APPLICATION NOTE NO. 87
Calculating Calibration Coefficients for
WET Labs
ECO-BB Scattering/Turbidity Meter
(voltage or RS-232 output sensor)
Revised
May 2011
Print version
(Note: This application note is for use with Seasave and SBE Data Processing
version 7.21d and later.
An earlier
version of the application note is compatible with earlier
software versions.
This Application Note applies to the use of the WET Labs ECO-BB series of scattering/turbidity meters with Sea-Bird CTDs. The ECO-BB has a response that is linear over the measurement range provided.
The type of output available from the ECO-BB has varied:
- ECO-BB with 0 - 5 m -1 sr -1 range: WET Labs provides a calibration sheet for calculation of output in m -1 sr -1 .
- ECO-BB with 0 - 25 NTU range: WET Labs provided two calibration sheets, for calculation of output in NTU or m -1 sr -1, as desired by the user.
- ECO-BB with 0 - 100 NTU range: WET Labs provided a calibration sheet for output in NTU only.
Note: If you require NTU output in a currently manufactured product, WET Labs ECO-NTU or ECO-FLNTU are calibrated for calculation of output in NTU. See Application Note 62.
Each calibration sheet provides a Dark Outputs and Scale Factor consistent with the output units. Definition of the sensor in Sea-Bird software follows for m -1 sr -1 and NTU (NTU output available only for older ECO-BB meters).
Notes:
- Use the voltage values in the Sea-Bird software if using a voltage output ECO-BB.
- Use the counts values in the Sea-Bird software if using an RS-232 output ECO-BB (compatible only with SBE 16plus V2, 16plus-IM V2, or 19plus V2 CTDs).
- For the voltage (analog) output ECO BB: Select OBS/Nephelometers/Turbidity – Turbidity Meter, WET Labs, ECO-BB for one of the external voltage channels on the CTD.
- For the RS-232 (digital) output ECO BB (SBE 16plus V2, 16plus-IM V2, or 19plus V2 CTDs only): Select Turbidity Meters – Turbidity Meter, WET Labs, ECO-BB for one of the RS-232 channels on the CTD.
b(qc) = (Output – Dark Output) * Scale Factor (b(qc) in units of m -1 sr -1)
where:
- Output (volts for analog output sensor; counts for digital output sensor) = in-situ output of ECO-BB
- Dark Output (volts for analog output sensor; counts for digital output sensor) = measured output for a seawater blank (pure, de-ionized water) with black tape over detector
- Scale factor (b(qc)/volts for analog output sensor; b(qc)/counts for digital output sensor) = multiplier
Note:
The configuration file can only be saved as an .xmlcon file (not a .con file) if the ECO-BB is selected as one of the sensors.
To get NTU units, select the ECO-NTU instead of the ECO-BB in the Sea-Bird software.
turbidity (NTU) = (Output – Dark Output) * Scale Factor
where:
- Output (volts) = in-situ output of the turbidity meter
- Dark Output (volts) = measured output for a seawater blank (pure, de-ionized water) with black tape over detector
- Scale factor (NTU/volts) = multiplier
Note:
The configuration file can only be saved as an .xmlcon file (not a .con file) if the ECO-NTU is one of the sensors.
While the factory-supplied Scale Factor can be used to obtain approximate values, field calibration is highly recommended. The relationship between ECO-BB output and β(Θc) or turbidity is somewhat variable, and is not easy to determine in the laboratory. Particle shape and size are some of the factors that affect the relationship. To accurately measure with the ECO-BB, perform calibrations on seawater samples with distributions of particles that are similar to what is expected in-situ. Determine β(Θc) or turbidity independently, and use those values, as well as readings from the ECO-BB, to determine the correct Scale Factor. The Scale Factor is correct as long as the distribution of particle sizes and shapes does not change; the condition does change with season and geographic location.
Date | Description |
May 2007 | Initial release. |
March 2010 | - WET Labs changes: Until recently,
there was only ECO-BB, for both NTU & m-1
sr-1. Now they have
different products for different units: - ECO-BB now only for m-1 sr-1 (calibrated 0-to-5 m-1 sr-1). - ECO-NTU for NTU; ECO-NTU can have secondary calibration to m-1 sr-1 for extra charge. - Add information on .xmlcon configuration file. - Update Sea-bird address. |
August 2010 | SBE Data Processing and Seasave
7.20g software revision:
- Wet Labs ECO-BB added to list of OBS/Nephelometers/Turbidity sensors (previously needed to select User Polynomial for this sensor). - Also, for older ECO-BB meters outputting in NTU: Wet Labs ECO-NTU added to list of OBS/Nephelometers/Turbidity sensors (previously needed to select User Polynomial for this sensor). |
May 2011 | SBE Data Processing and Seasave 7.21d software revision: WET Labs RS-232 output sensors, including WET Labs Triplet, are compatible with RS-232 sensor channel on SBE 16plus V2, 16plus-IM V2, and 19plus V2 CTDs. |