Sea-Bird Electronics              Products            Support            Software            Sales            Service            Search/Sitemap

SeatermAF Revision History
Version 2.1.4 latest

div.gif (3429 bytes)

Brief description of SeatermAF  | Download instructions  |  Download from FTP site  |  Software FAQs

The Revision History describes each version of software -- including bug fixes, modifications, and new features. You may not need the latest version; our revisions often include improvements and new features which may have little or no impact on your operation. Review this site regularly to view the latest changes.

Version #  listing:

  2.1.1 Beta   1.20 1.15 1.10
  2.1.0 Beta 1.25 1.19 1.14 1.03
2.1.4 2.0f Beta 1.24 1.18 1.13 1.02
2.1.3 2.0d Beta 1.23 1.17 1.12 1.01
2.1.2 Beta 2.0c Beta 1.21 1.16 1.11 1.00

 


SeatermAF VERSION 2.1.4
17 September 2013

MODIFICATIONS:

  1. Update default location of .psa, .ini, and .txt diagnostics files for compliance with more operating systems.

SeatermAF VERSION 2.1.3
04 February 2013

ADDITIONS:

  1. Add ability to open .ini file from old SeatermAF (< version 2.0) and convert it to .psa setup file for use with SeatermAF V2 (File menu - Open old configuraton file (.ini) from SeatermAF < 2.0).
  2. Add Header Form information to .psa setup file.
  3. Add Configure Header Form button on Upload & Header Options tab in Configuration Options dialog box. This allows you to access Header Form from dialog box as well as from Configure menu (Configure - Header Form).

SeatermAF VERSION 2.1.2 Beta
14 January 2013

ADDITIONS:

  1. Fix 19plus V2 upload timing issue.

SeatermAF VERSION 2.1.1 Beta
05 December 2012

ADDITIONS:

  1. Add feature to prompt user for header information for uploaded XML data file from SBE 25plus.

SeatermAF VERSION 2.1.0 Beta
10 August 2012

MODIFICATIONS:

  1. Rebuilt with new installer.

SeatermAF VERSION 2.0f Beta
07 June 2012

MODIFICATIONS:

  1. Added Connect CTD button on Toolbar (and Connect CTD in Communications menu). This allows customer to view CTD data being transmitted to AFM or ECO after logging is started, to verify proper functioning before system is deployed.

BUG FIXES:

  1. 25plus: Corrected bug related to upload of 25plus data at low baud rates.
  2. Stationary logic: Corrected bug in programming stationary logic's time to hold in pressure window.

SeatermAF VERSION 2.0d Beta
23 May 2012

MODIFICATIONS:

  1. SBE 25plus: If upload fails or is cancelled, SeatermAF V2 sends Escape character to CTD to interrupt.

SeatermAF VERSION 2.0c Beta
22 May 2012

MODIFICATIONS/ADDITIONS:

  1. SeatermAF was completely rewritten to provide better functionality and user interface.
  2. Compatibility with new CTD, SBE 25plus, was added.

SeatermAF VERSION 1.25
13 December 2010

FIXES:

  1. SBE 17plus V2 Searam: SeatermAF did not properly save and program the bottle closure pressure for the first bottle (closure order #1) on the Bottle Closure Pressures/Times tab, if you did not select Bottom bottle closure enabled on the SBE 17plus Bottle Closure Logic tab.

SeatermAF VERSION 1.24
16 May 2008

ADDITIONS:

  1. Added support for SBE 50 Digital Oceanographic Pressure Sensor as interface with SBE 55 ECO Water Sampler.

SeatermAF VERSION 1.23
25 April 2008

FIXES:

  1. AFM: SeatermAF transmitted battery type information to the AFM incorrectly (transmitted code for Ni-cad batteries when they were designated by the user as alkaline, and vice versa).
  2. On the Bottle Closure Pressures/Times tab in the Configuration Options dialog box, SeatermAF sometimes reset the last entry in the table to 0 db.

SeatermAF VERSION 1.21
15 October 2007

FIXES:

  1. On the Bottle Closure Pressures/Times tab in the Configuration Options dialog box, SeatermAF sometimes did not retain the user-selected number of bottles.

SeatermAF VERSION 1.20
09 July 2007

MODIFICATIONS:

  1. Added compatibility with Windows Vista to Help files.

SeatermAF VERSION 1.19
19 March 2007

ADDITIONS:

  1. Compatibility with SBE 55 ECO Water Sampler added.

SeatermAF VERSION 1.18
10 March 2006

FIXES:

  1. SeatermAF did not set the battery type in the AFM when the Program button was clicked.

SeatermAF VERSION 1.17
15 August 2005

FIXES:

  1. If the AFM was set up to fire more than 17 bottles on elapsed time, SeatermAF inserted incorrect times for bottle closures 18 and later, and did not allow you to arm the AFM. This error occurred only in version 1.16.

SeatermAF VERSION 1.16
11 April 2005

FIXES:

  1. AFM requires certain commands (bbx, bbpp, bbtt, and btx)  to be sent in lower case. SeatermAF's Program button (on the toolbar) was sending these commands in upper case. SeatermAF now sends all commands in lower case.

SeatermAF VERSION 1.15
27 September 2004

ADDITIONS:

  1. SBE 17plus: Added ability to select Nickel Metal Hydride (NiMH) batteries for 17plus (17plus is now shipping standard with NiMH batteries and charger).

FIXES:

  1. SBE 50 with AFM: On AFM Bottle Closure Logic tab, SeatermAF sometimes incorrectly activated configuration (.con) file entry field. SBE 50 does not have a .con file. When user subsequently clicked Program on Toolbar or select Program Auto Fire in Utilities menu, SeatermAF programmed all bottle closure pressures as -1 db, regardless of the pressure values input on the Bottle Closure Pressures/Times tab.

SeatermAF VERSION 1.14
19 February 2003

ADDITIONS:

  1. Added support for SBE 50 Digital Oceanographic Pressure Sensor as interface with AFM.

SeatermAF VERSION 1.13
14 March 2002

FIXES:

  1. Fix initialization bug when SBE 25 is selected the first time the program is run. Error caused uploads to fail.

SeatermAF VERSION 1.12
22 October 2001

FIXES:

  1. Fix file handling problems with .ini files.
  2. Fix an execution error that occurred when the user arrowed past the end of the closure parameters and bottle position lists.

SeatermAF VERSION 1.11
11 June 2001

FIXES:

  1. AFM: If Close on elapsed time, program erroneously looked for a .con file and gave an error message.

SeatermAF VERSION 1.10
06 June 2001

ADDITIONS:

  1. AFM: Added support for SBE 19plus CTD.

MODIFICATIONS:

  1. AFM: AFM and applicable CTD now configured in same dialog box; Connect AFM and Connect CTD added to Toolbar and menu.
  2. Configure menu's Configuration Options dialog box tab labels changed to be more descriptive and consistent.
  3. SeatermAF.ini (replaces SeatermAF.cfg) is now located in Windows directory.
  4. Execution priority increased while uploading.

FIXES:

  1. AFM: Closure algorithm for closing bottles on elapsed time or when stationary was not correct.
  2. AFM: Pressure data (used to control bottle closure) was interpreted incorrectly for some instruments with older pressure sensors.

SeatermAF VERSION 1.03
27 October 2000

MODIFICATIONS:

  1. AFM: Reduced lower limits for Pressure to enable upcast and Pressure change to enable upcast from 10 db to 1 db, allowing for work in shallow water.

SeatermAF VERSION 1.02
23 June 2000

FIXES:

  1. AFM: closure pressures was not correctly checked for ascending order if upcast was enabled. 
  2. AFM: Pressure change to enable upcast was not correctly calculated.
  3. AFM: Program locked up if there was a validation error prior to programming the AFM.
  4. SBE 19: When trying to upload data, SBE 19 was not properly detected.

SeatermAF VERSION 1.01
13 June 2000

MODIFICATIONS:

  1. Header form is automatically sized for the number of header prompts. 
  2. When prompt for header is selected and a cast range is uploaded, header form pops up for each cast.
  3. You can minimize SeatermAF, and it will work in background while uploading data.
  4. Upload system time is written in DATCNV-compatible format.
  5. Connect baud rate is stored in .ini file.
  6. Program tries harder to wake up instrument before sending upload command, if it has fallen asleep during header entry.
  7. Menus and buttons are disabled (menus don't pull down, buttons don't click) while SeatermAF is sending a command or receiving data, so one command does not interrupt another.

SeatermAF VERSION 1.00
30 March 2000

Initial release of Win 95/98/NT terminal program for instruments that include Auto Fire capability.

div.gif (3429 bytes)

Sea-Bird Home     Phone: (+1) 425-643-9866     Fax: (+1) 425-643-9954     E-mail: seabird@seabird.com