Sea-Bird Electronics            Products            Support            Software            Sales            Service            Search/Sitemap

Inductive Cable Coupler (ICC)
Inductive Modem Power Requirements

[HRule Image]

The Inductive Cable Coupler (ICC) is available as a 20-turn coupler (Version 4a), 100-turn coupler (Version 4b), and 8-turn coupler (older OEM applications). Transmit power levels depend primarily on the transmitting device, the number of turns in the coupler, and the size of the gap in the coupler. Transmit power levels do not change significantly with the number of IM devices in the loop or the loop impedance, although very low IM loop impedance may increase transmit power levels; always test IM systems with at least 100 ohms in the IM loop.

The table below summarizes power requirements:

  20-turn ICC
(Version 4a)
100-turn ICC
(Version 4b)
8-turn ICC
(OEM Version)
Direct Cable Connection
(no ICC)
Surface Inductive Modem (SIM) 35 mA 20 mA Not allowed 20 mA
Inductive Modem Module (IMM) --
Low power transmit setting (1)
Not allowed 1.2 mA Not allowed 1.3 mA
Inductive Modem Module (IMM) --
Full power transmit setting (1)
12 mA 4 mA Not allowed 40 mA
Underwater Inductive Modem (UIM) (2) Not allowed Not allowed 15 mA Not allowed
Notes:
(1)  IMM power setting is set with SetEnableFullPwrTx= command.
(2) Underwater Inductive Modem is a component used by OEMs to integrate Sea-Bird's inductive modem technology with a third-party sensor.

ADDITIONAL INFORMATION / LINKS:

Documentation -- manual, photos, technical papers, application notes, etc.

Sales Information -- options, accessories, cables, mount kits, spares, etc.

Links to Other Instruments of Interest

[HRule Image]

Specifications are subject to change without notice.  

Sea-Bird Home     Phone: (+1) 425-643-9866     E-mail: seabird@seabird.com