Brochure/specifications
Manual
Reference
Sheet
Note: We do not publish prices on the website. Please contact us [seabird@seabird.com
or (+1) 425-643-9866] for pricing.
Part # | Description | Notes | ||
16plus V2 | SeaCATplus Version 2 Conductivity & Temperature (pressure optional) Recorder - Includes 64 MB memory, RS-232 interface, 6 differential A/D channels (0 - 5 volt input range), 1 RS-232 data input channel, AF24173 Anti-Foulant Devices, data I/O extension cable (17762), data I/O cable (801225), Seasoft software, & complete documentation. Specify: housing (depth) selection. | SBE 16plus
V2 is a CTD intended for moored applications.
Photos
of standard shipment. See:
|
||
SBE 16plus V2 Housing (depth) Selections -- MUST SELECT ONE | ||||
16p-1a | 600 m plastic housing |
16plus V2 in plastic & titanium housings |
||
16p-1b | 7000 m titanium housing | |||
16p-1c | 10,500 m titanium housing | |||
SBE 16plus V2 Strain Gauge Pressure Sensor Options | ||||
16p-2a | 20 m strain gauge pressure sensor |
Pressure sensor is installed in connector end cap, & is not field replaceable / swappable.
While highest pressure rating gives you most flexibility in using 16plus
V2, it is at expense of
accuracy & resolution. It is advantageous to use lowest range
pressure sensor compatible with your intended maximum operating
depth, because accuracy & resolution are proportional to pressure
sensor's full scale range. For example, comparing 2000 & 7000 m sensors --
|
||
16p-2b | 100 m strain gauge pressure sensor | |||
16p-2c | 350 m strain gauge pressure sensor | |||
16p-2s | 600 m strain gauge pressure sensor | |||
16p-2d | 1000 m strain gauge pressure sensor | |||
16p-2e | 2000 m strain gauge pressure sensor | |||
16p-2f | 3500 m strain gauge pressure sensor | |||
16p-2g | 7000 m strain gauge pressure sensor | |||
SBE 16plus V2 Digiquartz Pressure Sensor Options (substitutes for strain gauge option) | ||||
16p-2h | 45 psia (20 m) Digiquartz pressure sensor with temperature compensation |
Digiquartz pressure sensors provide better accuracy
(0.02% of full scale range vs 0.1% of full scale range) &
resolution (Digiquartz resolution dependent on user-programmable
integration time) than strain gauge pressure sensors. See SBE 16plus
V2 Strain Gauge Pressure Sensor Options
above for photo of pressure sensor port; when used with Digiquartz,
nylon pressure capillary fitting screws into port. Pressure sensor is installed in connector end cap, & is not field replaceable / swappable. While highest pressure rating gives you most flexibility in using 16plus V2, it is at expense of accuracy & resolution. It is advantageous to use lowest range pressure sensor compatible with your intended maximum operating depth, because accuracy & resolution are proportional to pressure sensor's full scale range. For example, comparing 2000 psia (1400 m) & 6000 psia (4200 m) sensors --
|
||
16p-2i | 100 psia (60 m) Digiquartz pressure sensor with temperature compensation | |||
16p-2j | 200 psia (130 m) Digiquartz pressure sensor with temperature compensation | |||
16p-2k | 300 psia (200 m) Digiquartz pressure sensor with temperature compensation | |||
16p-2l | 400 psia (270 m) Digiquartz pressure sensor with temperature compensation | |||
16p-2m | 1000 psia (680 m) Digiquartz pressure sensor with temperature compensation | |||
16p-2n | 2000 psia (1,400 m) Digiquartz pressure sensor with temperature compensation | |||
16p-2o | 3000 psia (2,000 m) Digiquartz pressure sensor with temperature compensation | |||
16p-2p | 6000 psia (4,200 m) Digiquartz pressure sensor with temperature compensation | |||
16p-2q | 10,000 psia (6,800 m) Digiquartz pressure sensor with temperature compensation | |||
16p-2r | 15,000 psia (10,500 m) Digiquartz pressure sensor with temperature compensation | |||
SBE 16plus V2 Submersible Pump Options | ||||
16p-4e | Add SBE 5M plastic pump for sensor flushing, Standard XSG/RMG connectors (includes data I/O & pump Y-cable 17797, & mount) |
SBE
5M pump is intended for providing
pumped conductivity only. Pump provides better conductivity response & anti-foulant
protection than non-pumped configuration. SBE 5M pump is rated to 600 m
with plastic housing or 10,500 m with titanium housing. Larger, more powerful SBE 5P or 5T is required if also planning to pump auxiliary sensors (such as dissolved oxygen, etc.). SBE 5P plastic pump is rated to 600 m; SBE 5T titanium pump is rated to 10,500 m. Operational characteristics of SBE 5P & 5T are identical. Data I/O - Pump bulkhead connector on 16plus V2 is standard, so pump can be added in field without requiring any modifications to 16plus V2. These all include pump, pump mount, & data I/O - pump Y-cable. 16p-4c, -4d, -4g, & -4h include Y-cable compatible with wet-pluggable bulkhead connectors on 16plus V2 & on pump. See DN 31551 for standard connectors, DN 32896 for wet-pluggable connectors. |
||
16p-4a | Add SBE 5M titanium pump for sensor flushing, Standard XSG/RMG connectors (includes data I/O & pump Y-cable 17797, & mount) | |||
16p-4f | Add SBE 5P plastic pump for sensor flushing, Standard XSG/RMG connectors (includes data I/O & pump Y-cable 17797, & mount) | |||
16p-4b | Add SBE 5T titanium pump for sensor flushing, Standard XSG/RMG connectors (includes data I/O & pump Y-cable 17797, & mount) | |||
16p-4g | Add SBE 5M plastic pump with Wet-pluggable MCBH connectors for sensor flushing (includes data I/O & pump Y-cable 172220, & mount. Requires Option 16p-5) | |||
16p-4c | Add SBE 5M titanium pump with Wet-pluggable MCBH connectors for sensor flushing (includes data I/O & pump Y-cable 172220, & mount. Requires Option 16p-5) | |||
16p-4h | Add SBE 5P plastic pump with Wet-pluggable MCBH connectors for sensor flushing (includes data I/O & pump Y-cable 172220, & mount. Requires Option 16p-5) | |||
16p-4d | Add SBE 5T titanium pump with Wet-pluggable MCBH connectors for sensor flushing (includes data I/O & pump Y-cable 172220, & mount. Requires Option 16p-5) | |||
SBE 16plus V2 Wet-Pluggable Connector Option | ||||
16p-5 | Wet-pluggable (MCBH) connectors instead of Standard (XSG) connectors on SeaCAT bulkhead connectors & on extender & data I/O cables |
Wet-pluggable connectors may be mated in wet
conditions. Their pins do not need to be dried before mating. By
design, water on connector pins is forced out as connector
is mated. However, they must not be mated or un-mated while
submerged. Wet-pluggable connectors have a non-conducting guide pin
to assist pin alignment (exception: 6-pin connectors) & require less force to mate, making them
easier to mate reliably under dark or cold conditions, compared to
our standard (Impulse XSG/RMG) connectors. Like standard connectors,
wet-pluggables need proper lubrication & require care during use
to avoid trapping water in sockets. |
||
SBE 16plus V2 Dissolved Oxygen Sensor Options | ||||
16p-7g | SBE 63 Optical Dissolved Oxygen Sensor, 600 m (cable & mount included, requires option 16p-4f or -4b) |
SBE 63 is available in 7000 m titanium
(16p-7e or -7f) or 600 m (16p-7g or -7h) housing, and its connector type
(standard or wet-pluggable) must match 16plus V2 connector type. SBE 63 is plumbed in line between conductivity sensor & pump; consequently, system requires more powerful SBE 5P plastic pump (16p-4f or -4h) or SBE 5T titanium pump (16p-4b or -4d) instead of SBE 5M pump (16p-4a, -4c, -4e, or -4g). Operational characteristics of SBE 5P & 5T are identical. 16p-7e through h includes SBE 63, mount, & straight cable to connect sensor to 16plus V2 bulkhead connector. If connecting 2 sensors to 1 bulkhead connector, extra charge to substitute Y-cable applies in addition to 16p-7_ price. Additional sensors not listed here -- fluorometers, transmissometers, turbidity meters, PAR sensors, etc. from third party manufacturers -- are compatible with 16plus V2. These sensors can be purchased from Sea-Bird & integrated with 16plus V2 at our factory, or you can purchase mount kits & cables from Sea-Bird & perform integration yourself. See Third Party portion of price list. |
||
16p-7h | SBE 63 Optical Dissolved Oxygen Sensor with Wet-pluggable connector, 600 m (cable & mount included, requires option 16p-4h or -4d & 16p-5) | |||
16p-7e | SBE 63 Optical Dissolved Oxygen Sensor, 7000 m (cable & mount included, requires option 16p-4b) | |||
16p-7f | SBE 63 Optical Dissolved Oxygen Sensor with Wet-pluggable connector, 7000 m (cable & mount included, requires option 16p-4d & 16p-5) | |||
SBE 16plus V2 Spares & Accessories | ||||
801542 | AF24173 Anti-Foulant Device pair (spare, bagged, labeled for shipping) |
Anti-foulant
devices fit into anti-foulant device cups at each end of
conductivity cell. Anti-foulant devices included with standard
shipment; these are spares. Useful life varies, depending on several factors. We recommend that customers consider more frequent replacement when high biological activity & strong current flow (greater dilution of anti-foulant concentration) are present. Moored instruments in high growth & strong dilution environments have been known to obtain a few months of quality data, while drifters that operate in non-photic, less turbid deep ocean environments may achieve years of quality data. Experience may be strongest determining factor in specific deployment environments. |
||
233540 | Anti-foulant cap without hose barb, black |
Cap without hose barb is used for:
|
||
233538 | Anti-Foulant cap with hose barb, black |
Barbed cap is used for:
|
||
80291 | 9-cell series alkaline battery pack (solder connected) |
Batteries are easily accessed by unscrewing battery
end cap. 16plus V2 comes standard with 9 D cell alkaline
batteries (nominal 14 Ah), but can be powered with Electrochem
lithium batteries for more power intensive / longer deployments. Notes:
|
||
801483 | Battery pack assembly kit for 80547 9 D 10.8v/42Ah Lithium pack with instructions, cells not included (DN 67143) | |||
801479 | Battery pack assembly kit for 80540 3 DD 10.8v/30Ah Lithium pack with instructions, cells not included (DN 67144) | |||
50434 | SBE 16plusV2/19plusV2 Spares Kit (Standard connectors) - Complete support kit containing spare data I/O cable, bulkhead connectors, dummy plugs, Triton X-100, o-ring lubricant, & other mechanical spares & maintenance items. | Order appropriate spares kit for connector type on 16plus V2: | ||
50435 | SBE 16plusV2/19plusV2 Spares Kit (Wet-pluggable connectors) - Complete support kit containing spare data I/O cable, bulkhead connectors, dummy plugs, Triton X-100, o-ring lubricant, & other mechanical spares & maintenance items. | |||
801225 | Data I/O Cable, Standard connector (RMG-4FS) with DB-9S, 2.4 m (DN 32421) |
These test cables are used for setting up system &
uploading data from memory after recovery. Data I/O cable connects
to one of following cables:
|
||
801374 | Data I/O cable, Wet-pluggable (MCIL-4FS) with DB-9S, 2.4 m (DN 32715) | |||
17762 | Data I/O extension cable, AG-206 to RMG-4MP, 0.8 m (DN 31605) | This cable connects 6-pin data I/O pump bulkhead connector to 801225 data I/O cable, for 16plus V2 used without a pump. | ||
177797 | Y-cable, Pump-Data I/O, Standard connectors (DN 31551) | These Y-cables connect 6-pin data I/O - pump bulkhead connector to pump (2-pin leg) & to data I/O cable (4-pin leg). | ||
172220 | Y-cable, Pump-Data I/O, Wet-pluggable (DN 32896) | |||
90087 | Universal plumbing kit (includes pump air release valve, Y-fitting, & tubing) -- Application note 64-1 | Application Note 64-1 details installation of plumbing for SBE 43 & pump on a CTD. | ||
50087.1 | Cell filler/storage device with hose barbs (Application Note 34) | For cleaning conductivity cell after each use & storing instrument between uses. See document 67043 & Application Note 2D: Instructions for Care and Cleaning of Conductivity Cells. | ||
20200 | USB to Serial Port Adapter, FTDI UC232R-10 (connects computers with USB ports to RS-232 instruments) | Many newer PCs & laptop computers have USB port(s) instead of RS-232 serial port(s). USB serial adapter plugs into USB port, & allows a serial device to be connected through adapter. Multi-port adapters are available from other companies; see Application Note 68. | ||
various | Plumbing | For assorted sizes of Tygon tubing, see SBE 5M Configuration Details. |
Product configuration options listing for all products
Additional cables, mount kits, and spares may be found at: Cables, Mount kits for Sea-Bird sensors, Mount kits for Third Party sensors, Spare parts
Sea-Bird Home Phone: (+1) 425-643-9866 Fax: (+1) 425-643-9954 E-mail: seabird@seabird.com