Sea-Bird Electronics              Products            Support            Software            Sales            Service            Search/Sitemap

SEASOFT-DOS (Obsolete -- For Reference Only)

div.gif (3429 bytes)

Revision History  |  Manual  |  Software download instructions  |  Go to FTP site  |  Software FAQs

SEASOFT©-DOS consists of modular, menu-driven routines for acquisition, display, processing, and archiving of oceanographic data acquired with Sea-Bird equipment. A DOS program designed to run on an IBM-compatible computer (XT/AT/386/486/Pentium), SEASOFT-DOS usually performs correctly when run under Windows.

Sea-Bird recommends that all users upgrade to the Windows version of SEASOFT, unless there is a compelling reason to continue running DOS. SEASOFT-DOS was last updated in early 2001. Processing capability added to our software after that time has been added only to the Windows versions (for example, processing data from new Sea-Bird instruments -- such as SBE 16plus, 16plus-IM, 16plus V2, 16plus-IM V2, 19plus, 19plus V2, 45, 49, 50 -- or interfacing to additional auxiliary sensors). While Sea-Bird continues to support customers using the DOS software by answering questions, we will not change, improve, or correct the DOS software.

SEASOFT-DOS supports the following Sea-Bird products:

To enhance data integrity, raw data is acquired from the instruments and stored as raw, unmodified data. A conversion program (DATCNV) uses the instrument configuration and calibration coefficients to create a converted engineering unit data file that is operated on by all SEASOFT-DOS post-processing modules. Each SEASOFT-DOS module that modified the converted data file adds information to the header of the converted file, permitting tracking of how the various oceanographic parameters were obtained.

DESCRIPTION OF SEASOFT-DOS MODULES

Category Module Name Module Description
Terminal Programs 
to communicate 
with instruments
TERM11 For SBE 11 deck unit, firmware version 2.0 or greater.
TERM1621 For SBE 16 SEACAT (not 16plus, 16plus V2, 16plus-IM, or 16plus-IM V2) and SBE 21 SEACAT Thermosalinograph.
TERM17 For SBE 17 SEARAM.
TERM19 For SBE 19 SEACAT Profiler (not 19plus or 19plus V2).
TERM25 For SBE 25 SEALOGGER CTD.
TERM37 For SBE 37 MicroCAT (not MicroCATs with firmware > 3.0).
TERMAFM For Carousel Auto Fire Module.
TMODEM For SBE 11plus modem channel.
Acquisition, 
Playback, and 
Conversion
CNV37 Convert SBE 37 MicroCAT data from .asc files, and store converted data in .cnv files (not MicroCATs with firmware > 3.0).
DATCNV Convert raw data (.HEX or .DAT files) to engineering units (ASCII or binary format), and store converted data in .CNV files.
ROSSUM Summarize data in rosette (water sampler bottle) .ros  files.
SEACON Define instrument configuration and calibration coefficients.
Seasave Acquire real-time data and display raw archived data.
Data Editing ASCIIIN Add header information to a file containing rows and columns of ASCII data.
MARKSCAN Create bottle scan range (.bsr) file from .mrk data file.
SECTION Extract rows of data from file.
SPLIT Split data into upcast and downcast files.
STRIP Extract columns of data from file.
TRANS Convert data format from ASCII to binary, or vice versa.
Data Processing 
on converted data from .cnv file
ALIGNCTD Align C, T, P, and oxygen.
BUOYANCY Compute buoyancy (Brunt-Väisälä) and stability frequency.
BINAVG Average data, based on pressure, depth, scan number, or time ranges.
CELLTM Perform conductivity thermal mass correction.
DERIVE Calculate derived variables (salinity, density, sound velocity, oxygen, potential temperature, dynamic height, etc.) from data.
FILTER Low pass filter columns of data.
LOOPEDIT Mark scan if scan fails pressure reversal or minimum velocity tests.
WILDEDIT Mark wild points in data.
WFILTER Filter data with triangle, cosine, boxcar, gaussian, or median windows.
Data Display 
and Plotting
ASCIIOUT Output data portion and/or header portion from .cnv file to an ASCII file.
CONTOUR Generate density contours to overlay on TS plots.
SEAPLOT Plot data (C, T, P as well as derived variables). Plots can be screen dumped to a printer or plotted on an HP pen plotter or HP Laser Jet III.
Calibration OXSAT * Compute oxygen saturation as a function of temperature and salinity.
OXFIT * Compute oxygen coefficients.
OXFITW * Compute oxygen coefficients using Winkler titration values.
PHFIT * Compute pH coefficients.
Miscellaneous CFGTOCON * Convert SEASOFT Version 3.* configuration (SEASOFT.CFG) files to SEASOFT V 4.0 format.
SEACALC Calculate derived variables.
     

Note: All modules, except those marked with an asterisk (*), are available as Windows programs (part of the SEASOFT V2 suite) as well as in SEASOFT-DOS:

If you discover reproducible bugs in SEASOFT, please document the steps taken to achieve them and send a report via e-mail to jbackes@seabird.com or Fax 425-643-9954 to John Backes. We welcome any suggestions for new features and enhancements to the Windows version of the software.

div.gif (3429 bytes)

Last updated: 23 Feb 2011