Sea-Bird Electronics            Products            Support            Software            Sales            Service            Search/Sitemap

MicroCAT C-T (P optional) Sensor
SBE 37-SIP
(S
erial Interface & integral Pump)

[HRule Image]

Print version     Current manual - RS-232 or RS-485 (older manuals)    Reference Sheet - RS-232 or RS-485     Configuration options & accessories

Deploy in orientation shown (connector at bottom) for proper operation
SBE 37 MicroCAT CTD Moored Sensor
Titanium housing; plastic housing also available

SUMMARY

DESCRIPTION

The SBE 37-SIP MicroCAT is a high-accuracy conductivity and temperature (pressure optional) sensor with Serial Interface and integral Pump, which includes a non-volatile memory. Externally powered, it is useful as a stand-alone monitoring device, and is easily integrated with current meters, ROVs, AUVs, towed sonars, and other instrumentation platforms. Constructed of non-corroding materials to ensure long life with minimum maintenance, the MicroCAT is rated for operation to 350 meters (plastic ShallowCAT housing) or 7000 meters, or pressure sensor full-scale range.

Calibration coefficients are stored in EEPROM, providing data output in ASCII engineering units (decimal or XML format); raw output is also available. The data always includes Conductivity, Temperature, and (optional) Pressure; time, sound velocity (Chen-Millero), salinity, depth, and density can also be output.

SENSORS

The MicroCAT retains the temperature and conductivity sensors used in our time-proven SeaCAT and SeaCAT plus. Electrical isolation of the conductivity electronics eliminates any possibility of ground-loop noise. The unique internal-field conductivity cell permits the use of expendable anti-foulant devices. The aged and pressure-protected thermistor has a long history of exceptional accuracy and stability.

The optional strain-gauge pressure sensor is available in eight ranges, to a maximum depth of 7000 meters. Compensation of the temperature influence on pressure offset and scale is performed by the MicroCAT’s CPU.

PUMP

The integral pump typically runs for 1.0 sec each time the MicroCAT samples, providing the following advantages:

COMMUNICATIONS AND INTERFACING

The MicroCAT communicates via RS-232 or RS-485 serial interface:

User-selectable operating modes include:

SOFTWARE

The MicroCAT is supplied with a powerful Windows software package, Seasoft© V2, which includes:

SPECIFICATIONS

  Measurement
Range
Initial
Accuracy
Typical
Stability
Resolution
Conductivity 0 to 7 S/m
(0 to 70 mS/cm)
± 0.0003 S/m 
(0.003 mS/cm)
0.0003 S/m
(0.003 mS/cm)
per month
0.00001 S/m
(0.0001 mS/cm)
Temperature
 (°C)
-5 to 45 ± 0.002 (-5 to 35 °C);
± 0.01 (35 to 45 °C)
0.0002 per month 0.0001
Optional Pressure 20 / 100 / 350 / 600 / 1000 / 2000 / 3500 / 7000 m
(meters of deployment depth capability)
± 0.1% of full scale range 0.05% of full scale range per year 0.002% of full scale range
         

Clock Stability: 5 sec/month

Memory: 8 Mbyte; capacity in excess of 530,000 samples

Input Power: 0.25 Amps at 9 - 24 VDC

Quiescent Current: 30 microAmps *

Communication Current: 4.3 milliAmps *

Acquisition Current: 9.1 milliAmps (excluding pump) *

* Power consumption values are for RS-232 interface; for RS-485 interface, see RS-485 manual.

Acquisition Time: 0.9 - 2.7 sec/sample, dependent on sampling mode and inclusion of pressure sensor

Pump Current: 25 milliAmps

Housing, Depth Rating, and Weight:

Housing Material Depth Rating Weight
Plastic 350 m (1148 ft) With pressure, without clamp or guide:
2.3 kg (5.0 lbs) in air
1.4 kg (3.2 lbs) in water

With pressure, with clamp and guide:
2.5 kg (5.6 lbs) in air
1.6 kg (3.5 lbs) in water

Titanium 7000 m (23,000 feet) Without pressure sensor or clamps
3.0 kg (6.5 lbs) in air
1.8 kg (4.0 lbs) in water
     

Dimensions: in millimeters (inches)

Connector (RS-232 shown. For RS-485: pin 2 is RS-485A, pin 3 is RS-485B):

ADDITIONAL INFORMATION / LINKS:

Documentation -- manual, photos, technical papers, application notes, etc.

Sales Information -- options, accessories, cables, mount kits, spares, etc.

Software -- components of Seasoft V2

Links to Other Instruments of Interest

[HRule Image]

Specifications are subject to change without notice.

Sea-Bird Home     Phone: (+1) 425-643-9866     E-mail: seabird@seabird.com