Sea-Bird Electronics              Products            Support            Software            Sales            Service            Search/Sitemap

FIELD SERVICE BULLETIN NO. 18

PN 90402 SBE 45 Power, Navigation, & Remote Temperature Interface Box Wiring Error
April 2009

div.gif (3429 bytes)

Print version               Link to all Field Service Bulletins

Equipment Affected

PN 90402, an optional Interface Box for the SBE 45, with serial number 0052 through 0091 (exception: 0080 and 0084 are not affected). These Interface Boxes were shipped from March 2007 through March 2009.

 

Description of Problem

For the specified serial numbers, the wiring terminating at the DC In connector is reversed. Consequently, if using DC power for input to the Interface Box, the Interface Box will not function when you turn the Power On/Off switch to the On position.

This wiring error has no impact if you are powering the Interface Box with AC power.

 

Solution

OR

div.gif (3429 bytes)

Sea-Bird Home     Phone: (+1) 425-643-9866     E-mail: seabird@seabird.com