APPLICATION NOTE NO.
11Chelsea
Calculating Calibration Coefficients for Chelsea PAR Light Sensor
with Built-In Log Amplifier
Revised February 2010
This application note applies to the Chelsea PAR Light sensor, which has a built-in log amplifier.
This PAR sensor is compatible with the following Sea-Bird CTDs:
Note: The CTD voltage channel for use with the PAR sensor can be single-ended or differential.
SEASOFT computes PAR using the following:
PAR = [multiplier * 109 * 10 (V - B) / M / Calibration constant] + offset
Enter the following coefficients in the CTD configuration (.con or .xmlcon) file:
M = 1.0 / (log 10 e * A1 * 1000) = 1.0 / (0.43429448 * A1 * 1000) (Note 2)
B = - M * log 10 e * A0 = - A0 / (A1 * 1000) (Note 2)
Calibration constant = 10 9 / 0.046 = 2.174 x 10 10
Multiplier = 1.0 for output units of μEinsteins/m2.sec (Note 3)
Offset = (Note 4)
Notes:
In our SEASOFT V2 suite of programs, edit the CTD configuration (.con
or .xmlcon) file using the Configure
Inputs
menu in Seasave V7 (real-time data
acquisition software)
or the Configure menu in SBE Data Processing
(data processing software).
Select
Par/Irradiance, Biospherical /Licor as the auxiliary voltage sensor; the
algorithm applies to the Chelsea PAR sensor as well.
A0 and A1 are constants from
the Chelsea
calibration sheet with an equation of form:
PAR (in
μWatts/cm2) = A0 + (A1 * mV)
The multiplier can be used to
calculate irradiance in units other than
μEinsteins/m2.sec.
See Application Note 11General for multiplier
values for other units.
The multiplier can also be used to scale the data, to compare the shape of data sets taken at disparate light levels. For example, a multiplier of 10
would make a 10
μEinsteins/m2.sec light level plot as 100
μEinsteins/m2.sec.
Offset: To determine the offset, enter M, B, Calibration constant, and Multiplier, and set Offset to 0.0 in the configuration (.con or .xmlcon) file. In Seasave V7, display the calculated PAR output with the sensor covered (dark); then enter the negative of this reading as the offset in the configuration (.con or .xmlcon) file.
1. Chelsea computes: PAR = K * e (A0 + A1 * 1000 * V) (V = sensor output in volts)
2.
SEASOFT computes: PAR = [multiplier * 109 * 10 (V
- B) / M / Calibration constant] + offset
(V = sensor output in volts)
3. To
determine Calibration constant, let multiplier = 1.0 and offset = 0, and set
equations 1 and 2 equal to each other.
K * e (A0 + A1 * 1000 * V) = 10 9 * 10 (V - B)
/ M / Calibration constant
If e (A0 + A1 * 1000 * V) = 10 (V - B) / M , then K
= 10 9 / Calibration constant
→ Calibration constant = 10 9 / K
where K = 0.046 for PAR units of
μEinsteins/m2.sec
4. If e x = 10
y →
log 10 e x = y and
x log 10 e = y.
Let x = A0 + A 1 * 1000 * V and
y = (V - B) / M
Let W = log 10 e = 0.43429448 →
(A0 + A 1 * 1000 * V) W = (V - B) / M
→ W * A0 + W * A1 * 1000 * V = (V / M) – (B / M)
→ (W * A0) + (W * A1 * 1000 * V) = – (B / M) + (V / M)
Equating like terms:
(W * A1 * 1000 * V) = (V / M) →
M = 1.0 / (W * A1 * 1000)
(W * A0) = – (B / M)
→ B = - M * W * A0 = - A0 / (A1 * 1000)
Sea-Bird Home Phone: (+1) 425-643-9866 Fax: (+1) 425-643-9954 E-mail: seabird@seabird.com