Sea-Bird Electronics            Products            Support            Software            Sales            Service            Search/Sitemap

MicroCAT C-T (P optional) Recorder
SBE 37-SMP
(S
erial interface, Memory, & integral Pump)

[HRule Image]

Print version     Current manual - RS-232 or RS-485 (for older manuals, click here)   Reference Sheet - RS-232 or RS-485     Configuration options & accessories

Intended for deployment in orientation shown (connector at bottom) for proper operation

 Optional plastic (ShallowCAT) housing shown; standard titanium housing also available

SUMMARY

DESCRIPTION

The SBE 37-SMP MicroCAT is a high-accuracy conductivity and temperature (pressure optional) recorder with Serial interface, internal battery, non-volatile FLASH Memory, and integral Pump. The MicroCAT is designed for moorings or other long duration, fixed-site deployments. Constructed of titanium and other non-corroding materials to ensure long life with minimum maintenance, the MicroCAT's depth capability is 7000 meters (23,000 feet); it is also available with an optional 250-meter (820-feet) plastic ShallowCAT housing.

Calibration coefficients are stored in EEPROM, allowing the MicroCAT to output data in ASCII engineering units (decimal or XML format); raw output is also available. The data always includes Conductivity, Temperature, Pressure (if optional pressure sensor is installed), and time. If desired, the MicroCAT can calculate and output salinity and sound velocity (Chen-Millero).

SENSORS AND INTERFACE ELECTRONICS

The MicroCAT retains the temperature and conductivity sensors used in our time-proven SEACAT and SEACAT plus products. Electrical isolation of the conductivity electronics eliminates any possibility of ground-loop noise. The MicroCAT’s unique internal-field conductivity cell permits the use of expendable anti-foulant devices. The aged and pressure-protected thermistor has a long history of exceptional accuracy and stability.

The optional strain-gauge pressure sensor is available in eight ranges, from 0 - 20 meters to 0 - 7000 meters. Compensation of the temperature influence on pressure offset and scale is performed by the MicroCAT’s CPU.

Temperature is acquired by applying an AC excitation to a hermetically sealed VISHAY reference resistor and an ultra-stable aged thermistor (drift rate typically less than 0.002 °C per year). The thermistor resistance to reference resistance ratio is determined by a 24-bit A/D converter, which also processes the pressure sensor signal. Conductivity is acquired with an ultra-precision Wien-Bridge oscillator.

PUMP

The integral pump runs for 1.0 second each time the MicroCAT samples, providing the following advantages:

COMMUNICATIONS AND INTERFACING

The MicroCAT communicates directly with a computer via standard RS-232 interface. Data can be uploaded at up to 115.2K baud. Real-time data can be transmitted at distances of up to 1600 meters (5200 feet) at 600 baud, simultaneous with recording. Firmware upgrades can be downloaded through the communications port, without opening the instrument. An optional RS-485 interface allows multiple MicroCATs to share a common 2-wire cable, minimizing cable complexity for C-T chains.

User-selectable operating modes include:

SOFTWARE

The MicroCAT is supplied with a powerful Windows 2000/XP software package, SEASOFT© V2, which includes:

DATA STORAGE, FORMAT, AND BATTERY ENDURANCE

Temperature and conductivity are stored 6 bytes/sample, time 4 bytes/sample, and optional pressure 5 bytes/sample; memory capacity is in excess of 530,000 samples. The MicroCAT is powered by a 7.8 Amp-hour (nominal) battery pack consisting of twelve AA lithium batteries (Saft LS14500) which, when removed from the MicroCAT, can be shipped via commercial aircraft. The pack provides sufficient internal battery capacity for more than 380,000 samples for a typical sampling scheme. *

SPECIFICATIONS

  Measurement
Range
Initial
Accuracy
Typical
Stability
Resolution
Conductivity 0 to 7 S/m
(0 to 70 mS/cm)
0.0003 S/m 
(0.003 mS/cm)
0.0003 S/m
(0.003 mS/cm)
per month
0.00001 S/m
(0.0001 mS/cm)
Temperature
 (°C)
-5 to 35 0.002 0.0002 per month 0.0001
Optional Pressure 20 / 100 / 350 / 600 / 1000 / 2000 / 3500 / 7000 m
(meters of deployment depth capability)
0.1% of full scale range 0.05% of full scale range per year 0.002% of full scale range
         

Clock Stability: 5 seconds/month

Quiescent Current: 30 microAmps *

Sampling and Communication Current: *
Communication 4.3 milliAmps
Sampling (excluding pump)
     9.1 milliAmps if transmitting real-time data;
     7.9 milliAmps if not transmitting real-time data

* Power consumption / battery endurance values are for standard RS-232 interface; for optional RS-485 interface, see RS-485 manual.

Pump Current: 0.025 Amp-seconds/sample

Acquisition Time: 1.9 - 2.9 seconds/sample, dependent on sampling mode and inclusion of pressure sensor

Power Supply: 7.8 Amp-hour (nominal) battery pack

Optional External Power: 0.25 Amps at 9 - 24 VDC

Housing, Depth Rating, and Weight:

  Housing Material Depth Rating Weight
(with standard mounting clamp and guide)
Standard Titanium 7000 m (23,000 feet)  
Optional (ShallowCAT) Plastic 250 m (820 ft) 3.4 kg (7.5 lbs) in air
1.6 kg (3.5 lbs) in water
       

Dimensions in millimeters (inches)

Connector:

ADDITIONAL INFORMATION / LINKS:

Documentation -- manual, photos, technical papers, application notes, etc.

Sales Information -- options, accessories, cables, mount kits, spares, etc.

Software -- components of Seasoft V2

Links to Other Instruments of Interest

[HRule Image]

Specifications are subject to change without notice.

Last modified: 21-Apr-2011

Sea-Bird Home     Phone: (+1) 425-643-9866     Fax: (+1) 425-643-9954     E-mail: seabird@seabird.com