosalinograph integrated with a remote temperature sensor (SBE 38 or SBE 3), sound velocity calculation was incorrect. Seasave V7 was using SBE 21 temperature to calculate sound velocity, instead of using temperature from remote temperature sensor.

SBE Data Processing --

No changes; version number was updated to keep it in line with Seasave V7.


VERSION 7.10
19 March 2007

First official release version of Seasave V7. This and all future release of SBE Data Processing will have same revision number as Seasave V7.
Note: Sea-Bird recommends that all users upgrade to Seasave V7, but continues to supply and support Seasave-Win32. See the Seasave-Win32 Revision page for all future revisions of Seasave-Win32.

SBE Data Processing and Seasave V7 --

1. ADDITIONS:
a. Add Import / Export buttons to calibration coefficients dialog boxes, to allow transfer of coefficients from one system to another without the need for typing and possibility of resulting typographical errors.
b. Add Murphy-Larson equation compatibility for SBE 43 Dissolved Oxygen Sensor (Sea-Bird will begin shipping SBE 43s with Murphy-Larson calibration coefficients in Fall 2007).
c. Add compatibility with SBE 55 ECO Water Sampler.

Seasave V7 --

1. BUG FIXES:
a. Numerous small bug fixes.


VERSION 7.0 and 7.0h

Beta Version releases of Seasave V7.


VERSION 5.37m
04 January 2007

SBE Data Processing and Seasave-Win32 --

1. ADDITIONS:
a. Add Scan time added option to configuration (.con) file setup for SBE 16, 19, 19plus, 21, 25, and 49, to append the time to each scan of data (Note: SBE 9plus already had this option; SBE 16plus automatically includes the time with each scan of data).

SBE Data Processing --

1. ADDITIONS:
a. Data Conversion and Derive -- Add Miscellaneous tab in Data Conversion dialog box and Derive dialog box. Miscellaneous tab defines parameters required for output of specific variables (depth, average sound velocity, descent rate, acceleration, oxygen, plume anomaly, and potential temperature anomaly). Entries on this tab are used only if you are calculating and outputting the associated variable. Previously, these values were accessed using the Data button in the Select Output variables dialog box.

2. BUG FIXES:
a. Incorrectly calculated dV/dt (derivative of oxygen voltage with respect to time for SBE 43) if user selected dV/dt as an output variable.
Note: dV/dt was calculated and used correctly within the equation for calculating oxygen (see Application Note 64 for the oxygen equation).


VERSION 5.37e
06 March 2006

SBE Data Processing --

1. BUG FIXES:
a. Derive -- Bug fix described for 5.37d was not actually implemented in 5.37d. Version 5.37e fixes this bug.


VERSION 5.37d
30 Jan 2006

SBE Data Processing --

1. BUG FIXES:
a. Derive -- For salt water depth calculation, Derive now uses latitude from the header if it is available (i.e., if NMEA data was appended in Seasave). Previously it always ignored latitude in the header and just used the latitude entered when depth was selected as a derived variable.

SBE Data Processing and Seasave-Win32 --

1. MODIFICATIONS:
a. The allowable range for conductivity pcor is now -1.0e-06 to -1.0e-10 (previously it was -1.0e-07 to -1.0e-10). SBE Data Processing and Seasave-Win32 continue to check the allowable range, and reset pcor to the nominal value of -9.57e-08 if outside the allowable range.


VERSION 5.37c for SBE Data Processing
22 Dec 2005

SBE Data Processing --

1. BUG FIXES:
a. Wndow Filter -- Window Filter was occasionally crashing when Exclude scans marked bad was checked and the filter type was a median filter.


VERSION 5.37c for Seasave-Win32
22 Nov 2005

NOTE: Version 5.37c is an interim version that applies to Seasave-Win32 only. Continue to use Version 5.37b of SBE Data Processing.

Seasave-Win32 --

1. BUG FIXES:
a. SBE 19plus CTD -- Seasave-Win32 could not acquire data if 19plus was used with an SBE 33 or SBE 36 Deck Unit.


VERSION 5.37b for SBE Data Processing
08 Nov 2005

SBE Data Processing --

1. BUG FIXES:
a. Bin Average -- Bin Average was not excluding scans marked with the bad flag when determining the starting value for the first bin.


VERSION 5.37b for Seasave-Win32
07 Nov 2005

NOTE: Version 5.37b is an interim version that applies to Seasave-Win32 only. Continue to use Version 5.37a of SBE Data Processing.

Seasave-Win32 --

1. BUG FIXES:
a. If elapsed time was selected as a display variable for an SBE 49, Seasave-Win32 did not calculate it correctly.


VERSION 5.37a
26 Oct 2005

SBE Data Processing and Seasave-Win32 --

1. MODIFICATIONS:
a. Surface PAR sensor can now be selected as an auxiliary voltage sensor for SBE 21, instead of PAR sensor.


VERSION 5.37
25 Oct 2005

SBE Data Processing and Seasave-Win32 --

1. ADDITIONS:
a. Add D & A OBS-3+ to list of auxiliary voltage sensors that can be defined.
b. Add support for NMEA depth data to be appended to SBE 21 data string in 90488 Interface Box.
c. PAR sensor can now be selected as an auxiliary voltage sensor for SBE 21.


VERSION 5.36
14 Oct 2005

SBE Data Processing and Seasave-Win32 --

1. BUG:
a. Software did not allow a second Turner SCUFA OBS sensor.

SBE Data Processing --

1. BUG:
a. Sea Plot -- Command line parameter /s to start processing now did not work.


VERSION 5.35
15 July 2005

SBE Data Processing --

1. ADDITIONS:
a. Sea Plot – If a user-defined variable name was input in ASCII In, Sea Plot will now display the user-defined variable name.
b. Data Conversion -- Can process .hex files written for a 911 or 911plus by an in-house development version of Seasave.


VERSION 5.34b
13 June 2005

SBE Data Processing --

1. BUG:
a. 
ASCII Out – Julian days conversion did not work for the SBE 16plus or 19plus.


VERSION 5.34a
18 May 2005

SBE Data Processing --

1. ADDITIONS:
a. ASCII Out – Add command line option to label date column with mon/day/yr and time column with hh:mm (customer request for compatibility with their software).


VERSION 5.34
11 May 2005

SBE Data Processing --

1. BUG FIXES:

a. Fix potential buffer overflow problem.

2. ADDITIONS:
a. Loop Edit -- Add ability to mark scans associated with an initial surface soak, so that surface soak data can be eliminated in further processing.
b. Bin Average -- Add ability to average by depth, with interpolation (previously could average by depth, but without interpolation).
c. ASCII Out -- Add ability to convert elapsed and/or system time to mm/dd/yyyy format (Example: April 11, 2005 would be output as 04/11/2005).

Seasave-Win32 --

No changes were made to Seasave; the revision number was changed to keep it consistent with SBE Data Processing.


VERSION 5.33
08 February 2005

SBE Data Processing --

1. ADDITIONS:
a. ASCII Out -- Add /x parameter for use when running from the Command Line Options dialog box, from the command line, or with batch file processing:
/xascii_out:first_column_value=string
where string = value (maximum of 11 characters) placed in each row of column inserted before first column of data.
b. Help files -- Add more equations defining how derived parameters are calculated.

Seasave-Win32 --

1. BUG FIXES:
a. SBE 31 -- Seasave-Win32 would not allow you to set number of scans to average in the Deck Unit (in Acquire and Display Real-Time Data Setup dialog) to anything except for 1.

2. ADDITIONS:
a. Help files -- Add equations defining how derived parameters are calculated.


VERSION 5.32a
15 October 2004 for SBE Data Processing

NOTE: Version 5.32a is an interim version that applies to SBE Data Processing only. Continue to use Version 5.32 of Seasave.

SBE Data Processing --

1. BUG FIXES:
a. Data Conversion -- Data Conversion did not convert data correctly from an SBE 38 Remote Temperature Sensor used with an SBE 21 Thermosalinograph. This bug affected only versions 5.28 through 5.32.


VERSION 5.32
24 August 2004

SBE Data Processing --

1. BUG FIXES:
a. SBE Data Processing would not convert .psu files created with Version 5.29b or earlier to .psa files.
b. ASCII In -- Negative numbers were output in .cnv file without a leading space in column.
c. Data Conversion and Derive -- If you selected CPAR, program prompted for angle from sun. This information was not needed or used, and prompt has been removed from software.

Seasave-Win32 --

1. BUG FIXES:
a. Header notes with line lengths greater than 95 characters would cause Seasave-Win32 to crash.


VERSION 5.31b
11 June 2004 for SBE Data Processing

NOTE: Version 5.31b is an interim version that applies to SBE Data Processing only. Continue to use Version 5.31a of Seasave.

SBE Data Processing --

1. BUG FIXES:
a. Sea Plot -- When plotting elapsed time (hours, minutes, or seconds), the actual units plotted were always minutes, but the axis label was always seconds, regardless of which units you selected for the plot. This bug affected only versions 5.31and 5.31a.


VERSION 5.31a
18 May 2004 (SBE Data Processing) and 26 May 2004 (Seasave-Win32)

SBE Data Processing --

1. BUG FIXES:
a. All Modules -- If you deleted the directory that held the last used .psu or .psa file for a module, you would get an error message to contact Sea-Bird when next trying to open the module.
b. Sea Plot -- If you ran Sea Plot in batch mode with a w/, Sea Plot did not automatically print the output plot.

Seasave-Win32 --

1. BUG FIXES:
a. When using an SBE 45 MicroTSG and a PN 90402 SBE 45 Power, Navigation, and Remote Temperature Interface Box, and acquiring navigation data: If acquisition was started and stopped, and then restarted without exiting Seasave, the time written into the header line for NMEA UTC (Time) for the second and subsequent files was incorrect. In post-processing of the data, Data Conversion (in SBE Data Processing) would then use this incorrect time with the correct date to compute Julian days. This bug affected Seasave-Win32 versions 5.28a through 5.31.

Note: Raw data files created by Seasave-Win32 that have this problem can be corrected by deleting the data between the first date and last time on the NMEA UTC (Time) line in the header. Once you have edited the raw data file, rerun Data Conversion to correctly compute Julian days.
Example: Start and stop acquisition 4 times without exiting Seasave.
The line below is an example from the header of the fourth file acquired without restarting Seasave:
     * NMEA UTC (Time) = Sep 03 2003  21:06:18 2003  03:17:23 2003  05:24:53 2003  06:34:18
To correct the file, delete the information in bold, so the line looks like this:
     * NMEA UTC (Time) = Sep 03 2003  06:34:18
Caution: See Application Note 67 for information on the correct method for editing a .hex data file with a text editor. If you do not follow these instructions, you may corrupt the data file.


VERSION 5.31
12 May 2004

SBE Data Processing --

1. ADDITIONS:
a. Add c
ommand line utility, ConReport.exe, to software. This program provides same reports generated using Report button on CTD configuration dialogs. ConReport accepts wildcards for file names, so multiple reports may be produced in a single invocation and reports may be placed into a specified directory. ConReport is installed in the same directory as SBE Data Processing during SBE Data Processing installation.

2. MODIFICATIONS:
a.
New program setup files are ASCII text, formatted in XML (extensible markup language). Files may be viewed and edited in any text editor or XML editor. Old binary setup files (.PSU) can be read, but not written; all setup files will be written in the new XML format (.PSA). Old binary setup files will remain, until manually deleted.
b. Sea Plot -- Sea Plot always generated Scan Count data and added it to the variable box, even when Scan Count was already included in the file. This would present 2 selections for Scan Count. Now scan count is generated only when it is not already present in the file.

3. BUG FIXES:
a. Filter – Did not always filter selected variables. The file header always listed the actual variables that were filtered.
b. Sea Plot --
-- Sea Plot now allows processing of duplicated data variables. For example, you may want to convert salinity 3 times in a file, and then apply 3 different filters to salinity data. You need ability to select and differentiate from these modified salinities in plots. A duplicate count suffix is appended to duplicated variables (1st, 2nd, 3rd, etc). This should not be confused with duplicate sensor numbering - for example "Salinity, 2 [PSU], 1st" and "Salinity, 2 [PSU], 2nd" would be duplicated salinity conversions based on secondary temperature and conductivity measurements from a 911 with secondary CT sensors. Prior versions would only plot the first instance of a duplicated variable appearing in a header. This was misleading, as you could select, for example, the third salinity in the variable box, but the first salinity would be plotted with no indication (except perhaps the plot shape) that something was wrong.
-- Sea Plot sometimes crashed when switching between TS and non-TS plots.
-- Values below 1.0 would not plot on a log scale. Very small values, down to 1.0e-8, may now be plotted on a log scale.
-- Some older converted headers had a 2-digit year in the 'start_time' entry, newer headers use a 4-digit year. The routine that parsed the header would fail when reading this 2 digit year (a Y2K problem). Sea Plot can now handle both 2- and 4-digit years.

Seasave-Win32 --

1. MODIFICATIONS:
a. Seasave's algorithms were not changed; this revision incorporates minor changes to the Help files, and keeps the Seasave-Win32 revision number in line with the SBE Data Processing revision number.


VERSION 5.30b
01 March 2004

NOTE: Version 5.30b is an interim version that applies to Seasave-Win32 only. Continue to use Version 5.30a of SBE Data Processing.

Seasave-Win32 --

1. BUG FIX:
a. When using an SBE 45 MicroTSG and a PN 90402 SBE 45 Power, Navigation, and Remote Temperature Interface Box, and acquiring navigation data, the month shown in the NMEA Data display (Seasave's View menu, NMEA data display) was incorrect. Note that the data written to the saved .hex data file was correct.


VERSION 5.30a
30 October 2003 (SBE Data Processing) and 24 November 2003 (Seasave)

SBE Data Processing --

1. BUG FIXES:
a. Window Filter did not work -- this bug applied to version 5.30 only.
b. Some Help file links did not work properly -- this bug applied to version 5.30 only.

Seasave-Win32 --

1. BUG FIXES:
a. Dr. Haardt Fluorometer with voltage level gain range selection did not switch between ranges -- it always used the low gain range.


VERSION 5.30
17 October 2003

Seasave-Win32 and SBE Data Processing --

1. BUG FIXES:
a. SBE 21 with SBE 3 frequency sensor did not work properly.

SBE Data Processing --

1. ADDITIONS:
a. Sea Plot -- Add ability to create TS plots with density or thermosteric anomaly contours. Note that this is an official release version (not a Beta version) of Sea Plot. Because of major changes, the Sea Plot setup file, seaplot.psu, is not backward compatible with previous versions of Sea Plot. Sea Plot will create a new default .psu file the first time version 5.30 is run.

2. MODIFICATIONS:
a. Run menu was re-ordered to list common processing modules first, in the order they are typically used, followed by less common modules.
b. Rosette Summary module was renamed Bottle Summary.


VERSION 5.29b
18 July 2003

Seasave-Win32 and SBE Data Processing --

1. BUG FIXES:
a. SBE 25 .con file -- If user selected bottom contact switch for an A/D channel, software saved selection of bottom contact switch to a different channel.

2. MODIFICATIONS
a. SBE Data Processing (Data Conversion, Rosette Summary, and Derive modules) and Seasave-Win32 -- now use secondary salinity to compute secondary potential temperature if secondary conductivity is available in SBE 9 / 9plus data. Previously software used primary salinity in calculation.

3. ADDITIONS -- 
a. Add Bottom contact switch to list of available A/D sensors for SBE 19plus in .con file.

SBE Data Processing --

1. BUG FIXES:
a. Sea Plot -- did not let users select more than approximately 15 files (only affected V 5.29a). 


VERSION 5.29a
20 June 2003

Seasave-Win32 --

1. MODIFICATIONS:
a. Output latitude and longitude from a NMEA navigation device with up to three digits to right of decimal point.

SBE Data Processing --

1. BUG FIXES:
a. Split -- did not process SBE 19plus data. 

2. ADDITIONS
a. Sea Plot -- Add overlay plot capabilities, so data from multiple files can be viewed on one plot.
b. SBEBatch Batch processing -- 
- Program names and options can be separated by any combination of space and tab characters. Previously, only a space character was accepted.
- Add W option, to pause processing at start of a module to allow user to review setup before processing data.
- Add D option, to pause processing at end of a module to allow user to review output before continuing with rest of processing.
- SBE Batch can now launch system commands, such as copying or renaming a file, deleting a file from an intermediate step, etc. Additionally, it can launch non-Sea-Bird programs.


VERSION 5.28f
09 May 2003

NOTE: Version 5.28f is an interim version that applies to SBE Data Processing only. Continue to use Version 5.28e of Seasave.

SBE Data Processing --

1. BUG FIXES:
a. Data Conversion -- For 911plus, pressure for first scan was computed with a pressure temperature of 0.0. This bug affected only version 5.28 to 5.28e.
b. Sea Plot -- Versions 5.28d and 5.28e were inadvertently built with a development version of Sea Plot.


VERSION 5.28e
08 May 2003

Seasave-Win32 --

1. BUG FIXES:
a. SBE 49 did not work with a PDIM module.
b. Seasave-Win32 was not storing SBE 49 temperature serial number in the header.

SBE Data Processing --

1. BUG FIXES:
a. Data Conversion -- Program would not calculate pressure temperature output for 16plus, 19plus, and 49.


VERSION 5.28d
06 May 2003

Seasave-Win32 and SBE Data Processing --

1. BUG FIXES:
a. Seasave-Win32 and Data Conversion -- SBE 49 temperature coefficients were not stored correctly. Three lines were added to .con file:
    Line     Description
     245       SBE 49 temperature sensor serial number.
     246       SBE 49 temperature sensor calibration date.
     247       SBE 49 temperature sensor A0, A1, A2, A3, slope, and offset
b. Seasave-Win32 and Data Conversion -- SBE 49 conductivity was not computed correctly.

Seasave-Win32 --

1. BUG FIXES:
a. Archived data with header line lengths greater than 100 characters were not processed correctly. Header lines can now have up to 300 characters.


VERSION 5.28c
24 March 2003

Seasave-Win32 and SBE Data Processing --

1. BUG FIXES:
a. When configuring an SBE 25 with one or more A/D voltage sensors, software did not correctly save all configuration information. This bug was present in version 5.27 and later.


VERSION 5.28b
19 March 2003

Seasave-Win32 --

1. MODIFICATIONS:
a. Start initial playback of archived data at a faster rate than it was set for in version 5.28a.
b. Add additional control of playback speed for archived data: No Wait in ArchivedData menu (or type Ctrl N). This plays back data at rate at which it was acquired.

SBE Data Processing --

1. BUG FIXES:
a. Wild Edit: If input .cnv file contained scan count, time, byte count, or pump status before variables that user selected for Wild Edit, then output variables were not ones selected.

2. MODIFICATIONS:
a. A duplicate copy of SBEBatch.exe, our batch file processor, is now placed in Windows system folder. This allows user to run SBEBatch.exe from anywhere without having to specify path. Previously, SBEBatch.exe was only placed in same directory as SBE Data Processing (default location c:/Program Files/Sea-Bird/SBEDataProcessing-Win32).


VERSION 5.28a
04 March 2003

Seasave-Win32 --

1. BUG FIXES:
a. After Seasave-Win32 synchronized .hex or .dat file and the .con file, user was brought to prior .con file when selecting or modifying .con file.
b. If path name had a '.' when selecting an archived data file, Seasave-Win32 replaced everything past  '.' with '.dat'.
c. View Data Header button did not work in Start Archived Data Display dialog box.

2. MODIFICATIONS:
a. Seasave-Win32's .cfg file now remembers whether NMEA Data Display and Fire Bottle Control displays were open.
b. Seasave-Win32 now warns user if they are trying to save to a .con file that is read only (for instance, trying to save to a .con file on a CD-ROM).
c. Seasave-Win32 now turns off SBE 11plus Deck Unit alarm when data acquisition is halted.

3. ADDITIONS
a. Add -ss command line option, allowing user to start Seasave-Win32 and view data, but to delay writing data to a file until user prompts to begin saving data.
b. Add ability for user to control playback speed of archived data (select Faster or Slower in ArchivedData menu).

Seasave-Win32 and SBE Data Processing --

1. BUG FIXES:
a. Batch file processor was not processing /c option correctly.
b. Sea Plot did not work with batch file processor.

2. ADDITIONS
a. Add support for SBE 49 FastCAT.
b. Add support for SBE 45 MicroTSG thermosalinograph and optional 90402 - SBE 45 Interface Box.
c. Add support for SBE 38 secondary temperature sensor for SBE 21 thermosalinograph.
d. Add support for SBE 50 RS-232 sensor to SBE 16plus.
e. Add support for Chelsea UV Aquatracka fluorometer.
f. Add support for secondary Chelsea Aqua 3 fluorometer.

3. MODIFICATIONS
a. If SBE 21 or SBE 45 has a remote temperature sensor, remote temperature is now used to calculate density and sound velocity.
b. Change WetLabs ECO-AFL description to ECO-AFL/FL to cover WetLabs' new instruments.


VERSION 5.27c
20 December 2002 (for SBE Data Processing) and 22 January 2003 (for Seasave)

Seasave-Win32 --

1. BUG FIXES:
a. Seasave-Win32 sometimes crashed when clicking the Select Variable button in a Display Variables Setup dialog box.

SBE Data Processing --

1. BUG FIXES:
a. Sea Plot sometimes crashed after plotting one file, if the user then selected a file that had a different set of variables.


VERSION 5.27b
09 December 2002

SBE Data Processing --

1. BUG FIXES:
a. Loop Edit -- Used depth rather than pressure when both were in the input .cnv file, if depth followed pressure. This caused too many scans to be marked bad if the user had low pass filtered pressure, but not depth, prior to running Loop Edit.
b. Wild Edit -- Crashed if window size was greater than 100.
c. Align CTD, ASCII Out, Filter, Rosette Summary, Strip, Wild Edit, and Window Filter -- On the Data Setup tab, you click a button to select the output variables, specify filter values for variables, etc. In the dialog box that appears, if the number of variables in the header is large enough so that a vertical scroll bar is needed to display all the variables, the scroll bar was not displayed and not all the variables could be seen. This bug only affects SBE Data Processing V5.27 and 5.27a.

2. MODIFICATIONS:
a. Data Conversion -- Now uses NMEA time for Julian days if NMEA UTC (Time) in the header contains both date and time. Previous versions used system time for Julian days. Seasave-Win32 versions prior to 5.27 wrote NMEA date and time into the system time line in the header if NMEA date and time was available, so this modification only affects SBE Data Processing V5.27 and 5.27a.

Seasave-Win32 --

1. BUG FIXES:
a. If the NMEA message contained time but not date, and time in the message had fractional seconds, then NMEA UTC (Time) was incorrect.


VERSION 5.27a
04 November 2002

SBE Data Processing --

1. BUG FIXES:
a. Sea Plot Beta 1.2 version did not work and would not plot. Sea Plot revision was rolled back to version 1.1.

SBE Data Processing and Seasave-Win32 --

1. BUG FIXES:
a. Minor corrections were made to support new sensors added in version 5.27.


VERSION 5.27
30 October 2002 
November 1: A bug was found in this version and it was removed from the FTP site. All additions and modifications in 5.27 are included in 5.27a.

Seasave-Win32 and SBE Data Processing --

1. ADDITIONS:
a. Support for new sensors: Capsum METS methane, Seapoint Rhodamine, Sequoia LISST-25 Suspended Sediment, and WETStar CDOM.
b. Support for new sensor for SBE 16plus only: Pro-Oceanus Gas Tension Device (GTD) for dissolved oxygen and/or nitrogen.
c. Support for secondary sensors for: PAR, WET Labs WETStar, Seapoint Turbidity, and Turner SCUFA.
d. Sound velocity can be calculated from data from secondary T & C sensors.
e. User-input descriptive text for User Polynomial sensors makes sensor identification easier in data file.

2. MODIFICATIONS:
a. Seasave; Data Conversion and Derive in SBE Data Processing: Use harmonic mean to calculate Average Sound Velocity; convert pressure to depth for calculation (with user-input latitude).

Seasave-Win32 --

1. BUG FIXES:
a. Turner 10-AU-005 did not display log scale.

2. ADDITIONS:
a. Command line parameter allows you to automatically start Seasave-Win32 and data acquisition.
b. Keyboard shortcuts allow you to start data acquisition or display of archived data without using a mouse.

3. MODIFICATIONS:
a. System Upload Time in data file header is now always computer time and date, regardless of whether a NMEA navigation device is connected.

SBE Data Processing --

1. BUG FIXES:
a. Derive: Was using depth for pressure when both were in .cnv file and depth followed pressure.
b. Derive and Rosette Summary: Was ignoring match instrument configuration to input file check box selection on File Setup tab.
c. Data Conversion: Both BL file and AFM file appeared in header as source of bottle scan range information.

2. ADDITIONS:
a. Rosette Summary: Scan number is in list of variables that can be averaged.
b. Bin Average: Parameter for run options, command line, or batch file processing allows user to set center value for first bin.

3. MODIFICATIONS:
a. Align CTD: Maximum allowable number of scans to advance is increased to 1000.


VERSION 5.26f
24 September 2002

NOTE: Version 5.26f is an interim version that applies to Seasave-Win32 only. Continue to use Version 5.26 of SBE Data Processing.

Seasave-Win32 --

1. BUG FIXES:
a. Seasave-Win32 was not closing serial ports or ASCII shared files correctly.


VERSION 5.26d
21 August 2002

NOTE: Version 5.26d is an interim version that applies to Seasave-Win32 only. Continue to use Version 5.26 of SBE Data Processing.

Seasave-Win32 --

1. BUG FIXES:
a. When Lat/Lon was not appended to every scan, Seasave-Win32 used a latitude of 0.0 for depth calculation, even if a different value was specified. This bug applied to real-time data acquisition only; playback of archived data was not affected.


VERSION 5.26c
15 August 2002

NOTE: Version 5.26c is an interim version that applies to Seasave-Win32 only. Continue to use Version 5.26 of SBE Data Processing.

Seasave-Win32 --

1. BUG FIXES:
a. The ASCII shared file was not closed when displaying archived data if the Generate Shared File option was enabled in the ASCII Output Set Up dialog box (Configure menu, ASCII Output). This caused the message Error, could not open shared file: filename to appear the next time real-time or archived data was displayed.


VERSION 5.26b
18 July 2002

NOTE: Version 5.26b is an interim version that applies to Seasave-Win32 only. Continue to use Version 5.26 of SBE Data Processing.

Seasave-Win32 --

1. BUG FIXES:
a. NMEA date and time data was not correctly displayed for an SBE 11plus Deck Unit transmitting data to the computer using the RS-232 interface.


VERSION 5.26a
11 July 2002

NOTE: Version 5.26a is an interim version that applies to Seasave-Win32 only. Continue to use Version 5.26 of SBE Data Processing.

Seasave-Win32 --

1. BUG FIXES:
a. An invalid bottle number or communication error message displayed, and bottles were not always fired, for a system with all of the following setup parameters (problem occurred only if all of these setup selections were made):
Configure menu, NMEA (lat/lon) Interface -- Store Lat/Lon in Data File selection is Add to Header Only.
Configure menu, Water Sampler Configuration -- Firing Sequence selection is Table Driven.
Configure menu, Instrument Configuration (old or new) -- NMEA position data added is selected.
RealtimeData menu, Start Acquisition -- Store on Disk is selected.


VERSION 5.26
19 June 2002

Seasave-Win32 --

1. BUG FIXES:
a. For systems with two SBE 43 Dissolved Oxygen Sensors -- In the configuration (.con) file, calibration coefficients for the secondary dissolved oxygen sensor were overwritten by the primary dissolved oxygen sensor coefficients.

SBE Data Processing --

1. BUG FIXES:
a. For systems with two SBE 43 Dissolved Oxygen Sensors -- In the configuration (.con) file, calibration coefficients for the secondary dissolved oxygen sensor were overwritten by the primary dissolved oxygen sensor coefficients.

2. ADDITIONS:
a. Sea Plot Beta Version 1.1
-- Numerous enhancements were made, including: logarithmic plotting; multiple file operation; date/time plotting; calculation and plotting of salinity and density if C, T, P in file; and batch processing. Additional features will be added to Sea Plot over the next few months.
b. SeacalcW
-- Add Windows version of Seacalc to SBE Data Processing. SeacalcW is a seawater calculator that computes a number of derived variables from one user-input scan of temperature, pressure, etc.


VERSION 5.25b
8 April 2002

Seasave-Win32 --

1. BUG FIXES:
a. Seasave-Win32 was not closing serial ports correctly. In Win XP this caused the following message to display: Cannot initialize comport 0. In Win XP it also prevented SEATERM from opening the COM port.
b. SBE 16 and 19 (not 16plus or 19plus) with Paroscientific Digiquartz pressure sensor -- Pressure temperature was not being computed correctly. This bug was introduced in Seasave-Win32 version 1.20.

SBE Data Processing --

1. BUG FIXES:
a. Data Conversion
SBE 16 and 19 (not 16plus or 19plus) with Paroscientific Digiquartz pressure sensor -- Pressure temperature was not being computed correctly. This bug was introduced in SBE Data Processing Beta version 1.3.


VERSION 5.25a
26 March 2002

NOTE: Version 5.25a is an interim version that applies to Seasave-Win32 only. Continue to use Version 5.25 of SBE Data Processing.

Seasave-Win32 --

1. BUG FIXES:
a. SBE 19 (not 19plus) and 25 -- When real-time data acquisition was restarted without exiting Seasave, the CTD would wake up before the message to switch on the CTD with the magnetic switch was displayed. This prevented it from starting logging.


VERSION 5.25
09 January 2002

SBE Data Processing --

1. ADDITIONS:
a. Sea Plot Beta Version
-- Add Sea Plot module, for plotting converted data (C, T, P as well as derived variables and data from auxiliary sensors), to SBE Data Processing. Sea Plot can plot up to 5 variables (1 Y-axis and 4 X-axes or 1 X-axis and 4 Y-axes) on a single plot. Sea Plot allows you to view plots on screen, and then print or save to a file or the clipboard. This release is a Beta version of Sea Plot -- additional features will be added to Sea Plot over the next few months.


VERSION 5.24
14 December 2001

SBE Data Processing --

1. BUG FIXES:
a. Data Conversion 
-- Potential Temperature Anomaly description was incorrect in the output header.
-- When 'Create bottle (.ROS) file only' was selected, the start time was omitted from the output .ROS file.

Seasave-Win32 --

1. BUG FIXES:
a. ASCII Output was limited to 8 variables.
b. If computer is communicating with deck unit using GPIB interface, change logic to determine if NMEA data is available from deck unit.


VERSION 5.23
20 November 2001

SBE Data Processing --

1. BUG FIXES:
a. Rosette Summary
-- The value -1.#IND000 was sometimes displayed in the standard deviation field if all values were the same.
-- If the .ros input file (created in Data Conversion) contained depth, or contained pressure followed by depth, Rosette Summary used depth instead of pressure to compute derived variables. In doing so, it treated the depth as if it were pressure. For example, it interpreted 5 meters as 5 decibars, resulting in small errors in variables derived from pressure. Note that if both pressure and depth were in the file, but the pressure column was after the depth column, the error did not occur.

Seasave-Win32 --

1. BUG FIXES:
a. Seasave-Win32 sometimes crashed if a file path name was longer than 80 characters.
b. Label for Oxygen, SBE, percent saturation incorrectly just said Oxygen, SBE.
c. On some computer monitors, field to view NMEA Date/Time in NMEA Data dialog box (View menu / NMEA data display) was not large enough. Change field from 1 line to 2 lines.

2. ADDITIONS:
a. Add command line argument -nmeatime. For NMEA messages that contain time but not date, -nmeatime directs Seasave-Win32 to use the system date and NMEA time for the System Upload time. Modify time in the header to be:

If NMEA Message Includes -nmeatime Included in Command Line? Information in System Upload Time in Header Information in NMEA UTC Time in Header
date and time (no effect) NMEA date and time NMEA date and time
time only no computer date and time NMEA time
time only yes computer date and NMEA time NMEA time
no date, no time (no effect) computer date and time (none)

SBE Data Processing and Seasave-Win32 --

1. BUG FIXES:
a. Configuration (.con) file 
-- secondary SBE 43 calibration date was being overwritten with primary SBE 43 calibration date.
-- PAR serial number was being written to line for IFREMER nephelometer serial number.


VERSION 5.22
23 August 2001

SBE Data Processing --

1. BUG FIXES:
a. Align CTD -- Line length for list of aligned variables in header was not long enough.
b. Bin Average -- Scans per bin was being added after flag column instead of before.
c. Derive -- Oxygen saturation for SBE 43 was being computed as 0.

2. ADDITIONS:
a. Add CNV37.exe and CNV39.exe to SBE Data Processing. These programs are used to convert SBE 37 MicroCAT and SBE 39 data to formats compatible with SBE Data Processing. Note that CNV37 and CNV39 are not available in SBE Data Processing's menus; the .exe files can be found in the same folder as SBEDataProc.exe.

Seasave-Win32 --

1. BUG FIXES:
a. SBE 19plus -- Comm port configurations was not prompting for baud rate between SBE 32 and CTD and baud rate between SBE 33 and computer.


VERSION 5.21
16 July 2001

SBE Data Processing --

1. BUG FIXES:
a. Incorrectly got inconsistent variables missing from input file message.
b. SBE 16plus and 19plus -- Configuration (.con) file for instruments with NMEA did not have correct number of bytes per scan.

2. MODIFICATIONS:
a. Edit fields on File Setup tab now scroll horizontally.
b. When tabbing through fields on File Setup tab, file and directory change dialogs will not unexpectedly pop up.
c. Input file drop down on File Setup and Header View tabs was made non-editable.

Seasave-Win32 --

1. BUG FIXES:
a. Crashed if filename, including path, was greater than 80 characters.
b. Valid variable was not switched to x-axis #3 if overlay display x-axis #3 was set to none and x-axis #4 was set to a valid variable.


VERSION 5.2
15 June 2001

SBE Data Processing --

1. BUG FIXES:
a. Data Conversion, Rosette Summary, and Derive -- Offset for SBE 43 was being read as 0.0 from .con file.


VERSION 5.1
14 June 2001

SBE Data Processing -- 

BUG FIXES:
a. Data Conversion -- Temperature and pressure were not found when processing SBE 16plus or 19plus data files.

Seasave-Win32 -- 

MODIFICATIONS:
a. Add logic for overlay plots to rearrange x-axis variables if none is selected for one x-axis and a variable to plot is selected for a higher numbered x-axis.


VERSION 5.0
06 June 2001

NOTE: For this and all future software releases, the Windows software real-time data acquisition and analysis software (Seasave and SBE Data Processing) will have the same version number. The 4.0 series and below indicates Seasoft-DOS; the 5.0 series and above indicates Seasoft-Windows.

SBE Data Processing -- 

ADDITIONS:
a. Data Conversion -- Add support for SBE 19plus used with Auto Fire Module (AFM).
b. Data Conversion -- Add command line, run options, and batch file processing /xdatcnv:pump and /xdatcnv:nomatch.
c. Derive -- Add command line, run options, and batch file processing /xderive:pump.
d. ASCII In -- data in input files can now be separated by spaces, commas, or tabs (or any combination).

BUG FIXES:
a. Data Conversion -- If Append to .NAV file was selected in Seasave's Configure/NMEA Interface, and Data Conversion could not find the .nav file, Data Conversion would crash.

Seasave-Win32 -- 

ADDITIONS:
a. Add option to display both the bottle number and pressure to the Bottle Line Configuration.

BUG FIXES:
a. New style .con file changes did not take effect until .con file was looked at a second time or until Seasave-Win32 was restarted.
b. Lat/Lon configuration could be modified by information in header when displaying an archived data file.


OLDER VERSIONS

When the Windows versions of Seasave and SBE Data Processing were incorporated in Seasoft and began having the same revision number:

div.gif (3429 bytes)

Sea-Bird Home     Phone: (+1) 425-643-9866     Fax: (+1) 425-643-9954     E-mail: seabird@seabird.com