Sea-Bird Electronics            Products            Support            Software            Sales            Service            Search/Sitemap

MicroCAT C-T-DO (P optional) Recorder
SBE 37-SIP-IDO
(S
erial Interface, integral Pump, & Integrated Dissolved Oxygen)

[HRule Image]

Print version     Current manual - RS-232 or RS-485 (not yet available)    Reference Sheet - RS-232 or RS-485 (not yet available)     Configuration options & accessories

Intended for deployment in orientation shown (sensor end up) for proper operation.

Standard titanium housing with optional clamps shown; optional plastic (ShallowCAT) housing also available.

SUMMARY

DESCRIPTION

The SBE 37-SIP-IDO MicroCAT is a high-accuracy conductivity and temperature (pressure optional) sensor with Serial Interface, integral Pump, and Integrated Dissolved Oxygen sensor. It includes an internal memory, but is externally powered. Constructed of titanium and other non-corroding materials for long life with minimal maintenance, the MicroCAT is useful as a stand-alone monitoring device, and is easily integrated with current meters, ROVs, AUVs, towed sonars, and other instrumentation platforms..

Calibration coefficients are stored in EEPROM, allowing output of C, T, P, DO, and time in ASCII engineering units (decimal or XML; raw output available); time, salinity, sound velocity (Chen-Millero), and depth can also be output.

SENSORS

Temperature and Conductivity sensors are based on our field-proven SEACAT and SEACATplus products. Electrical isolation of conductivity electronics eliminates any possibility of ground-loop noise. Our unique internal-field conductivity cell permits the use of expendable anti-foulant devices, for long-term bio-fouling protection. The aged and pressure-protected thermistor has a long history of exceptional accuracy and stability.

The IDO is a frequency-output version of our field-proven SBE 43 Dissolved Oxygen sensor, with the same performance specifications.

The optional strain-gauge pressure sensor is available in eight ranges, from 0 - 20 meters to 0 - 7000 meters. Compensation of the temperature influence on pressure is performed by the MicroCAT’s CPU.

PUMP

The integral pump runs each time the MicroCAT samples, providing the following advantages:

With Adaptive Pump Control, the MicroCAT calculates the pumping time for best oxygen accuracy as a function of the previous sample's temperature and pressure (maximizing data quality while minimizing power consumption).

COMMUNICATIONS AND INTERFACING

The MicroCAT communicates via standard RS-232 interface. Real-time data can be transmitted up to 1600 meters at 600 baud (power considerations may limit distance), simultaneous with recording; data can be uploaded at up to 115.2K baud. The user can upgrade firmware through the external connector, without opening the housing. An optional RS-485 interface allows multiple MicroCATs to share a common 4-wire cable (power, common, data+, data-), minimizing cable complexity for C-T chains.

User-selectable operating modes include:

SOFTWARE

The MicroCAT is supplied with a powerful Windows 2000/XP software package, SEASOFT© V2, which includes:

SPECIFICATIONS

  Measurement
Range
Initial
Accuracy
Typical
Stability
Resolution
Conductivity 0 to 7 S/m
(0 to 70 mS/cm)
0.0003 S/m 
(0.003 mS/cm)
0.0003 S/m
(0.003 mS/cm)
per month
0.00001 S/m
(0.0001 mS/cm)
Temperature
 (°C)
-5 to 35 0.002 0.0002 per month 0.0001
Oxygen 120% of surface saturation
(all natural waters, fresh and salt)
2% of saturation 0.5% per 1000 hours 0.035% of saturation
Optional Pressure 20 / 100 / 350 / 600 / 1000 / 2000 / 3500 / 7000 m
(meters of deployment depth capability)
0.1% of full scale range 0.05% of full scale range per year 0.002% of full scale range
         

Clock Stability: 5 seconds/month

Memory: 8 Mbyte; capacity in excess of 444,000 samples with pressure

Acquisition Time: 2.2 - 5.0 seconds/sample, dependent on sampling mode and inclusion of pressure sensor

Input Power: 0.25 Amps at 9 - 24 VDC

Power Consumption (all with pressure)*:

   Quiescent: 0.0004 Watts

   CTD-DO Sample Acquisition (excluding pump): 0.17 Watts

   CTD-DO Sample Waiting (not sampling, pump running, excluding pump):
     
Receive line valid 0.056 Watts
      Receive line not valid 0.016 Watts

   CTD-DO Between Samples:
     
Receive line valid 0.056 Watts
      Receive line not valid 0.0004 Watts

   Pump: 0.12 Watts

   Communications: 0.065 Watts

* Power consumption values are for standard RS-232 interface; for optional RS-485 interface, see RS-485 manual.

Housing, Depth Rating, Weight:

  Housing Material Depth Rating Weight
Standard Titanium 7000 m (23,000 feet) 3.6 kg in air, 2.3 kg in water
Optional (ShallowCAT) Plastic 250 m (820 ft)  
       

Dimensions

Connector:

ADDITIONAL INFORMATION / LINKS:

Documentation -- manual, photos, technical papers, application notes, etc.

Sales Information -- options, accessories, cables, mount kits, spares, etc.

Software -- components of Seasoft V2

Links to Other Instruments of Interest

[HRule Image]

Specifications are subject to change without notice.

Last modified: 08-Apr-2011

Sea-Bird Home     Phone: (+1) 425-643-9866     Fax: (+1) 425-643-9954     E-mail: seabird@seabird.com