Part # | Description | Notes | ||
19plus V2 | SEACATplus Version 2 PROFILER Pumped Conductivity, Temperature, Depth Recorder - Includes SBE 5M submersible pump, 64 MB memory, 6 differential A/D channels (0 - 5 volt input range), 1 RS-232 data input channel, data I/O & pump Y-cable (PN 17709), 2.5 meter data I/O cable (PN 801225), SEASOFT software, & complete documentation. Specify: housing (depth) & pressure sensor selections. Configurations that include pumped SBE 43 or other pumped auxiliary sensors require SBE 5T or 5P; see option 19p-4_. |
19plus V2 in plastic housing with SBE 5M plastic pump & associated plumbing (plumbing configuration shown for 19plus V2 to be deployed in vertical orientation) 19plus V2 is intended primarily as a profiling CTD, sampling at 4 Hz, but can also be used in moored mode to obtain long-term time series data. 19plus V2 is powered by internal batteries, & records data in memory. In addition, in profiling mode, 19plus V2 can be deployed:
Notes: |
||
SBE 19plus V2 Housing (depth rating) Selections -- MUST SELECT ONE | ||||
19p-1a | 600 m plastic housing for CTD & SBE 5M plastic pump | SBE 5M pump is rated to 600 m with plastic housing or 10,500 m with titanium housing. | ||
19p-1b | 7000 m titanium housing for CTD & SBE 5M titanium pump | |||
SBE 19plus V2 Standard Strain Gauge Pressure Sensor Selections -- MUST SELECT ONE (or select optional Digiquartz below) | ||||
19p-2a | 20 m strain gauge pressure sensor |
Pressure sensor is installed in connector end cap, & is not field replaceable / swappable. While highest
pressure rating gives you most flexibility in using 19plus
V2, it is at expense of
accuracy & resolution. It is advantageous to use lowest range
pressure sensor compatible with your intended maximum operating
depth, because accuracy & resolution are proportional to
pressure sensor's full scale range. For example, comparing 2000 &
7000 m sensors --
|
||
19p-2b | 100 m strain gauge pressure sensor | |||
19p-2c | 350 m strain gauge pressure sensor | |||
19p-2h | 600 m strain gauge pressure sensor | |||
19p-2d | 1000 m strain gauge pressure sensor | |||
19p-2e | 2000 m strain gauge pressure sensor | |||
19p-2f | 3500 m strain gauge pressure sensor | |||
19p-2g | 7000 m strain gauge pressure sensor | |||
SBE 19plus V2 Optional Digiquartz Pressure Sensor (instead of standard strain gauge) | ||||
19p-2i | 45 psia (20 m) Digiquartz with temperature compensation |
Digiquartz pressure sensors provide better accuracy
(0.02% of full scale range vs 0.1% of full scale range) &
resolution than strain gauge pressure sensors. See SBE 19plus V2 Strain Gauge Pressure Sensor Options
above for photo of pressure sensor port; when used with Digiquartz,
nylon pressure capillary fitting screws into port. Pressure sensor is installed in connector end cap, & is not field replaceable / swappable. While highest pressure rating gives you most flexibility in using 19plus V2, it is at expense of accuracy & resolution. It is advantageous to use lowest range pressure sensor compatible with your intended maximum operating depth, because accuracy & resolution are proportional to pressure sensor's full scale range. For example, comparing 2000 psia (1400 m) & 6000 psia (4200 m) sensors --
|
||
19p-2j | 100 psia (60 m) Digiquartz with temperature compensation | |||
19p-2k | 200 psia (130 m) Digiquartz with temperature compensation | |||
19p-2l | 300 psia (200 m) Digiquartz with temperature compensation | |||
19p-2m | 400 psia (270 m) Digiquartz with temperature compensation | |||
19p-2n | 1000 psia (680 m) Digiquartz with temperature compensation | |||
19p-2o | 2000 psia (1,400 m) Digiquartz with temperature compensation | |||
19p-2p | 3000 psia (2,000 m) Digiquartz with temperature compensation | |||
19p-2q | 6000 psia (4,200 m) Digiquartz with temperature compensation | |||
19p-2r | 10,000 psia (6,800 m) Digiquartz with temperature compensation | |||
19p-2s | 15,000 psia (10,500 m) Digiquartz with temperature compensation | |||
SBE 19plus V2 Pump Upgrade Option | ||||
19p-4a | SBE 5P plastic pump instead of SBE 5M plastic pump (required if using SBE 43 DO sensor) |
Standard
SBE
5M pump is intended for providing pumped conductivity only, for
improved conductivity response over a non-pumped system. Larger, more powerful SBE
5P or 5T is
required if also planning to pump auxiliary sensors (such as
dissolved oxygen, etc.). SBE 5P plastic pump is rated for 600 m; SBE 5T titanium pump is rated to 10,500 m. Operational characteristics of SBE 5P & 5T are identical. |
||
19p-4b | SBE 5T titanium pump instead of SBE 5M titanium pump (required if using SBE 43 DO sensor) | |||
SBE 19plus V2 Wet-Pluggable Connector Option | ||||
19p-5 | Wet-pluggable (MCBH) connectors instead of Standard (XSG) connectors on SEACAT bulkhead connectors, data I/O & pump Y-cable, & data I/O cable |
Wet-pluggable connectors may be mated in wet
conditions. Their pins do not need to be dried before mating. By
design, water on connector pins is forced out as connector
is mated. However, they must not be mated or un-mated while
submerged. Wet-pluggable connectors have a non-conducting guide pin
to assist pin alignment (exception: 6-pin connectors) & require less force to mate, making them
easier to mate reliably under dark or cold conditions, compared to
our standard (Impulse XSG/RMG) connectors. Like standard connectors,
wet-pluggables need proper lubrication & require care during use
to avoid trapping water in sockets. |
||
SBE 19plus V2 Auxiliary Sensor & Integration Options | ||||
19p-6a | SBE 43 Dissolved Oxygen Sensor (Profiling Configuration), 7000 m (cable & mount included, requires option 19p-4a or -4b) |
General Information:
Sensor-Specific Information:
Moored Application Information:
|
||
19p-6b | SBE 43 Dissolved Oxygen Sensor (Profiling Configuration) with Wet-pluggable connector, 7000 m (cable & mount included, requires option 19p-4a or -4b & 19p-5) | |||
19p-6c | SBE 43 Dissolved Oxygen Sensor (Profiling Configuration), 600 m plastic (cable & mount included, requires option 19p-4a or -4b) | |||
19p-6d | SBE 43 Dissolved Oxygen Sensor (Profiling Configuration) with Wet-pluggable connector, 600 m plastic (cable & mount included, requires option 19p-4a or -4b & 19p-5) | |||
19p-7a | SBE 18 pH sensor, for use in profiling mode only, 1200 m (cable & mount included) | |||
19p-7b | SBE 18 pH sensor, for use in profiling mode only, with Wet-pluggable connector, 1200 m (cable & mount included – Requires option 19p-5) | |||
19p-8a | SBE 27 pH/ORP sensor, for use in profiling mode only, 1200 m (cable & mount included) | |||
19p-8b | SBE 27 pH/ORP sensor, for use in profiling mode only, with Wet-pluggable connector, 1200 m (cable & mount included – Requires option 19p-5) | |||
YMOLD | Extra charge for Y-cable to connect two (2) or more sensors to one (1) auxiliary sensor input bulkhead connector on CTD | |||
19p-9a | Underwater power/data interface module (PDIM) integrated for use with SBE 36 CTD Deck Unit or 33 Carousel Deck Unit (includes PDIM, interface cable, & mounting hardware) |
SBE 36 CTD Deck Unit & PDIM provide surface power &
real-time data acquisition & control for 19plus V2, allowing
deployment with 10,000 m long single-conductor sea cables. SBE
36 is installed on ship, while PDIM is mounted on or near 19plus
V2.
See SBE 36 description for more information on interface. SBE 33 Deck Unit is typically used to provide surface power & real-time data acquisition & control for 19plus V2 & SBE 32 (32, 32C, or 32SC) Carousel Water Sampler or SBE 55 ECO Water Sampler. However, it can be used with PDIM instead of Carousel or ECO to provide same functionality as SBE 36 & PDIM. PDIM housing is anodized aluminum, & is rated to 6800 m. PDIM connector type (standard or wet-pluggable) must match 19plus V2 connector type. 19p-9a & -9b do not include SBE 36 or 33 Deck Unit; order separately.
|
||
19p-9b | Underwater power/data interface module (PDIM) with Wet-pluggable connectors integrated for use with SBE 36 CTD Deck Unit or 33 Carousel Deck Unit (includes PDIM, interface cable, & mounting hardware – Requires option 19p-5) | |||
SBE 19plus V2 Hardigg Shipping Case Option | ||||
19p-10 | Hardigg shipping case (AL4915-1105) instead of wood crate, for SBE 19plus V2 in optional cage |
Hardigg
shipping case with custom foam inserts holds SBE 19plus
V2 with auxiliary sensors in optional cage (SBE 9plus shown; SBE 19plus V2 similar).
Note: 19plus V2 with
optional Digiquartz pressure sensor requires a longer cage (PN
801270) than 19plus V2 with standard strain-gauge
pressure sensor. PN 801270 cage does not fit in this Hardigg
shipping case; contact Sea-Bird to discuss purchasing a
larger case. |
||
SBE 19plus V2 Spares & Accessories | ||||
801269 | Stainless-steel protective cage, 40 in. (101 cm) tall (SEACATplus) |
Many
users purchase optional cage to protect system from damage during
deployment.
See document 67171 for installation details. Notes:
|
||
801270 | Stainless-steel protective cage, 48 in. (123 cm) tall (SEACATplus with Digiquartz) | |||
50396 | Stainless-steel protective cage, 38 in. (96 cm) tall, used in SBE 32 Carousel CTD extension stand | |||
50434 | SBE 16plus V2/19plus V2 Spares Kit (Standard connectors) - Complete support kit containing spare data I/O cable, bulkhead connectors, dummy plugs, Triton X-100, o-ring lubricant, & other mechanical spares & maintenance items. | Order appropriate spares kit for connector type on 19plus V2: | ||
50435 | SBE 16plus V2/19plus V2 Spares Kit (Wet-pluggable connectors) - Complete support kit containing spare data I/O cable, bulkhead connectors, dummy plugs, Triton X-100, o-ring lubricant, & other mechanical spares & maintenance items. | |||
801511 | 9 NiMH cells in rechargeable drop-in pack (10.8V/8 Ah) |
|
||
90504 | NiMH battery charger for 6, 9, or 12 cells / 8 Ah D-cell packs. Provides full recharge in 2 hours, includes AC power cord & battery charging cable | |||
801641 | 9 Ni-Cad cells in drop-in rechargeable pack (10.8V/4.4 Ah) |
19plus
V2
comes standard with alkaline batteries, but can be powered with
rechargeable Ni-Cad batteries instead. Batteries are easily accessed
by unscrewing battery end cap. Note: As of January 2011, Sea-Bird does not manufacture the 90226 Ni-Cad battery charger, because parts are no longer available. We are still able to provide the battery pack, for customers who already have a charger. |
||
50288 | SBE 19plus19plus V2 Moored Mode Conversion Kit -- anti-foulant holders & hardware for user conversion from profiling to moored mode configuration. AF24173 anti-foulant devices (PN 801542) are NOT included. Note: When installed on 19plus/19plus V2 at time of order, T-C Duct & related parts normally installed on profilers are provided as a separate Profiling Conversion Kit, PN 50345 (order 801542 anti-foulant devices separately, or PN 50288.1 Kit with anti-foulant devices) | 19plus V2 is intended primarily for use as profiling instrument, & does not come standard with anti-foulant device fittings. Conversion kit allows user to retrofit 19plus V2 with anti-foulant device fittings for moored applications. Anti-foulant device fittings attach to each end of conductivity cell. See document 67114. Only 1 conversion kit is needed per instrument; order replacement Anti-Foulant Devices as needed (see next item). | ||
801542 | AF24173 Anti-foulant Device (pair, bagged, labeled for shipping) |
Anti-foulant
devices fit into anti-foulant device cups at each end of
conductivity cell. Anti-foulant devices included with standard
shipment; these are spares. Useful life varies, depending on several factors. We recommend that customers consider more frequent replacement when high biological activity & strong current flow (greater dilution of anti-foulant concentration) are present. Moored instruments in high growth & strong dilution environments have been known to obtain a few months of quality data, while drifters that operate in non-photic, less turbid deep ocean environments may achieve years of quality data. Experience may be strongest determining factor in specific deployment environments. |
||
20200 | USB to Serial Port Adapter, FTDI UC232R-10 (connects computers with USB ports to RS-232 instruments) | Many newer PCs & laptop computers have USB port(s) instead of RS-232 serial port(s). USB serial adapter plugs into USB port, & allows a serial device to be connected through adapter. Multi-port adapters are available from other companies; see Application Note 68. | ||
31634 | Hardigg shipping case (AL4915-1105), for SBE 19plus/19plus V2 in optional cage |
Hardigg
shipping case with custom foam inserts holds SBE 19plus
V2
with auxiliary sensors in optional cage (SBE 9plus shown; SBE 19plus V2 similar).
Note: 19plus V2 with optional Digiquartz pressure sensor requires a longer cage (PN 801270) than 19plus V2 with standard strain-gauge pressure sensor. PN 801270 cage does not fit in this Hardigg shipping case; contact Sea-Bird to discuss purchasing a larger case. Notes
regarding features of Hardigg case supplied by Sea-Bird: |
||
50275 | SBE 19plus/19plus V2 magnetic switch assembly | Replacement for magnetic switch, which is used to start & stop logging. See document 67101. | ||
41124 | Battery cover/contact plate, replaces 80076.1 | Cover plate for alkaline batteries. | ||
801225 | Data I/O cable, Standard connector (RMG-4FS) w/ DB-9S, 2.4 m (DN 32421) | Data I/O cable connects to 4-pin end of pump-data I/O Y-cable. These test cables are used for setting up system & uploading data from memory after recovery. Applicable cable is included with 19plus V2; these are spares. | ||
801374 | Data I/O cable, Wet-pluggable (MCIL-4FS) w/ DB-9S, 2.4 m (DN 32715) | |||
17709 | Y-cable, Pump-Data I/O (DN 31551) | These Y-cables connect 6-pin data I/O - pump bulkhead connector to pump (2-pin leg) & to data I/O cable (4-pin leg). | ||
171883 | Y-cable, Pump-Data I/O, Wet-pluggable (DN 32896) | |||
90087 | Universal plumbing kit (includes pump air release valve, Y-fitting, & tubing) -- Application note 64-1 | Application Note 64-1 details installation of plumbing for SBE 43 & pump on a CTD. | ||
50087 | Cell filler/storage device (Application Note 34) | For cleaning conductivity cell after each use & storing instrument between uses. See document 67043 & Application Note 2D: Instructions for Care and Cleaning of Conductivity Cells. | ||
various | Plumbing spares | For air bleed valve & y-fitting, & assorted sizes of Tygon tubing, see SBE 5M Configuration Details. | ||
Underwater Cable for Hand-Hauled, Real-Time Profiling (see Application Note 59 for proper use & limitations) | ||||
801150 | Load-bearing Data cable, Standard connector, with DB-9S, 30 m (100 ft) (DN 32284) |
These cables are for hand-hauling 19plus
V2 &
acquiring real-time data. Cables are not intended for static working
loads above 45 kg (100 lbs); working loads above 18 kg (40 lbs) may
be difficult to handle without winch. Minimum recommended cable bend
radius is 10 cm (4 in.) (e.g., 20 cm sheave block nominal diameter).
See Application Note 59:
A Load-Bearing Underwater Cable for Hand-Hauled, Real-Time Profiling. Select a cable with connector (standard XSG or wet-pluggable MCBH) to match your 19plus V2. |
||
801295 | Load-bearing Data cable, Standard connector, with DB-9S, 50 m (165 ft) (DN 32284) | |||
801140 | Load-bearing Data cable, Standard connector, with DB-9S, 61 m (200 ft) (DN 32284) | |||
801153 | Load-bearing Data cable, Standard connector, with DB-9S, 100 m (330 ft) (DN 32284) | |||
801301 | Load-bearing Data cable, Standard connector, with DB-9S, 126 m (415 ft) (DN 32284) | |||
801148 | Load-bearing Data cable, Standard connector, with DB-9S, 183 m (600 ft) (DN 32284) | |||
801371 | Load-bearing Data cable, Wet-pluggable, with DB-9S, 30 m (100 ft) (DN 32643) | |||
801372 | Load-bearing Data cable, Wet-pluggable, with DB-9S, 60 m (200 ft) (DN 32643) | |||
801337 | Load-bearing Data cable, Wet-pluggable, with DB-9S, 100 m (330 ft) (DN 32643) | |||
801338 | Load-bearing Data cable, Wet-pluggable, with DB-9S, 200 m (660 ft) (DN 32643) |
Product configuration options listing for all products
Additional cables, mount kits, and spares may be found at: Cables, Mount kits for Sea-Bird sensors, Mount kits for Third Party sensors, Spare parts
Last modified: 25 Apr 2011
Sea-Bird Home Phone: (+1) 425-643-9866 Fax: (+1) 425-643-9954 E-mail: seabird@seabird.com