Sea-Bird Electronics              Products            Support            Software            Sales            Service            Search/Sitemap

APPLICATION NOTE NO. 73
Using Instruments with Pressure Sensors at Elevations Above Sea Level
Revised October 2010

div.gif (3429 bytes)

Print version

This application note covers use of a Sea-Bird instrument that includes a pressure sensor at elevations above sea level, such as in a mountain lake or stream.

 

Background

Sea-Bird pressure sensors are absolute sensors, so their raw output includes the effect of atmospheric pressure. As shown on the Calibration Sheet that accompanies the instrument, our calibration (and resulting calibration coefficients) is in terms of psia. However, when outputting pressure in engineering units, most of our instruments output pressure relative to the ocean surface (i.e., at the surface the output pressure is 0 decibars). Sea-Bird uses the following equation in our instruments and/or software to convert psia to decibars.

Pressure (db) = [pressure (psia) – 14.7] * 0.689476
where 14.7 psia is the assumed atmospheric pressure (based on atmospheric pressure at sea level)

This conversion is based on the assumption that the instrument is being used in the ocean; the surface of the ocean water is by definition at sea level. However, if the instrument is used in a mountain lake or stream, the assumption of sea level atmospheric pressure (14.7 psia) in the instrument and/or software can lead to incorrect results. Procedures are provided below for measuring the pressure offset from the assumed sea level atmospheric pressure, and entering the offset in the instrument and/or software to make the appropriate correction.

Inclusion of calibration coefficients in the instrument itself or in a file used by our software to interpret raw data varies, depending on the instrument. Commands used to program the instrument vary as well. Therefore, there are variations in the correction procedure, depending on the instrument. These instruments are addressed below:

 

SBE 9plus and 25

Sea-Bird real-time data acquisition software (Seasave) and post-processing software (SBE Data Processing) use calibration coefficients programmed in a configuration (.con or .xmlcon) file to convert raw data from these instruments to engineering units.

Follow this procedure to correct the pressure:

  1. With the instrument in the air, place it in the orientation it will have when deployed.
  2. In Seasave, in the .con or .xmlcon file, set the pressure offset to 0.0.
  3. Acquire data in Seasave, and display the pressure sensor output in decibars.
  4. Calculate offset = (0 – instrument reading).
  5. Enter the calculated offset in the .con or .xmlcon file.

Offset Correction Example:
Pressure displayed at elevation is -1.655 db.
Offset = 0 – (-1.655) = + 1.655 db
Enter offset in .con or .xmlcon file.

 

SBE 16plus and 16plus V2 (RS-232 versions), 19plus and 19plus V2, and 49

Sea-Bird real-time data acquisition software (Seasave) and post-processing software (SBE Data Processing) use calibration coefficients programmed in a configuration (.con or .xmlcon) file to convert raw data from these instruments to engineering units. These instruments are also able to directly output data that is already converted to engineering units (pressure in decibars), using calibration coefficients that are programmed into the instrument.

Follow this procedure to correct the pressure:

  1. With the instrument in the air, place it in the orientation it will have when deployed.
  2. In Seasave, in the .con or .xmlcon file, set the pressure offset to 0.0.
  3. Acquire data in Seasave, and display the pressure sensor output in decibars.
  4. Calculate offset = (0 – instrument reading).
  5. Enter the calculated offset in the .con or .xmlcon file.
  6. Also enter the calculated offset in the instrument (using the POffset= command in the terminal program*).
    * Note: SBE 16plus V2 and 19plus V2 use SeatermV2 terminal program; the other instruments use Seaterm.

Offset Correction Example:
Pressure displayed at elevation is -1.655 db.
Offset = 0 – (-1.655) = + 1.655 db
Enter offset in .con or .xmlcon file and in instrument.

 

SBE 16plus and 16plus V2 (RS-485 versions), and 16plus-IM and 16plus-IM V2

Sea-Bird real-time data acquisition software (Seasave) and post-processing software (SBE Data Processing) use calibration coefficients programmed in a configuration (.con or .xmlcon) file to convert raw data from these instruments to engineering units. These instruments are also able to directly output data that is already converted to engineering units (pressure in decibars), using calibration coefficients that are programmed into the instrument.

Follow this procedure to correct the pressure:

  1. With the instrument in the air, place it in the orientation it will have when deployed.
  2. In the terminal program*, set the pressure offset to 0.0 (#iiPOffset=0) and set the output format to converted data in decimal form (#iiOutputFormat=3).
    * Note: SBE 16plus V2 and 16plus-IM V2 use SeatermV2 terminal program; the other instruments use Seaterm.
  3. Acquire data using the #iiTP command.
  4. Calculate offset = (0 – instrument reading).
  5. Enter the calculated offset in the instrument (using the #iiPOffset= command).
  6. Also enter the calculated offset in the .con or .xmlcon file, using SBE Data Processing.

Offset Correction Example:
Pressure displayed at elevation is -1.655 db.
Offset = 0 – (-1.655) = + 1.655 db
Enter offset in .con or .xmlcon file and in instrument.

 

SBE 37 (all IDO [Integrated Dissolved Oxygen] models, and all other models with firmware version > 3.0)

The SBE 37 is able to directly output data that is already converted to engineering units (pressure in decibars), using calibration coefficients that are programmed into the instrument. If using SeatermV2 (version 1.1 and later) to upload data, SeatermV2 creates a configuration (.xmlcon) file along with a .hex data file. Sea-Bird post-processing software (SBE Data Processing) uses the calibration coefficients in the .xmlcon file to convert raw data to engineering units.

Follow this procedure to correct the pressure:

  1. With the SBE 37 in the air, place it in the orientation it will have when deployed.
  2. In the SeatermV2 terminal program, set the pressure offset to 0.0 and pressure sensor output to decibars. *
  3. Acquire data. *
  4. Calculate offset = (0 – instrument reading).
  5. Enter the calculated offset in the SBE 37 in SeatermV2. *
  6. If you have already uploaded data, also enter the calculated offset in the .xmlcon file, using SBE Data Processing.

Offset Correction Example:
Pressure displayed at elevation is -1.655 db.
Offset = 0 – (-1.655) = + 1.655 db
Enter offset in the SBE 37.

* NOTE: Commands for setting pressure offset, setting output format, and acquiring data vary:

Instrument

Pressure Offset Command

Output Format Command

Command to Acquire Data **

MicroCATs with Inductive Modem (IM) or RS-485 telemetry
(37-IM, 37-IMP, 37-IMP-IDO, 37-SM, 37-SMP, 37-SMP-IDO, 37-SI, 37-SIP, 37-SIP-IDO)
#iiPOffset= #iiOutputFormat=1 #iiTSn:100
(measures and outputs data 100 times)
MicroCATs with RS-232 telemetry
(37-SM, 37-SMP, 37-SMP-IDO, 37-SI, 37-SIP, 37-SIP-IDO)
POffset= OutputFormat=1 TSn:100
(measures and outputs data 100 times)

** See MicroCAT manual for location of pressure data in output data string.

 

SBE 37 (all models with firmware version < 3.0, except IDO [Integrated Dissolved Oxygen] models)

The SBE 37 is able to directly output data that is already converted to engineering units (pressure in decibars), using calibration coefficients that are programmed into the instrument. These SBE 37s do not use a configuration (.con or .xmlcon) file.

Follow this procedure to correct the pressure:

  1. With the SBE 37 in the air, place it in the orientation it will have when deployed.
  2. In the Seaterm terminal program, set the pressure offset to 0.0 and pressure sensor output to decibars. *
  3. Acquire data. *
  4. Calculate offset = (0 – instrument reading).
  5. Enter the calculated offset in the SBE 37 in Seaterm. *

Offset Correction Example:
Pressure displayed at elevation is -1.655 db.
Offset = 0 – (-1.655) = + 1.655 db
Enter offset in the SBE 37.

* NOTE: Commands for setting pressure offset, setting output format, and acquiring data vary:

Instrument

Pressure Offset Command

Output Format Command

Command to Acquire Data

MicroCATs with Inductive Modem (IM) or RS-485 telemetry
(37-IM, 37-IMP, 37-SM, 37-SMP, 37-SI, 37-SIP)
#iiPOffset= #iiFormat=1 #iiTP (measures and outputs pressure 30 times)
MicroCATs with RS-232 telemetry
(37-SM, 37-SMP, 37-SI, and 37-SIP)
POffset= Format=1 TP (measures and outputs pressure 100 times)
       

SBE 50

The SBE 50 is able to directly output data that is already converted to engineering units (psia, decibars, or depth in feet or meters), using calibration coefficients that are programmed into the instrument. The SBE 50 does not use a configuration (.con or .xmlcon) file.

Follow this procedure to correct the pressure:

  1. With the SBE 50 in the air, place it in the orientation it will have when deployed.
  2. In the Seaterm terminal program, set the pressure offset to 0.0 (POffset=0) and set the output format to the desired format (OutputFormat).
  3. Acquire data using the TS command a number of times.
  4. Calculate offset = (0 – instrument reading).
  5. Enter the calculated offset in the SBE 50 (use POffset= in Seaterm). The offset must be entered in units consistent with OutputFormat=. For example, if the output format is decibars (OutputFormat=2), enter the offset in decibars.

Offset Correction Example:
Pressure displayed at elevation with OutputFormat=2 (db) is -1.655 db.
Offset = 0 – (-1.655) = + 1.655 db
Enter offset in the SBE 50.

 

SBE 52-MP

The SBE 52-MP is able to directly output data that is already converted to engineering units (decibars), using calibration coefficients that are programmed into the instrument. The SBE 52-MP does not use a configuration (.con or .xmlcon) file.

Follow this procedure to correct the pressure:

  1. With the SBE 52-MP in the air, place it in the orientation it will have when deployed.
  2. In the Seaterm terminal program, set the pressure offset to 0.0 (POffset=0).
  3. Acquire data using the TP command.
  4. Calculate offset = (0 – instrument reading).
  5. Enter the calculated offset in the SBE 52-MP (use POffset= in Seaterm).

Offset Correction Example:
Pressure displayed at elevation is -1.655 db.
Offset = 0 – (-1.655) = + 1.655 db
Enter offset in the SBE 52-MP.

 

SBE 39-IM

The SBE 39-IM directly outputs data that is already converted to engineering units (pressure in decibars), using calibration coefficients that are programmed into the SBE 39-IM. The SBE 39-IM does not use a configuration (.con or .xmlcon) file.

Follow this procedure to correct the pressure:

  1. With the SBE 39-IM in the air, place it in the orientation it will have when deployed.
  2. In the Seaterm terminal program, set the pressure offset to 0.0 (#iiPOffset=0).
  3. Acquire data using the #iiTP command.
  4. Calculate offset = (0 – instrument reading).
  5. Enter the calculated offset in the SBE 39-IM (use #iiPOffset= in Seaterm).

Offset Correction Example:
Pressure displayed at elevation is -1.655 db.
Offset = 0 – (-1.655) = + 1.655 db
Enter offset in the SBE 39-IM.

 

SBE 39

The SBE 39 directly outputs data that is already converted to engineering units (pressure in decibars), using calibration coefficients that are programmed into the SBE 39. The SBE 39 does not use a configuration (.con or .xmlcon) file. The SBE 39 is a special case, because its programmed calibration coefficients do not currently include a pressure offset term. The lack of a pressure offset term creates two difficulties when deploying at elevations above sea level:

Range
(m or db) +
Range (psi) =
Range (db) / 0.689476
0.1 * Range (psi) Atmospheric Pressure (psi) at
elevation at which error occurs =
[14.7 – 0.1 * Range (psi)]
Approximate Corresponding Elevation (m)
20 29 2.9 11.8 psi 1800
100 145 14.5 0.2 psi No where on Earth!
350 507 50.7
1000 1450 145
2000 2900 290
3500 5076 507
7000 10152 1015

+ Notes:

  • Although decibars and meters are not strictly equal, this approximation is close enough for this Application Note. See Application Note 69 for conversion of pressure (db) to depth (m) for fresh or salt water applications.
  • Equations used in conversions ‑
    As shown above:    pressure (db) = [pressure (psia) – 14.7] * 0.689476;
    Rearranging:           pressure (psia) = [Pressure (db) / 0.689476] + 14.7
    Measuring relative to atmospheric:   pressure (psi; relative to atmospheric pressure) = Pressure (db) / 0.689476

From the table, it is apparent that the only practical limitation occurs with a 20 meter pressure sensor. To use the SBE 39 in this situation, change the sensor range internally to 100 meters by entering PRange=100 in the SBE 39 (using Seaterm). This changes the electronics' operating range, allowing you to record pressure data at high elevations, but slightly decreases resolution. After the data is recorded and uploaded, perform post-processing to adjust for the pressure offset. Note that Sea-Bird software cannot currently perform this adjustment for the SBE 39.

CAUTION: Changing PRange in the SBE 39 does not increase the actual maximum water depth at which the instrument can be used (20 meters) without damaging the sensor.

Example 1: You want to deploy the SBE 39 with a 20 m pressure sensor in a mountain lake at 1400 meters (4590 feet). This is lower than 1800 meters shown in the table, so you do not need to adjust the sensor range. After the data is recorded and uploaded, perform post-processing to adjust for the pressure offset.

Example 2: You want to deploy the SBE 39 with a 20 m pressure sensor in a mountain lake at 2000 meters (6560 feet). This is higher than 1800 meters shown in the table, so you need to adjust the sensor range. In Seaterm, set PRange=100 to allow use of the SBE 39 at this elevation. After the data is recorded and uploaded, perform post-processing to adjust for the pressure offset.

 

SBE 26plus and 53 BPR

Unlike our other instruments that include a pressure sensor, the SBE 26plus and 53 output absolute pressure (i.e., at the surface the output pressure is atmospheric pressure at the deployment elevation). Therefore, no corrections are required when using these instruments above sea level. SBE 26plus / 53 software (SEASOFT for Waves) includes a module that can subtract measured barometric pressures from tide data, and convert the resulting pressures to water depths.

 


Application Note Revision History
Date Description
June 2004 Initial release.
July 2005 Add information on SBE 39-IM, 52-MP, 53.
February 2009 - Add information on V2 Seacats and newer MicroCATs (version 3 and greater firmware).
- Add information on SeatermV2 software.
October 2010 - Add information on IDO MicroCATs. (37-SMP-IDO, SIP-IDO, IMP-IDO)
- Add information on creation of  .xmlcon file for MicroCATs with data uploaded using SeatermV2 1.1 and later.
- Add information on .xmlcon files for all instruments that use configuration files.
- Update address.

div.gif (3429 bytes)

Last modified: 23 Feb 2011

Sea-Bird Home     Phone: (+1) 425-643-9866     Fax: (+1) 425-643-9954     E-mail: seabird@seabird.com