APPLICATION NOTE NO. 62
Calculating Calibration Coefficients for WET Labs ECO-AFL and ECO-FL Fluorometer, ECO-NTU Turbidity Meter, and ECO-FL-NTU Fluorometer/Turbidity Meter (voltage or RS-232 output sensor)
Revised
May 2011

div.gif (3429 bytes)

Print version
(Note: This application note is for use with Seasave and SBE Data Processing version 7.21d and later. An earlier version of the application note is compatible with earlier software versions.)

This Application Note applies to the following WET Labs fluorometers and turbidity meters:

Note: The ECO-FL is also available as part of the WET Labs Triplet, an RS-232 output instrument that incorporates three WET Labs ECO sensors. The Triplet can include up to three ECO-FL or ECO-NTU sensors (or one of those sensors in combination with other ECO sensors). Set up each RS-232 output channel with the coefficients from the calibration sheet for the sensor on that channel.

 

Fluorometer Calibration Coefficients

The fluorometer has a response that is linear over the measurement range provided.

Analog Scaling Value / Range (enter in ECOView Host) 1 2 4 (factory default)
Chlorophyll a: Nominal Range (µg/l) 0 - 30 0 - 50 0 - 125
Rhodamine: Nominal Range (ppb) 0 - 55 0 - 110 0 - 230
Phycocyanin: Nominal Range (ppb) 0 - 100 0 - 200 0 - 400
Phycoerythrin: Nominal Range (ppb) 0 - 55 0 - 110 0 - 230
CDOM: Nominal Range (ppb) 0 - 125 0 - 250 0 - 500
Uranine: Nominal Range (ppb) 0 - 100 0 - 200 0 - 400

Note: Measurement ranges specified by WET Labs have varied over the years; consult the documentation that was provided with your sensor for the appropriate ranges, or contact WET Labs.

Chlorophyll a: Nominal Range (µg/l) 0 - 50
Rhodamine: Nominal Range (ppb) 0 - 175
Phycocyanin: Nominal Range (ppb) 0 - 175
Phycoerythrin: Nominal Range (ppb) 0 -175
CDOM: Nominal Range (ppb) 0 - 375
Uranine: Nominal Range (ppb) 0 - 300

Setting Up Configuration (.xmlcon) File for Fluorometer

  1. Use the Configure Inputs menu in Seasave V7 (real-time data acquisition software), or the Configure menu in SBE Data Processing (post-processing software), to create / modify the .xmlcon file (see software Help files).
  1. The software prompts for Dark Output and Scale Factor, and calculates:

concentration (µg/l or ppb, as applicable) = (Output - Dark Output) * Scale Factor     (see Note)

where:

The fluorometer comes with a calibration sheet that lists values for Dark Output and Scale Factor (see Note). Each of these values is supplied in terms of both voltage and counts. If you changed the analog scaling value (ECO-FL series only), change the Scale Factor to correspond.

Notes:

Example Chlorophyll a Concentration Calculation in Sea-Bird Software:

Dark Output = 0.05 volts and Scale Factor = 12.35 µg/l-volts (from calibration sheet)

Measured voltage from fluorometer = Output = 4.65 volts

Calculated concentration (µg/l) = (Output - Dark Output) * Scale Factor
                                             = (4.65 - 0.05) * 12.35 = 56.8 µg/l

While the factory-supplied Scale Factor can be used to obtain approximate values, field calibration is highly recommended.

Example Calculation of Scale Factor from field calibration:

Seawater sample analysis shows chlorophyll a concentration is 50 µg/l when fluorometer reads 3.2 volts; measured signal for seawater blank is 0.05 volts.

concentration (µg/l) = (Output - Dark Output) * Scale Factor
50 = (3.2 - 0.05) * Scale Factor

Solving: Scale Factor = (50) / (3.2 - 0.05) = 15.87 µg/l

Enter new Scale Factor in configuration (.xmlcon) file.

 

Turbidity Meter Calibration Coefficients

The turbidity meter has a response that is linear over the measurement range provided.

Analog Scaling Value / Range (enter in ECOView Host) 1 2 4 (factory default)
Sea-Bird PN 24345, 24346, and 24348: Nominal Range (NTU) 0 - 30 0 - 60 0 - 125
Sea-Bird PN 24367: Nominal Range (NTU) 0 - 250 0 - 500 0 - 1000

Note: If you require m -1 sr -1 output, WET Labs can provide a secondary calibration for these instruments in m -1 sr -1 at an additional cost. Alternatively, you can purchase the ECO-BB, which comes calibrated to m -1 sr -1. See Application Note 87 for details on how to use m -1 sr -1 output with Sea-Bird CTDs and software.

Setting Up Configuration (.xmlcon) File for Turbidity Meter

  1. Use the Configure Inputs menu in Seasave V7 (real-time data acquisition software), or the Configure menu in SBE Data Processing (post-processing software), to create / modify the .xmlcon file (see software Help files).
  1. The software prompts for Dark Output and Scale Factor, and calculates:

Turbidity (NTU) = (Output - Dark Output) * Scale Factor     (see Note)

where:

The turbidity meter comes with a calibration sheet that lists values for Dark Output and Scale Factor. Each of these values is supplied in terms of both voltage and counts. If you changed the analog scaling value (single-channel ECO-NTU only), change the Scale Factor to correspond.

Note: The configuration file can only be saved as an .xmlcon file (not a .con file) if the digital output version of the ECO-NTU is one of the sensors.

While the factory-supplied Scale Factor can be used to obtain approximate values, field calibration is highly recommended. The relationship between turbidity and NTU is highly variable, and is not easy to determine in the laboratory. Particle shape and size are some of the factors that affect the relationship. To accurately measure NTU with a turbidity meter, perform calibrations on seawater samples with distributions of particles that are similar to what is expected in-situ. Determine NTU independently, and use those values, as well as readings from the turbidity meter, to determine the correct Scale Factor. The Scale Factor is correct as long as the distribution of particle sizes and shapes does not change; the condition does change with season and geographic location.

 


Application Note Revision History
Date Description
March 2001 Initial release.
March 2003 Reflect addition of ECo-FL(RT), -FLD, -FL, -FLS, -FLB, and -FLSB to Wet Labs product line and to Seasave and SBE Data Processing .con file.
June 2003 Add information on Wet Labs ECO-FL-NTU.
December 2006 Add note that newer calibration sheets from Wet Labs may list Dark Counts instead of Vblank. You can use the Dark Count value as the entry for Vblank when setting up the .con file.
May 2007 Incorporate Seasave V7.
August 2009 - Update fluorometer and turbidity meter ranges to correspond to current listing.
- Add Rhodamine fluorometer.
- Update to Seasoft V2.
February 2010 - ECO-NTU can come with secondary calibration to 0 - 5 m -1 sr -1, if customer wants both units from same sensor.
- Add information on .xmlcon configuration file.
- Update Sea-bird address.
April 2010 Add Phycocyanin sensor.
May 2010 Add Phycoerythrin sensor.
August 2010 SBE Data Processing and Seasave 7.20g software revision: WET Labs ECO-NTU added to list of OBS/Nephelometers/Turbidity sensors (previously needed to select User Polynomial for this sensor).
February 2011 Add CDOM sensor.
May 2011 SBE Data Processing and Seasave 7.21d software revision: WET Labs RS-232 output sensors, including WET Labs Triplet, are compatible with RS-232 sensor channel on SBE 16plus V2, 16plus-IM V2, and 19plus V2 CTDs.