Print version Current Manual (for older manuals, click here) Reference Sheet Configuration options & accessories
|
SUMMARY
DESCRIPTION
The SBE 16plus-IM V2 (Version 2) SEACAT is a high-accuracy conductivity and temperature recorder (pressure is optional) with built-in Inductive Modem. Compared to the previous 16plus-IM, the V2 incorporates an electronics upgrade and additional features, with six differentially amplified A/D input channels, one RS-232 data input channel, and 64 MB FLASH memory. Data can be output in XML as well as ASCII and HEX formats.
The Inductive Modem (IM) system provides reliable, low-cost, real-time data transmission for up to 100 IM-enabled instruments using plastic-coated wire rope (typically 3 x 19 galvanized steel) as both the transmission line and mooring tension member. IM instruments clamp anywhere along the rugged mooring wire. Expensive and potentially unreliable multi-conductor electrical cables with fixed position underwater connectors are not required. IM moorings are easily reconfigured for changing deployments (positions changed or instruments added or removed), by sliding and re-clamping instruments on the cable. IM systems are much less expensive and more power-efficient than acoustic modems, and offer reliable communication over greater distances.
In a typical mooring, an Inductive Modem Module (IMM) or Surface Inductive Modem (SIM) housed in the buoy communicates with underwater IM instruments and is interfaced to a computer or data logger via an RS-232 serial port. The computer or data logger (not supplied by Sea-Bird) is programmed to poll each IM instrument on the mooring for its data, and send the data to a telemetry transmitter (satellite link, cell phone, RF modem, etc.). The 16plus-IM V2 saves data in memory for upload after recovery, providing a data backup if real-time telemetry is interrupted.
The SBE 16plus-IM V2 uses the same temperature and conductivity sensors (and optional strain gauge or Digiquartz® pressure sensors) proven in 10,000 SEACATs and MicroCATs. Sea-Bird’s unique internal-field conductivity cell permits the use of expendable anti-foulant devices, for long-term bio-fouling protection.
Calibration coefficients are stored in memory, permitting data output in ASCII engineering units (degrees C, Siemens/m, decibars, salinity [PSU], sound velocity [m/sec.], etc.). The sample interval, ranging from 10 to 14,400 seconds, is user-programmable in 1-second increments. Between samples, the 16plus-IM V2 powers down, drawing only 140 microAmps. Alkaline D-cells provide power for approximately 290,000 samples of C and T, depending on the sampling and telemetry schedule. Six differentially amplified A/D input channels and one RS-232 channel provide conditioned power (500 mA) for, and obtain data from, optional auxiliary sensors (oxygen, turbidity, fluorescence, etc.).
|
CONFIGURATION, OPTIONS, AND ACCESSORIES
A standard SBE 16plus-IM V2 is supplied with plastic housing for depths to 600 meters, 64 Mbyte FLASH memory, alkaline batteries, glass-reinforced epoxy bulkhead connectors, and expendable anti-foulant devices.
Options and accessories include:
SOFTWARE
The SBE 16plus-IM V2 is supplied with a powerful Windows 2000/XP software package, SEASOFT© V2, which includes programs for communication and data retrieval, and data processing (filtering, aligning, averaging) and plotting of CTD and auxiliary sensor data and derived variables.
SPECIFICATIONS
Dimensions in millimeters (inches)
Measurement Range | Initial Accuracy | Typical Stability | Resolution | |
Conductivity (S/m) | 0 - 9 | 0.0005 |
0.0003 S/m/month |
0.00005 typical |
Temperature (°C) | -5 to +35 | 0.005 |
0.0002 °C/month |
0.0001 |
Pressure (optional) -- Strain-Gauge | 0 to 20 / 100 / 350 / 600 / 1000 / 2000/ 3500 / 7000 meters | 0.1% of full scale range | 0.01% of full scale range/year | 0.002% of full scale range |
Pressure (optional) -- Quartz |
0 to 20 / 60 / 130 / 200 / 270 / 680 / 1400 / 2000 / 4200 / 7000 / 10500 meters | 0.02% of full scale range | 0.02% of full scale range/year | Depends on sample integration time; consult factory |
Memory: 64 Mbyte non-volatile FLASH memory
Data Storage:Recorded Parameter | Bytes/Sample |
T + C | 6 |
Pressure -- strain-gauge or Quartz | 5 |
each external voltage | 2 |
auxiliary RS-232 sensor | sensor dependent |
date and time | 4 |
Real-Time Clock: 32,768 Hz TCXO accurate to ±1 minute/year
Internal Batteries: 9 alkaline D-cells
Battery Endurance 1:
CT only: 290,000 samples
CTD only: 200,000 samples
CTD & 5M pump: 110,000 samples
1
With Duracell MN 1300 cells. Dependent on sampling scheme.
Auxiliary Voltage Sensors:
Auxiliary power out: up to 500 mA at 10.5 - 11 VDC
A/D resolution: 14 bits
Input range: 0 - 5 VDC
Depth Rating, Housing, and Weight:
Depth Rating | Housing Material | Weight in air | Weight in water |
600 m (1950 ft) | acetal copolymer plastic | 9 kg (20 lbs) | 4 kg (9 lbs) |
7000 m (22,900 ft) | 3AL-2.5V titanium | 17 kg (38 lbs) | 12 kg (27 lbs) |
ADDITIONAL INFORMATION / LINKS:
Documentation -- manual, photos, technical papers, application notes, etc.
Sales Information -- options, accessories, cables, mount kits, spares, etc.
Software -- components of Seasoft V2
Links to Other Instruments of Interest
Specifications are subject to change without notice.
Sea-Bird Home Phone: (+1) 425-643-9866 Fax: (+1) 425-643-9954 E-mail: seabird@seabird.com