APPLICATION NOTE NO. 34
Instructions for use of Conductivity Cell
Filling and Storage Device P/N 50087 and 50087.1

Revised March 2008

Print version

This application note provides instructions for use of PN 50087 / 50087.1 syringe and tubing assembly in rinsing, cleaning, and storing conductivity sensors. The tubing assembly consists of a length of 6.35 mm (1/4 inch) I.D. tube connected by a plastic reducing union to a short piece of 11.1 mm (7/16 inch) I.D. tube. Refer to Application Note 2D: Instructions for Care and Cleaning of Conductivity Cells for information on water and solutions recommended for use.

Note: This procedure can also be used with the SBE 16, 16plus, 16plus-IM, 19, and 19plus, which are no longer in production.

Procedure for Use

  1. To fill the conductivity cell, draw about 40-60 cc of purified water into the syringe.
  2. Connect the plastic tubing to the TC duct intake on the temperature sensor [Figure 1] (or to the open end of the conductivity cell on systems without the TC duct [Figure 2]), and inject solution into the cell and pump plumbing.
  1. (If applicable) Loop the rubber band around a bar on the CTD cage and back over the top of the syringe to secure the apparatus for storage.

REMOVE THE SYRINGE AND TUBING ASSEMBLY BEFORE DEPLOYMENT !